Maintenance of Turbines
1. Preparation
  • Removing the Instrument: Detach the rotary instrument from the chuck mechanism.
  • Pre-Disinfection: Perform external disinfection without immersing the instrument in a disinfectant bath.
2. Cleaning
  • Under Running Water: Clean under water at temperatures below 38°C.
  • Removing Dirt/Deposits: Use Bien-Air cleaning wire to remove dirt from spray nozzles, dry with compressed air, and wipe the instrument with a cloth. Avoid ultrasonic cleaners.
  • Internal and External Cleaning with Spraynet: Shake Spraynet, use appropriate attachments, and clean with the instrument tip pointing downwards. Protect with an absorbent cloth and press the Spraynet button for one second.
3. Disinfection
  • Surface Disinfectant: Follow manufacturer's instructions, do not immerse in disinfectant bath.
  • Thermal Disinfection: Follow manufacturer's instructions.
4. Lubrication
  • Lubrication with Lubrimed or Lubrifluid: Shake Lubrifluid, use appropriate attachments, and lubricate with the instrument tip pointing downwards. Conduct an operating test for 30 seconds at low speed to remove excess lubricant. Store instruments with the tip downwards.
5. Sterilisation
  • Preparation: Place the turbine in a sterilisation bag.
  • Process: Sterilise according to manufacturer's instructions. Only sterilise cleaned and lubricated instruments.
Maintenance Accessories
  • Various cleaning and lubrication products are available, including Aquacare, Spraynet, Lubrifluid, Lubrimed, and Bien-Air cleaning wire.
See more

Catalog excerpts

MAINTENANCE OF TURBINES.-1

Refer to the user manual for the product for the complete cleaning and sterilisation maintenance procedures. This medical device is supplied in "non sterile" condition. Please ensure that this procedure is followed before the device is used. • Remove the rotary instrument from the chuck mechanism B. PRE-DISINFECTION • External disinfection D o not immerse the instrument in a disinfectant bath. CLEANING A. CLEAN UNDER RUNNING WATER AT <38°C B. REMOVING DIRT/DEPOSITS • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth Do not immerse in an ultrasonic cleaner C . NTERNAL AND EXTERNAL CLEANING WITH I SPRAYNET • Shake the Spraynet before use • Use the Spraynet attachment ( for 2-/3-way and 4-way turbines, see inset below) • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second • Clean the exterior of the turbine with a cloth soaked in Spraynet C‘. LEANING THE INSIDE WITH SPRAYNET C FOR 2-/3-WAY AND 4-WAY TURBINES ONLY • Use the Spraynet attachment • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second S pray into the opening as shown in the illustrations for “2-/3-way” and “4-way” * All the references for the attachments are given on the last page

 Open the catalog to page 1
MAINTENANCE OF TURBINES.-2

DISINFECTION A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use) D o not immerse the instrument in a disinfectant bath OR B. THERMAL DISINFECTION (follow the manufacturer's instructions for use) A. LUBRICATION WITH LUBRIMED OR B. LUBRICATION WITH LUBRIFLUID • Shake the Lubrifluid before use • Use the Lubrifluid attachment 2 , 3 or 4 depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below) • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second • Then,...

 Open the catalog to page 2
MAINTENANCE OF TURBINES.-3

MAINTENANCE ACCESSORIES SPRAYNET 500 ml cleaning spray Pack of 6 LUBRIFLUID 500 ml lubricant Pack of 6 AQUACARE 500 ml cleaning spray Pack of 6 Bora L / LK, Boralina, Prestige L / LK, Tornado L / LK Bora L, Boralina, Prestige L, Tornado L, Boralina LUBRICARE Maintenance system LUBRIMED Pack of 6 cartridges BIEN-AIR CLEANING WIRE Pack of 10