Semi Automatic and Manual Manifold HTM/ISO Specification Sheet

Semi Automatic and Manual Manifold HTM/ISO Specification Sheet
1 / 12 PagesView full catalog

Semi Automatic and Manual Manifold HTM/ISO Specification Sheet

Product catalog summary
Semi Automatic Manifold Control Systems
The BeaconMedæs Semi Auto Manifold is engineered to supply piped medical gas from high-pressure cylinders, ensuring a continuous supply. It complies with ISO and HTM standards, emphasizing safety and functionality. The system automatically switches between duty and standby banks when pressure drops, with remote alarms for cylinder changes. It operates at nominal pressures of 4 bar (max 5 bar) and 7 bar (max 8.5 bar), with options for standard or full-feature units. The manifold can serve as a main or emergency reserve, with pressure regulation ensuring high flow rates without significant pressure drops. Materials used are halogen-free, and modular headers allow flexible configurations.
Manual Manifold Control Systems
The Manual Manifold is designed for small installations with low flow demands, requiring manual changeover of cylinder banks. It shares similar specifications with the semi-automatic system, including pressure regulation and material standards. The manual system is also available in standard or full-feature units and can function as a main or emergency reserve manifold.
Materials and Components
All polymers and elastomers used must be halogen-free, with specific prohibitions on certain materials like PTFE. Non-return valves must have metallic seats with ceramic balls. Modular header manifolds provide connection points for flexible tailpipes, with non-return valves at each connection to prevent system damage.
Installation and Configuration
Typical installations are detailed for various cylinder types, with considerations for optional high-pressure bank valves affecting manifold width. The document includes schematic diagrams and tables listing main features and components for both semi-automatic and manual systems.
Overview
This document provides detailed specifications and part numbers for various gas manifold systems and accessories offered by Atlas Copco Medical. It includes both semi-automatic and manual manifold systems, along with optional accessories and spare parts.
Manifold Systems
  • Semi-Automatic Manifolds: Available in standard and full feature models, with options for 4 bar and 7 bar pressure settings. They support gases like N2O, CO2, O2, AIR, and N2, with specific part numbers for each configuration.
  • Manual Manifolds: Offered in 4 bar and 7 bar pressure settings, supporting similar gases as the semi-automatic models, with distinct part numbers.
Accessories and Options
  • Exhaust Pipe Connection Kits: Available for both semi-automatic and manual systems.
  • Termination Box: For remote alarm integration.
  • Pressure Switches: Options with and without line contact monitors for 4 bar and 7 bar systems.
  • High Pressure Bank Valves: Available as a pair for enhanced system control.
Spare Parts
  • Manifold Header Components: Includes corner connectors and cylinder spare racks for various configurations.
  • Tailpipes: Specific tailpipes for different gases, including oxygen, nitrous oxide, air, and CO2, with various fitting standards.
Important Notes
The document emphasizes that specifications are subject to change for product improvement. It advises contacting the company for the latest specifications and information.
See more

Catalog excerpts

Semi Automatic and Manual Manifold HTM/ISO Specification Sheet-1

Semi Automatic and Manual Manifold Systems SPECIFICATION SEMI AUTOMATIC & MANUAL MANIFOLDS Semi Automatic Manifold Control Systems The BeaconMedæs Semi Auto Manifold is designed to supply piped medical gas where continuity of supply is essential, and where the gas is to be supplied from high pressure gas cylinders. All individual components conform to ISO and HTM requirements to form a medical gas control panel to which maximises safety requirements with simplified function. The Semi Automatic Manifold shall be supplied fully tested. All polymers and elastomers in the gas flow that can be subjected to working pressure greater than 3000 kPa shall be halogen-free. The use of PTFE, PCTFE, Viton and other halogenated polymers in these applications is strictly prohibited. Non-return valves fitted to header manifolds shall have a metallic seat with ceramic ball. Soft seat nonreturn valves utilising polymers or elastomers are not acceptable. The duty bank is determined by operating the right hand regulator leaver, which will adjust the set pressure to determine the lead regulator. Rotate clockwise to set right bank as duty, or anticlockwise for left bank duty. The duty bank will depleted until the pressure drops below the changeover parameters. The gas supply will then automatically changeover to the standby bank. A signal can be taken to a remote alarm from the contact gauges to alert the requirement to change cylinders. The duty selector leaver should then be switch over to the full running bank, and the empty cylinders changed. This cycle is then repeated to maintain continuous supply. The manifold control panel shall be supplied as nominally 4 bar (5 bar max.) and 7 bar (8.5 bar max), either standard units or full feature (see table 1 for details). The 4 bar manifold is available with 100 or 250 bar contents gauges depending on the pressure of the cylinders to be used. These units can be supplied as standard unit or full feature. See table 1 for list of critical components included and optional extras supplied as bolt on kits. The semi auto manifold can be used as either main supply or emergency reserve manifold (ERM). When used as a emergency backup manifold the line pressure regulator should be set slightly lower than the primary supply pressure. This will prevent it from feeding to the pipeline during normal operation of the primary system. The line pressure regulator can be increased to nominal distribution pressure in the event of emergency. The manifold shall be supplied with a non-return valve and lockable line isolation valve for connection to the distribution system, enabling a continuous supply of gas to the distribution system upon failure of the normal supply. Pressure Regulation There shall be two separate stages of pressure regulation to enable high peak flow rates without a significant reduction in downstream pressure. Multistage regulators combined into a single unit are not acceptable. The inlet of the 1st stage regulator shall be protected from the particulate matter by a 25µm sintered brass filter. Sintered aluminium bronzes shall not be used. Regulators shall comply with BS EN ISO 10524-2 and shall be supplied with documented test reports upon request, confirming successful completion of the oxygen ignition tests stated therein. The manifold control system shall be capable of supplying a flow of 1,000 l/min to a nominal 400 kPa distribution system, 2,000 l/min to a nominal 700 kPa distribution system, based on a maximum allowable pressure drop of 10% from static to full flow. All regulators shall be protected from over-pressurisation by relief valves. A pre-piped kit shall be available for indoor installation, to enable the gas to be taken away and vented to atmosphere safely. Relief valves shall not be vented into the manifold room. Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL www.beaconmedaes.com Modular Header Manifolds Modular header manifolds shall provide connection points for flexible cupro nickel tailpipes. They shall be available in ‘primary’ and ‘secondary’ configurations, with either single or double cylinder connection points. ‘Primary’ headers shall connect directly to the manifold control system with extensions for additional cylinders being provided by the addition of ‘secondary’ headers. Non-return valves shall be fitted to each tailpipe connection point to protect the system in the event of a tailpipe fracture. Corner connectors shall be available to enable installation of manifold headers around corners of the manifold room. A custom length corner connector shall also be available to enable header manifolds to be installed in a ‘U’ configuration across 3 adjacent walls of a manifold room. Schematic Diagram Lockable line ball valve (Valve 'B') Alternative supply source Test Point Medical gas pipeline distribution system Lockable line ball valve (Valve 'A') Exhaust (piped to safe location) Non-return valve 2nd stage relief valve Line pressure gauge 2nd stage regulator Optional Bank Valve 1st stage relief valve Contact gauge Contact gauge Sintered filter Sintered filter Optional Bank Valve Non-return valve (With gas specific connection) High pressure cylinders compete with Isolation valve

 Open the catalog to page 1
Semi Automatic and Manual Manifold HTM/ISO Specification Sheet-2

Image shown without cover fitted 10 (Valve shown in Closed Posi tion) 9 (Valve shown in Open ition) Table 1: Main features/components Notes: Item 12 is recommended for indoor installations. Item 16 & 17 are used to connect alarm outputs to a Medipoint medical alarm. Image showing typical optional header and tailpipe Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL

 Open the catalog to page 2
Semi Automatic and Manual Manifold HTM/ISO Specification Sheet-3

Typical Installation For Use With ‘J’ & ‘G’ Type Cylinder Typical panel weight is 15kg. Note - If option high pressure bank valves are used (see figure 1, item 19) the manifold width increases by 110mm per side. Note - ‘J’ type cylinders typically for Oxygen and Medical Air. ‘G’ type cylinders typically for Nitrous Oxide and N2O/O2 mix (Entonox). Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL www.bea

 Open the catalog to page 3

All BeaconMedaes catalogs and technical brochures

  1. Central alarm

    4  Pages

  2. Magnis MSV

    2  Pages

  3. mVAC

    12  Pages

Archived catalogs

  1. Local alarm

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.