JEROME® J405

Product catalog summary
Introduction
The Jerome® J405 Mercury Vapor Analyzer is designed for detecting mercury vapor in ambient air, with a detection range of 0.5 to 999 µg/m3. It is intended solely for gaseous vapor use and should not be exposed to liquids or dust. The device is user-friendly, requires minimal maintenance, and includes features such as a backlit display, auto zero bridge adjustment, and storage for 20,000 samples.
Unpacking and Setup
Upon receipt, ensure all components are present and undamaged. Retain packaging for future use. The device is ready for use but requires sensor regeneration before initial operation. It can operate on internal battery power or an external AC power supply.
Technical Specifications
The J405 features updated electronics with 24-bit a/d conversion and can operate for 24 hours on its internal battery. It offers multiple measurement units and includes diagnostic and error detection capabilities. The device measures mercury vapor concentrations with a display range of 0.5 μg/m3 to 999 μg/m3, a resolution of 0.01 μg/m3 Hg, and a response time of 16 seconds in sample mode.
Operation
The J405 features a user-friendly interface with options for sampling, regeneration, data management, and system settings. Daily operations include sensor regeneration and sampling, with specific procedures for different power sources.
Maintenance
Regular maintenance involves preventive checks, filter replacements, and battery management. A detailed schedule and instructions are provided in the manual.
Calibration and Troubleshooting
Calibration services are available from the factory, and functionality verification is recommended. The troubleshooting section addresses potential interferences and error messages.
Accessories and Parts
A range of accessories and maintenance parts are available, including filters and battery packs.
Applications
The J405 is suitable for regulatory compliance, ambient air analysis, quality control, and portable mercury vapor detection.
Warranty
The manual includes warranty information and customer service contact details.
Instrument Operation
The main screen displays system time, battery status, and mercury concentration. The device can operate in manual or auto-sampling modes, with intervals from 1 to 120 minutes. USB ports allow data transfer and menu navigation.
Sample Menu
Users can select between manual and auto-sampling modes, set sample intervals, and choose measurement units. Continuous sampling is not recommended as it may reduce sensor life.
Regen Menu
The sensor saturation meter indicates when regeneration is needed. The regeneration process requires external power and involves a 45-minute cycle.
Warmup Procedure
For accuracy at cleanup levels below 3 µg/m3, the J405 must be warmed up with the Zero Air Filter in the intake at the sampling location.
Set Interval and Hour
The J405 can be set to perform sensor regenerations at intervals of 6, 12, 24, or 48 hours, or when the sensor is saturated.
Data Management
The J405 can store up to 20,000 data points. Data can be retrieved via USB, saved to a USB flash drive, or printed directly.
System Menu
The SYSTEM menu is password-protected. Users can set alarm levels, toggle audible alarms, and choose display modes.
Daily Operations
Before daily use, verify the instrument's operation by checking the display, battery level, and performing a sensor regeneration.
Sensor Regeneration
Sensor regeneration is necessary to clean the sensor of accumulated mercury vapor. It should be performed at the start and end of each day.
Troubleshooting and Maintenance
Common issues such as power problems and battery failures are addressed with solutions like recharging or replacing the battery. Display warnings and error messages guide users on corrective actions.
Functional Test Kit Overview
The J405 Functional Test Kit (FTK) is used for verifying the functionality of the instrument. It is important to handle the mercury vial and thermometer with care due to the risk of mercury contamination.
See more

Catalog excerpts

JEROME® J405-1

USER MANUAL MERCURY VAPOR ANALYZER ARIZONA INSTRUMENT LLC 3375 N Delaware Street | Chandler, AZ 85225 USA 800.528.7411 | 602.470.1414 | f 602.281.1745 www.azic.com Email: General – [email protected] International – [email protected] Customer Service – [email protected]

 Open the catalog to page 1
JEROME® J405-2

JEROME® J405 Mercury Vapor Analyzer Operation Manual PROPRIETARY RIGHTS NOTICE This manual contains valuable information and material developed by Arizona Instrument LLC for use with the Jerome® J405 Mercury Vapor Analyzer. No part of this manual can be reproduced or transmitted in any form or by any means, electronic, mechanical or otherwise. This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of Arizona Instrument LLC. ALL RIGHTS RESERVED  Copyright 2006-2017 Arizona Instrument LLC Resisorb® is a...

 Open the catalog to page 2
JEROME® J405-4

All section numbers, titles and page numbers in BOLD in this manual are hyperlinks and can be clicked to simplify navigation within the PDF version of the manual. The BACK button found in most PDF software programs is also very helpful when navigating using the hyperlinks. An up-to-date electronic copy of this manual can be found at: http://www.azic.com/downloads.aspx NOTE: For EPA clean-up levels less than 3 μg/m3, it is necessary to run a warm-up routine before sampling. To initiate the automatic five minute warm-up, install a Zero Air Filter (AZI P/N Z2600 3905) in the intake and select Warmup...

 Open the catalog to page 4
JEROME® J405-5

This manual contains details that will optimize the results and the life of your instrument. Read and refer to the manual for complete details on operation, maintenance and troubleshooting, special voltage inputs and data output. The Jerome® J405 is easy to operate and ready for use upon receipt from the factory. • Remove the instrument from the packing material. Retain all packaging materials for any future shipment of the instrument. If the instrument is returned to AZI for any reason, it must be placed in the original packaging materials that have been tested and proven to be effective protection...

 Open the catalog to page 5
JEROME® J405-6

Perform a sensor regeneration by following these steps:  Connect the AC power supply/charger between the matching (DIN) connector on the back of the J405 and an AC power outlet or connect the external battery pack to the back of the J405.  If the J405 is turned off, press the I/O power button to turn the instrument on.  Press the RIGHT arrow button (►) to enter the main menu.  Press the DOWN arrow button (▼) to move the cursor to Regen.  Press the RIGHT arrow button (►) to select Regen from the menu.  Press the ENTER/START button on the keypad to select Regen Now from the Regen menu. ...

 Open the catalog to page 6
JEROME® J405-7

The Jerome® J405 Mercury Vapor Analyzer is an ambient air analyzer with a range of 0.5 to 999 micrograms of mercury vapor per cubic meter (µg/m3 Hg). CAUTION: The Jerome® J405 is for gaseous vapor use only. DO NOT allow the probe or the instrument's intake to be exposed to any liquid, dust or other foreign material. The J405 is designed to be easy to operate for quick and accurate analysis of mercury vapor levels. It has few maintenance requirements. However, please take a moment to read this manual before attempting operation. If you have any questions about your application or operation, please...

 Open the catalog to page 7
JEROME® J405-8

Regulatory detection compliance Regulatory clean up compliance Ambient air analysis Quality control Scrubber efficiency testing Accuracy check for other mercury vapor monitors and control systems Mercury vapor source detection Leak detection Portable mercury vapor detection. (The Jerome® J405 is operated from an internal 12V battery source. For charging the batteries, an external AC power supply/charger is included.)

 Open the catalog to page 8
JEROME® J405-9

JEROME® J405 TECHNICAL SPECIFICATIONS Units: µg/m3 Display Range* Resolution Response time-sample mode Response time-survey mode Flow rate Power requirements Fuse Internal battery pack Operating environment Case construction Dimensions Weight Digital meter display Data storage capacity Output Certifications 16 seconds 2 seconds 750 ± 50 cc/min (0.75 ± 0.05 liters/min) 12 VDC for the instrument (provided by the internal battery, AC power supply/charger, external battery pack or car accessory adapter) 100-240VAC, 47-63Hz, 3.2A for the AC power supply/charger Auto-resetting fuse Rechargeable nickel...

 Open the catalog to page 9
JEROME® J405-10

The heart of the J405 is the well-proven gold film sensing technology. A thin gold film, in the presence of mercury vapor, undergoes an increase in electrical resistance proportional to the mass of mercury vapor in the sample. When the SAMPLE button is pressed, an internal pump pulls ambient air through a scrubber filter (AZI P/N: 600-0262) and into the flow system. After three seconds, the sample solenoid bypass opens, closing off the scrubber filter from the flow system. The sample air passes through the AG filter (AZI P/N: 600-0293), an acidic gas filter that removes any acidic gases that...

 Open the catalog to page 10
JEROME® J405-11

The instrument’s microprocessor automatically re-zeroes the digital meter at the start of each sample cycle and holds the meter reading until the next sample cycle is activated, thus eliminating drift between samples. During the sample mode cycle, a sensor saturation meter on the LCD represents the percentage of sensor saturation, or adsorbed mercury collected on the gold film. With use, the sensor becomes saturated and needs to be cleaned. This is accomplished by a manually activated 45-minute sensor regeneration cycle, which removes the mercury from the sensor. The solenoid bypass closes during...

 Open the catalog to page 11
JEROME® J405-12

INSTRUMENT OPERATION J405 Main Screen Display The J405 Main Screen is depicted below, followed by explanations of the labeled items. SYSTEM TIME – As set in the SYSTEM menu. (See SYSTEM menu on page 20). BATTERY – Indicates current charge level and charging status (See Charging Internal Battery on page 28) SOUND ICON – Indicates if the audible alarm is turned on or muted, as set in the SYSTEM menu. (See SYSTEM menu on page 20). SYSTEM DATE – As set in the SYSTEM menu. (See SYSTEM menu on page 20). SENSOR SATURATION METER – Graphically indicates mercury saturation level of J405 sensor. (See REGEN...

 Open the catalog to page 12