spinit® HbA1c Quick Guide

spinit® HbA1c Quick Guide

spinit® HbA1c Quick Guide

Product catalog summary
Test Disc and Sample Collection:
The spinit® HbA1c test requires specific sample collection procedures. Use capillary blood directly from the finger, fresh venous whole-blood (EDTA), or capillary whole-blood of human origin. The spinit® HbA1c disc is ready-to-use and does not require any reagent preparation.

spinit® HbA1c Controls:
Controls should be stored refrigerated (2-8°C) and are stable for 50 days once opened. Allow control material to reach room temperature before use and mix thoroughly. Use Quality Control mode for testing, ensuring measured values fall within the acceptable range.

spinit® Instrument:
Switch on the instrument and wait for stabilization. Follow the user manual and instructions for use (IFU) for detailed guidance. Information codes indicate issues such as insufficient sample volume (code 114) or reagent volume (code 210), requiring a new sample or disc.

Test Kit Components:
Supplied items include spinit® HbA1c discs and strips. Not supplied items include gloves, disinfectant towel, lancet, and tissue paper. Store the test kit refrigerated until the expiry date.

Test Procedure:
1. Clean the sampling area and allow it to dry.
2. Remove a spinit® strip and close the vial immediately.
3. Prick the finger, wipe away the first blood drop, and collect the second drop with the strip.
4. Ensure the sample area is fully loaded; refill if necessary.
5. Insert the strip into the disc sample well, break the strip, and leave the sample inside.
6. Follow on-screen instructions to complete the test.
7. Discard used materials according to local regulations.

Additional Notes:
Do not reuse the disc. Ensure the disc is inserted into the instrument within 10 minutes of opening the pouch. For expected values, refer to the spinit® HbA1c test instructions for use.
See more

Catalog excerpts

spinit® HbA1c Quick Guide-1

Quick Guide TEST DISC : • • • • Keep the test kit refrigerated 2-8°C until expiry date. When performing a test, allow disc to reach room temperature. Only take the disc out of the pouch after reading the barcode. Once the pouch has been opened, the disc must be loaded and inserted into the spinit® instrument within 10 minutes. • Using capillary blood, the sample should be introduced within one minute into the disc after collection. SAMPLE COLLECTION: • Use capillary blood directly from finger, fresh venous whole-blood (EDTA) or capillary whole-blood of human origin. • spinit® HbA1c disc is ready-to-use and does not require any reagent preparation step. • spinit® strips: • The vial must be stored at room temperature; • Always close the vial immediately after removing a spinit® strip. Store the controls refrigerated (2-8°C). Opened controls are stable for 50 days. Allow the control material to reach room temperature before use. Mix the control material thoroughly by shaking the vial for 30 seconds. Use Quality Control mode for control testing. The measured value should be within the acceptable range stated in the spinit® HbA1c Controls IFU. spinit® INSTRUMENT: • Switch the spinit® instrument on by pressing the ON button. • Wait until spinit® stabilises; • Information codes for the spinit® HbA1c assay: INFO CODE Training materials: Consult spinit® instrument instructions: • spinit® user manual • spinit® HbA1c controls IFU • spinit® HbA1c IFU Not enough sample volume to be measured. Not enough reagent volume or previously used disc. Collect a new sample and repeat the test using a new disc. Necessary materials to perform the test: Supplied with the kit: • spinit® HbA1c discs • spinit® strips

 Open the catalog to page 1
spinit® HbA1c Quick Guide-2

How to perform a spinit® HbA1c test 1 • Remove disc from refrigerator and wait to reach room temperature unopened; • Press Start button and follow on-screen • Scan the barcode from the disc pouch; • Place the disc on a plain surface with label • Take the disc out of the pouch; • Press the reagent blister using your thumb; Enough sample and correctly filled; • Prepare the sampling area by cleaning the • • finger using a disinfectant and allow it to dry; Remove a spinit® strip out of the vial and close the vial immediately after; Prick the patients finger using a lancet and wipe away the first...

 Open the catalog to page 2

All Biosurfit, SA catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.