Overview: The HFlow-1 is a flowmeter funnel designed to measure both bulk density and flow rate in pharmaceutical and metal powders. It offers three standard funnels: Hall, Carney, and Gustavsson, and complies with USP, Ph. Eur., ASTM, and ISO standards, making it suitable for diverse applications.
Specifications:
  • Funnels: Hall (Φ 5 mm, 60°), Carney (Φ 2.5 mm, 60°), Gustavsson (Φ 2.5 mm, 30°).
  • Orifice Diameter: Hall (5 ± 0.2 mm), Carney (2.5 ± 0.4 mm), Gustavsson (2.5 ± 0.03 mm).
  • Volume of Cups: 25 ± 0.05 ml.
Measurement Procedures:
  • Flow Rate Measurement: Weigh 50 g sample, add to funnel, start timing when powder exits, and record time when last powder exits.
  • Bulk Density Measurement: Empty and weigh cup, add sample, adjust height, scrape, and weigh again to calculate bulk density using mass and volume.
Compliance:
  • Flow Rate: USP <1174>, Ph. Eur. <2.9.36>, ISO <4490>, ASTM , ASTM , ISO <13517>.
  • Bulk Density: ISO <4490>, ASTM , ISO <3923-1>, ASTM , ISO <13517>.
Interoperability: The HFlow-1 can be combined with the Bettersizer 2600 laser diffraction analyzer to balance particle size distribution and flow properties of additive manufacturing (AM) powders.
Application Example: In feedstock preparation, undersized particles may aggregate, affecting powder bed fusion processes by increasing interparticle friction, which can be analyzed using the HFlow-1.
Contact Information: For more details, visit the Bettersize Instruments website or contact via email at [email protected].
See more

Catalog excerpts

HFlow-1-1

The HFlow-1 measures both bulk density and flow rate in pharmaceutical and metal powders, offering a choice of three standard funnels—Hall, Carney, and Gustavsson. Complied with USP, Ph. Eur., ASTM, and ISO standards, it accommodates diverse applications, allowing customers to choose the best fit for their measurement requirements. Bulk Density Measurement with 3 Funnels Flow Rate Measurement with 3 Funnels Carney Funnel Gustavsson Funnel Measurement Procedure Bulk Density 02 Adjust height Flow Rate where ρΒ is the bulk density, M is mass in grams, and V Β is the bulk volume in milliliters or cubic centimeters. 02 Weigh sample Start timing when powder begins to exit the orifice. * Both procedures are applicable across three types of funnels, according to corresponding standards. Record the elapsed time when the last of powder exits the o

 Open the catalog to page 1
HFlow-1-2

Analyzing bulk density and flow rate of various metal powders applied in additive manufacturing (AM). Hall, Carney, and Gustavsson funnels equipped in HFlow-1 have different orifice diameters and angles respectively. They can be employed to test the bulk density and flow rate of a wide range of metal AM powders for evaluating printability. The combination of the laser diffraction analyzer Bettersizer 2600 and HFlow-1 assists in balancing the particle size distribution and flow property of AM powders. In the feedstock preparation process, undersized particles tend to aggregate, which can be detrimental...

 Open the catalog to page 2

All Bettersize catalogs and technical brochures

  1. BetterPyc 380

    9  Pages

  2. PowderPro A1

    5  Pages

  3. BeVision M1

    7  Pages

  4. BeVision S1

    7  Pages

  5. PowderPro A1

    5  Pages

  6. benano-series

    11  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About MedicalExpo
MedicalExpo is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. MedicalExpo is an entity of VirtualExpo Group.