Overview: The document provides detailed specifications and guidelines for hip joint head replacements, specifically focusing on products offered by BEZNOSKA Ltd. These include heads for both hemiarthroplasty and total hip joint arthroplasty.
Specifications:
  • All hip joint heads feature a self-chucking taper 12/14 (euro-taper) with variable depth, affecting the endoprosthesis neck length.
  • Hemiarthroplasty heads are available in metal, with sizes ranging from 38 mm to 56 mm, and biarticular heads from 42 mm to 56 mm. They are made of corrosion-proof stainless steel (ISO 5832-1, ISO 5832-9) and have a UHMWPE lining (ISO 5834-2).
  • Total hip endoprosthesis heads are available in metal or ceramic, with metal heads in diameters of 22, 28, 32, and 36 mm, and ceramic heads (BIOLOX®delta) in 28, 32, and 36 mm.
Materials and Standards:
  • Metal heads are made from corrosion-proof steel (ISO 5832-9) and cobalt alloy CoCrMo (ISO 5832-12).
  • Ceramic heads are made from BIOLOX®delta, a non-allergenic, wear-resistant material (ISO 6474).
Procedures:
  • During surgery, trial articulations are performed using plastic trial heads, which are part of the standard instrumentation set.
  • Correct sizing of the endoprosthesis head is crucial to avoid joint instability or damage.
Instrumentation:
  • Instrumentation sets include matrices for measuring extracted heads and trial heads for both rasp and definitive stem fittings.
Recommendations:
  • Ceramic heads are recommended for use with BEZNOSKA femoral stems due to their superior mechanical properties and reduced wear.
Contact Information: BEZNOSKA, s.r.o., Dělnická 2727, 272 01 Kladno, Czech Republic. Website: www.beznoska.com, Email: [email protected], Phone: +420 312 660 670.
See more

Catalog excerpts

Hip Joint Head Replacements-1

Hip Joint Head Replacements Hip Joint Head Replacements

 Open the catalog to page 1
Hip Joint Head Replacements-2

All hip joint heads delivered by Messrs. BEZNOSKA Ltd. are round and provided with a self-chucking taper 12/14 (euro-taper) of variable depth that determines the final length of the endoprosthesis neck. Heads that are supposed to have a long neck have a conical extender that ensures a firm fixation of the head on the implant stem. The heads are made for hemiarthroplasty and total hip joint arthroplasty. The heads for hemiarthroplasty are metal and come in two options: Standard hemiarthroplasty heads are sized from 38 mm to 56 mm, graded by 2 mm (10 sizes). Biarticular heads are sized from 42...

 Open the catalog to page 2
Hip Joint Head Replacements-3

■ Instrumentation Set During the surgery, the surgeon has to perform a preliminary trial articulation (on a rasp) and a final trial articulation of the implant (on a femoral stem). For these trial fittings, the set of instruments contains two plastic trial heads in two options each - one set of heads for the rasp and one set for the definitive stem. These plastic trial heads (see Fig. 1) are always considered as an integral part of regular instrumentation set delivered for implantation of a concrete type of femoral stem. The plastic trial heads for the ceramic heads are delivered in a separate...

 Open the catalog to page 3
Hip Joint Head Replacements-4

Hemiarthroplasty Head The head is made of corrosion-proof (stainless) steel as per ISO 5832-1, successfully time-tested by leading global manufacturers. The head surface is polished to mirror-like gloss. Biarticular Head The head cover and the inner metal head are made of corrosion-proof steel as per ISO 5832-1 and as per ISO 5832-9, successfully time-tested by leading global manufacturers. The external surfaces of both elements are polished, especially the inner metal head that is polished to mirror-like gloss. The head liner is made of UHMWPE as per ISO 5834-2.

 Open the catalog to page 4
Hip Joint Head Replacements-5

Hemiarthroplasty Head The head is made of corrosion-proof (stainless) steel as per ISO 5832-1, successfully time-tested by leading global manufacturers. The head surface is polished to mirror-like gloss. Biarticular Head The head cover and the inner metal head are made of corrosion-proof steel as per ISO 5832-1 and as per ISO 5832-9, successfully time-tested by leading global manufacturers. The external surfaces of both elements are polished, especially the inner metal head that is polished to mirror-like gloss. The head liner is made of UHMWPE as per ISO 5834-2. Hlavice kobaltova CoCrMo

 Open the catalog to page 5
Hip Joint Head Replacements-6

Ceramic Head - BIOLOX®delta BIOLOX®delta is extremely pure ceramic material on the basis of aluminum oxide as per ISO 6474.The product is bioinert, biocompatible, abrasion-resistant, and non-allergenic. A clinically-tested advantage of BIOLOX®delta heads is a low polyethylene wear compared to metal heads, in combination with UHMWPE cup or cup liner. Ceramic head

 Open the catalog to page 6
Hip Joint Head Replacements-7

■ Combination of heads with other BEZNOSKA implants Cemented cup Cemented cup Non-cemented Non-cemented cup Reconstruction Oval revision cup type Poldi type 02 cup type SF type DUO, insert acetabular cage type TC Ceramic Head Hemiarthroplasty Head Biarticular Head