Refer to the user manual for the product for the complete cleaning and sterilisation maintenance procedures. This medical device is supplied in "non sterile" condition. Please ensure that this procedure is followed before the device is used. • Remove the rotary instrument from the chuck mechanism B. PRE-DISINFECTION • External disinfection D o not immerse the instrument in a disinfectant bath. CLEANING A. CLEAN UNDER RUNNING WATER AT <38°C B. REMOVING DIRT/DEPOSITS • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth Do not immerse in an ultrasonic cleaner C . NTERNAL AND EXTERNAL CLEANING WITH I SPRAYNET • Shake the Spraynet before use • Use the Spraynet attachment ( for 2-/3-way and 4-way turbines, see inset below) • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second • Clean the exterior of the turbine with a cloth soaked in Spraynet C‘. LEANING THE INSIDE WITH SPRAYNET C FOR 2-/3-WAY AND 4-WAY TURBINES ONLY • Use the Spraynet attachment • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second S pray into the opening as shown in the illustrations for “2-/3-way” and “4-way” * All the references for the attachments are given on the last page
Open the catalog to page 1DISINFECTION A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use) D o not immerse the instrument in a disinfectant bath OR B. THERMAL DISINFECTION (follow the manufacturer's instructions for use) A. LUBRICATION WITH LUBRIMED OR B. LUBRICATION WITH LUBRIFLUID • Shake the Lubrifluid before use • Use the Lubrifluid attachment 2 , 3 or 4 depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below) • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second •...
Open the catalog to page 2MAINTENANCE ACCESSORIES SPRAYNET 500 ml cleaning spray Pack of 6 LUBRIFLUID 500 ml lubricant Pack of 6 AQUACARE 500 ml cleaning spray Pack of 6 Bora L / LK, Boralina, Prestige L / LK, Tornado L / LK Bora L, Boralina, Prestige L, Tornado L, Boralina LUBRICARE Maintenance system LUBRIMED Pack of 6 cartridges BIEN-AIR CLEANING WIRE Pack of 10
Open the catalog to page 3All Bien-Air Dental catalogs and technical brochures
-
Tornado
5 Pages
-
Lubricare_2
3 Pages
-
Bora2
5 Pages
-
Nova
7 Pages
-
CHIROPRO PLUS
11 Pages
Archived catalogs
-
MC2 + MC3 Micromotors
8 Pages
-
MCX MICROMOTOR
8 Pages
-
Optima
7 Pages
-
bien-air_tornado
13 Pages
-
Laboratory
2 Pages
-
SMALL BODY – BIG HEART
8 Pages
-
EVO 15
11 Pages
-
ENT CATALOGUE
13 Pages
-
ENT PRODUCTS OVERVIEW
2 Pages
-
WIRE DRIVER
2 Pages
-
STRAIGHT SHANK BURS
2 Pages
-
NEURO PRODUCTS OVERVIEW
2 Pages
-
NEURO CATALOGUE
7 Pages
-
MICROSAWS BLADES
2 Pages
-
Products oVErVIEW
1 Pages
-
CATALOGUE
22 Pages
-
Bien Air Surgery
6 Pages
-
CATALOQUE 2013-2014 BIEN-AIR DENTAL
180 Pages
-
MILL200
2 Pages
-
MILL300
2 Pages
-
ECONOMICAL MAINTENANCE
2 Pages
-
OSSEO DUO
8 Pages
-
SCAN200 / D710
2 Pages
-
CAD-CAM SOLUTIONS
16 Pages
-
MICROMOTORS
12 Pages
-
ELECTRONICS CONTROLS
12 Pages
-
Bora Turbines
8 Pages
-
PHYSIODISPENSERS
20 Pages
-
POWERED MICROSAW RHINOPLASTY
8 Pages
-
CATALOGUE 2011 BIEN-AIR DENTAL
172 Pages