Overview: The document details the specifications and functionalities of the TotalAlert Infinity™ medical gas notification system by BeaconMedæs. It is designed to comply with HTM 02-01, BS EN 60601-1-2, and BS EN ISO 7396-1 standards, providing central and area alarm signals for medical gas systems.
Specifications: The system features an 8” touch screen LCD for configuration and monitoring, with a built-in Human-Machine Interface (HMI). It supports Ethernet connectivity for remote monitoring and configuration, and includes a web server for remote access to alarm signals and pressures. The system is compatible with BACnet for integration with Building Automation Systems.
Panel Functionality: The alarm panel displays a green NORMAL condition when no alarms are active. It includes a motion sensor for backlight sleep mode, a full diagnostic self-test feature, and antimicrobial protection on the front panel. The system supports custom alarm instructions and mirroring of digital readings from other panels.
Connectivity and Communication: The system allows for electronic notifications via email or text, and supports WiFi access to panel web pages. It includes a history log for event tracking and supports 4-20mA transducer setups for monitoring various conditions.
Power and Installation: The alarm panel operates on 100 to 250 VAC and converts to 24 VDC internally. It is designed for safe user interaction with low voltage electronics and protective guards. Installation guidelines specify environmental conditions for transport and storage.
Models and Configurations: Various models are available, including central, zone, and combination alarms, each with specific input and output capabilities. The document lists example model numbers and configurations for different monitoring needs.
Dimensions: The document provides detailed dimensions for both concealed and surface installation of the alarm panels.
See more

Catalog excerpts

TotalAlert Infinity HTM/ISO Specification Sheet-1

TotalAlert Infinity™ Medical Gas Notification System MEDICAL GAS NOTIFICATION SYSTEM SPECIFICATION Medical Gas Notification System The TotalAlert Infinity™ medical gas notification system provides central and area alarm signals as required by the latest edition of HTM02-01. The medical gas notification system shall fully comply with the requirements of HTM 02-01, BS EN 60601-1-2 and BS EN ISO 7396-1. The Right Instruction The medical gas notification system shall allow: • Set up customised instructions for each alarm signal, to appear on the screen when the signal is in alarm. • Create custom instructions with the touch screen interface on the alarm panel LCD. • Incorporate user-defined instructions into an emergency response plan by the facility. Panel Functionality • Central, Zone, and Combination medical gas notification systems shall have an 8” touch screen LCD display. Central Alarm Panel (up to 20 BeaconMedaes Line Contact Monitoring signals.) • Configuration of the medical gas notification system shall be done via build-in touch-screen Human-Machine Interface (HMI), no additional tools shall be required. • Duplication (cloning) of the configuration should be foreseen e.g. by using removable memory board. Remote Ethernet monitoring and secured configuration shall be foreseen without additional software. • A Green NORMAL condition on the home screen shows that there are no alarms active. • The home page includes a location/area served badge for the alarm panel, customisable with the keyboard feature of the touch screen during the set up process. • LCD display contains an adjustable background with many colour/ brightness options. • LCD display backlight includes a “sleep mode” option via motion sensor to preserve screen life, adjustable from 5 to 60 minutes. By default, the alarm sleep mode option is not active and the backlight remains on all of the time. If option is active and an alarm occurs during “sleep mode”, the backlight turns on and remains on during the alarm condition. Combination alarm can monitor up to 20 BeaconMedaes Line Contact Monitoring signals and up to 6 digital gas sensors. • A full diagnostic self-test of the alarm tests the alarm signals (Green, Yellow, and Red), the LED indicators on the front panel, the audible horn, and displays all of the alarm set points. Initiate test with the touchscreen controls. Antimicrobial Additive User accessible overlay material on the front panel shall include a silver antimicrobial additive for inherent antimicrobial protection. Efficiency of the additive according to ISO 22196 shall be >99% for sample of species such as E coli and MRSA. Ethernet Connectivity with Embedded Web Page • A built-in web server in each panel allows a remote operator to display the alarm’s web page and view alarm signals, pressures, and other networked devices. • Ethernet communication is compatible with BeaconMedaes TotalAlert Embedded source equipment, TotalAlert Embedded manifolds, and TotalAlert alarm panels. Zone Alarm Panel (up to 8 gases) • The panel’s web page contains links to all other TotalAlert devices on BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● [email protected] Constructionline Reg No. 75463 ● CHA

 Open the catalog to page 1
TotalAlert Infinity HTM/ISO Specification Sheet-2

the network, including other alarm panels and source equipment. • Electronic notification is available through e-mail/text messaging. • Accessible through any SMTP gateway. • Allows for remote alerts of alarm conditions. History Log • Event log accessible on LCD display records all alarms and panel activity, displays most recent 100 events. • WARNING LED alerts the user to warnings and communication issues which are not alarm causing conditions. • Audible SPEAKER produces a minimum sound pressure of 58 dBA at a distance of 1m. • MUTE button indicator LED flashes red during an alarm condition...

 Open the catalog to page 2
TotalAlert Infinity HTM/ISO Specification Sheet-3

• Gas specific digital gas sensor contains a transducer capable of providing factory calibrated signals to the alarm panel. Source alarm signal with custom, alarm specific instruction • Remotely installed sensors may be located up to 1,524 m (5,000 feet) from the alarm panel. • Pipeline connections are 1/4” BSPP threaded connection with sealing washer. • Digital gas sensors include a “heartbeat” flashing green LED light to designate that sensor is functioning correctly. Zone Alarm Panel Configuration • Zone alarm panel monitors 1 to 8 digital gas sensor modules, installed remotely. • Display...

 Open the catalog to page 3
TotalAlert Infinity HTM/ISO Specification Sheet-4

TotalAlert Infinity™ Alarm Models Product number BeaconMed*s • Telford Crescent, Staveley, Derbyshire S43 3PF • Tel: +44 (0) 1246 474 242 • Fax: +44 (0) 1246 472 982 Company Reg No. 2957933 (English) • Company VAT Reg No. 598612590 • www.beaconmedaes.com • [email protected] Constructionline Reg No. 75463 • CHAS accredited contractor • SAFEcontractor Reg No. N00184450

 Open the catalog to page 4
TotalAlert Infinity HTM/ISO Specification Sheet-5

In an effort to continuously improve our products, 4107951155.01 the right is reserved to change the specification of the 2212020128 items described herein at any time. Please contact us for further information and up to date specifications. 08/10/17 Page 5 of 5 TotalAlert Infinity™ Alarm Dimensions 16.4 mm 202.0 mm Finished Wall Concealed Installation Dimensions Front Panel Hinged For Internal Access Surface Installation Dimensions BeaconMedæs Telford Crescent, Staveley, Derbyshire Patent Pending © BeaconMedæs 22-07-

 Open the catalog to page 5

All BeaconMedaes catalogs and technical brochures

  1. Central alarm

    4  Pages

  2. Magnis MSV

    2  Pages

  3. mVAC

    12  Pages

Archived catalogs

  1. Local alarm

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.