USER MANUAL COMPUTRAC® MAX® 4000/4000XL MOISTURE ANALYZER Firmware Revision 1.7X October 2019 AMETEK BROOKFIELD 3375 N Delaware Street | Chandler, AZ 85225 USA 800.528.7411 | 602.470.1414 | f 602.281.1745 www.azic.com Email: Sales – sales.computrac@ametek.com International – international.computrac@ametek.com Customer Service – customerservice.computrac@ametek.com

Open the catalog to page 1
Computrac® MAX® 4000XL - 2

Computrac® MAX® 4000 / 4000XL PROPRIETARY RIGHTS NOTICE This manual contains valuable information developed by AMETEK Brookfield for use with the Computrac® line of moisture analyzers. No part of this manual can be reproduced or transmitted in any form or by any means, electronic, mechanical, or otherwise. This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of AMETEK Brookfield. ALL RIGHTS RESERVED © Copyright 2008-2019 AMETEK Brookfield PLEASE NOTE: There are 2 different instrument models in the...

Open the catalog to page 2
Computrac® MAX® 4000XL - 3

AMETEK Brookfield’s Computrac® MAX® 4000 series moisture analyzers are advanced, highly customizable instruments. As a result, the manual for the MAX® 4000 is extensive, comprehensive and detailed. For best results, read this manual in its entirety and keep the manual available for reference during instrument operation. The manual is organized in distinct sections to facilitate easily locating the information needed when referring back to the manual. The sections and their descriptions are as follows: §2 TABLE OF CONTENTS starting on page 4 §3 WARNINGS AND SAFETY INFORMATION starting on...

Open the catalog to page 3
Computrac® MAX® 4000XL - 6

WARNINGS AND SAFETY INFORMATION The MAX® 4000 moisture analyzer complies with the Underwriter Laboratories Inc. standards and European Council Directives for Electrical Equipment for Laboratory Use, electromagnetic compatibility, and the stipulated safety requirements. The MAX® 4000 and its accessories pass Class A emissions. However, improper use or handling of the instrument can result in damage to the equipment and/or injury to personnel. The MAX® 4000 is Equipment Class I, Pollution Degree 2, and Installation Category II. The Temperature Calibration Interface (TCI) is ESD sensitive....

Open the catalog to page 6
Computrac® MAX® 4000XL - 7

LITHIUM COIN CELL BATTERY The MAX® 4000 contains a lithium battery to keep power on the clock when the instrument is turned off. CAUTION – The battery may explode if mistreated. Do not recharge, disassemble, or dispose of in fire. PREVENT EXCESS HEAT BUILD-UP AROUND THE INSTRUMENT • When setting up the instrument, leave enough space to prevent heat from building up and to keep your instrument from overheating. Leave 20 cm (8 inches) around the instrument and 1 m (3 feet) above the instrument. • Do not put any flammable substances on, under or near the instrument, as the area around the...

Open the catalog to page 7
Computrac® MAX® 4000XL - 8

ENVIRONMENTAL CONDITIONS • STORAGE AND SHIPPING o Temperature should be between 0 °C and +40 °C (+32 °F and +104 °F). o Relative Humidity should be between 10% and 80%. • OPERATIONAL CONDITIONS – INDOOR USE. o Optimum results will be achieved when the instrument is set on a smooth, level, non-vibrating surface in a non-condensing, non-explosive environment of 0 - 40 ºC at 50% relative humidity or 0 - 30 ºC at 80% relative humidity. o Do not expose the instrument unnecessarily to extreme temperatures, moisture, shocks, blows or vibration. WARNING SYMBOLS Protective Ground: Identifies the...

Open the catalog to page 8
Computrac® MAX® 4000XL - 9

PACKING AND SHIPPING INSTRUCTIONS IMPORTANT Movement, handling, and packaging of the instrument must be done with EXTREME CARE to avoid permanent, expensive internal damage. The MAX® 4000 uses a delicate and very sensitive electronic force balance to measure small weights and weight losses. The balance is capable of measuring weight to a resolution of 0.1 milligrams. Careful handling is necessary to prevent damage to the balance. With proper care, the instrument will provide years of dependable service. Unpack the instrument carefully and set it down gently. Select a place where it will be...

Open the catalog to page 9
Computrac® MAX® 4000XL - 10

Call AMETEK Brookfield Customer Service for Return Material Authorization (RMA) information prior to returning an instrument. Boxes and packing materials are available from AMETEK Brookfield for all shipments. Pack the MAX® 4000 in a Computrac® MAX® 4000 shipping container. Use the packing diagram below that matches your shipping system as a guide. If using the foam on the right below, orient the bottom foam tray as shown. A packing guidance diagram is also included on one of the box flaps. Be sure to place the foam block underneath the pan support when packing the instrument. See UNPACKING...

Open the catalog to page 10
Computrac® MAX® 4000XL - 11

The Computrac® MAX® 4000 and MAX® 4000XL are high performance programmable loss-on-drying analyzers with the following features: • Typically, the instrument results are precise to better than: 0.002% SD typical for < 0.020% moisture (MAX® 4000XL only) 10% RSD from 0.020% - 0.100% moisture (MAX® 4000XL only) 2.0% RSD from 0.100% - 100% moisture (MAX® 4000 and MAX® 4000XL) Programming is through the front panel keypad with a user friendly menu system. A color LCD that is easy to read and provides color graphing. AMETEK Brookfield's pioneering prediction method that can automatically end a...

Open the catalog to page 11
Computrac® MAX® 4000XL - 12

A User Login function that, if activated, can be used to restrict access to calibration functions and instrument and test configuration options except by authorized users, preventing unwanted changes to operating parameters. When User Login is active on the MAX® 4000XL, a user is required to login prior to testing, and the user name is recorded with each test result. On 21 CFR, Part 11 compliant instruments, the Delete Test Results and Delete Audit Log Events functions are disabled throughout the system. For those familiar with moisture analysis instruments, the MAX® 4000 should be simple...

Open the catalog to page 12
Computrac® MAX® 4000XL - 13

Instrument Models There are 2 different instrument models in the MAX® 4000 series, the MAX® 4000 and the MAX® 4000XL, which have various feature differences, as detailed below. Throughout this manual, MAX® 4000 is used to refer to both the MAX® 4000 and the MAX® 4000XL when describing common features or functionality. When there are specific differences between the two models, these are indicated through the usage of MAX® 4000XL for features that are only available in the MAX® 4000XL. Refer to the label on the front of the instrument, the splash screen displayed during power-up or access...

Open the catalog to page 13