Product Overview
Bertec's AM6501 amplifier is designed for use with force plates and force transducers, providing robust performance for research, clinical, and industrial applications worldwide. It features a digital input and an analog output with a fixed gain of 1, facilitating immediate data collection.
Design Specifications
The amplifier converts digital output from the force transducer to an analog output. It includes a female 9-pin D-Sub connector for digital input and a female 15-pin D-Sub connector for analog output. The device offers six output channels with analog signals ranging from ±5V to ±10V. Six analog scale factors are provided to convert these signals into force and moment units. An auto zero button on the front panel helps remove tare load offset.
Unique Features
  • Crosstalk-free calibrated outputs
  • Six channels of ±5V or ±10V analog output
  • Operates on mains power of 100–250V, 50–60Hz
  • Digital gain ratios accurate to 30ppm and analog output autozero accurate to 50ppm
See more

Catalog excerpts

Analog Amplifier AM6501-1

Analog Amplifier AM6501 Product Details and Specifications Bertec’s amplifiers are designed for use with our line of force plates and force transducers. Their solid construction will handle the rigors of any application as our products are used internationally in research, clinical, and industrial facilities. The AM6501 amplifier has a digital input and provides an analog output with fixed gain (gain=1) to allow immediate data collection. Unique Features - Crosstalk-free calibrated outputs Design The AM6501 amplifier takes the digital output of the force transducer and converts it to an analog output. The digital input is a female 9-pin D-Sub connector; whereas, the analog output is a female 15-pin D-Sub connector (see pin assignment). The six output channels provide an analog signal in the range of ±5V or ±10V. Six analog scale factors are provided with the force plate to convert the analog signals to force and moment units. The auto zero button on the front panel is utilized to remove tare load offset. Front Panel - 6 channels of +/- 5 V OR +/- 10 V analog output - Mains powers 100–250V, 50–60Hz - 30ppm accurate digital gain ratios and 50ppm accurate analog output autozero Back Panel STATUS POWER ZERO ANALOG OUTPUT INDICATORS FOR STATUS (TOP) AND POWER (BOTTOM) TRANSDUCER INPUT 2500 Citygate Drive Columbus, Ohio 43219 bertec.com POWER INPUT

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.