ioRT-50

ioRT-50

ioRT-50

Product catalog summary
Overview:
The ioRT-50 is a mobile x-ray therapy unit designed for superficial and intra-operative x-ray therapy, particularly for treating tumors and tumor beds post-resection. It is manufactured by WOmed, a subsidiary of Eckert & Ziegler BEBIG.

Specifications:
  • Maximum high tension: 50 kV for superficial therapy, 70 kV for intra-operative therapy.
  • Maximum tube current: 8 mA for superficial therapy, 7 mA for intra-operative therapy.
  • Maximum power: 500 W.
  • Cooling system: Two-circuit water-water cooler.
  • Selectable energies: 2 for superficial therapy, 1 for intra-operative therapy.
  • Applicator types: Open applicators with focus to skin distance of 15 and 30 mm; spherical applicators with diameters ranging from 35 to 55 mm.
  • Power supply: 230 V AC, 10 AT.

Design and Operation:
  • The unit is mounted on a mobile trolley with a tube arm, allowing for easy positioning and stability during treatment.
  • Operated via a touchscreen panel on the trolley or remotely through a tablet PC.
  • Features include an emergency switch, key switch, x-ray START/STOP switches, and a display for device conditions and treatment time.

Applications:
  • Superficial x-ray therapy: Utilizes open applicators for skin treatment.
  • Intra-operative radiation therapy (IORT): Designed for use in operating theatres with a transmission target tube for easy access to treatment sites.

Additional Features:
  • High-capacity offline cooler for intra-operative use.
  • Micro-processor controlled with service software and remote access capabilities.
  • Spherical applicators provide homogeneous dose distribution and high dose rates for short treatment times.

Contact Information:
Wolf-Medizintechnik GmbH, Am Wachtelberg 15, 07629 St. Gangloff, Germany. Phone: +49 (0) 36606 60441, Fax: +49 (0) 36606 60442, Email: [email protected], Website: www.womed.net
See more

Catalog excerpts

ioRT-50-1

Equipment for x-ray therapy Efficient application for superficial and intra-operative x-ray therapy WOmed An Eckert & Ziegler BEBIG Company

 Open the catalog to page 1
ioRT-50-2

ioRT-50Efficient application for superficial and intra-operative x-ray therapy ioRT-50 x-ray therapy units are intended for superficial x-ray therapy. The ioRT-50 unit consists of a mobile trolley with a mounted x-ray tube stand. It is operated via an adjusted touchscreen panel or remotely via a tablet PC. The tube head is mounted on a spring-loaded arm and can be turned on two axes. The spring-loaded arm itself is easy to move for height setting. A clamping mechanism enables the preselected stand position to remain stable while treating the standing, sitting or lying patient.The trolley control...

 Open the catalog to page 2
ioRT-50-3

ioRT-50 x-ray therapy units deliver intra-operative radiation therapy (IORT) to tumors and tumor beds following resection. The transmission target tube design enables the treatment site to be easily reached. The mobile trolley is designed for use inside the operating theatre. All components are inside the trolley cover. For intra-operative use, the cooling system can be prepared in advance, so it can be run offline during the entire treatment procedure. A set of steam-sterilizable spherical applicators enable accurate adjustment to the target. The dose delivery is homogeneous around the spherical...

 Open the catalog to page 3
ioRT-50-4

Wolf-Medizintechnik GmbH Am Wachtelberg 15, 07629 St. Gangloff, Germany Phone: +49 (0) 36606 60441, Fax: +49 (0) 36606 60442 [email protected], www.womed.net WOmed An Eckert & Ziegler BEBIG Company

 Open the catalog to page 4