Overview: The document provides technical specifications and operational details for the Anaesthetic Gas Scavenging System (AGSS) by BGS GENERAL srl, designed to evacuate anesthetic gases using a system of blowers and control mechanisms.
Specifications:
  • The system includes two side channel blowers, a protective cover casing, an electrical and control panel, a remote start/stop/alarms panel, a safety suction filter, and a depression control valve.
  • Nominal capacity is 210 m3/h, with a capacity of 137 m3/h at -10Kpa.
  • Maximum differential pressure is 22 Kpa.
  • Power supply is 230V, 50Hz, with motor power of 2 x 1.5 kW.
  • Sound pressure level is 64 dB(A), and the total weight is 110 kg.
  • Compliance with UNI EN 7396-2:2007 and HTM 02-01 standards.
Components:
  • Suction and exhaust ports are 42 mm in diameter.
  • Includes a suction filter (FMV100), non-return valve, vacuum switch, and vacuum regulator valve.
  • Control panel operates at 230V, 50Hz.
Options:
  • Additional remote panel and multi-voltage power supply options are available.
Remote Alarm Panel:
  • Features include indicators for power supply, standby, normal operation, insufficient vacuum, and blower blockages.
  • Buttons for stop, start, test, and mute functions.
Dimensions:
  • Junction box for wall installation with dimensions: 155 mm x 185 mm x 70 mm.
See more

Catalog excerpts

CGA 130-1

CENTRALE DI EVACUAZIONE GAS ANESTETICI (AGSS) ANAESTHETIC GAS SCRAVENGING SYSTEM (AGSS) CODICE / CODE 1314MF Centrale di Scarico Exhaust Aspirazione Suction Quadro elettrico Control panel soffianti a canale laterale, telaio con carter di copertura e protezione , quadro elettrico di alimentazione e comando, pannello remoto per avvio/arresto/allarmi, filtro di sicurezza in aspirazione, valvola di regolazione depressione. Le soffianti sono dotate di: piedi antivibranti, valvola di non ritorno. Anaesthetic gas scavenging system consisting of 2 side channel blowers, protection cover casing, electrical and control panel, remote start / stop /alarms panel, safety suction filter and depression control valve. The blowers are equipped with: vibration absorbing feet, nonreturn valve. * = tolleranza / tollerance / tolerance / tolerancia ±10% BGS GENERAL srl Via Zibido, 3 - 20080 Zibido San Giacomo (MI) +39 02.90005069 r.a. - [email protected] - www.bigiesse.it

 Open the catalog to page 1
CGA 130-2

CENTRALE DI EVACUAZIONE GAS ANESTETICI (AGSS) ANAESTHETIC GAS SCRAVENGING SYSTEM (AGSS) DIMENSIONI / DIMENSIONS OPZIONI / OPTIONS Alimentazione - Power supply 3~ multivolts 50/60 Hz Aspirazione Suction Filtro aspirazione Suction filter Pannello remoto supplementare Mandata Exhaust Valvola di non ritorno Non return valve Soffiante Blower Vuotostato Vacuum switch Valvola regolazione vuoto Vacuum regulator valve Quadro elett. Control panel Additional remote panel PANNELLO ALLARME REMOTO / REMOTE ALARM PANEL 155 mm 185 mm ALARM FOR ANAESTHETIC GAS EVACUATION SYSTEM ALLARME EVACUAZIONE GAS ANESTETICI...

 Open the catalog to page 2

Archived catalogs

  1. CGA 100

    2  Pages

  2. Medevice

    20  Pages

  3. CGA 30

    2  Pages

  4. 3PBO-M

    2  Pages

  5. CGA 70

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.