Overview: The 6540 Series Photostability Chambers are designed for near UV and visual light testing according to ICH Q1B Guidelines, Option 2. These chambers feature a user-friendly color touch-screen interface for controlling light and temperature conditions, with automatic light shut-off after reaching specified exposure levels.
Standard Features:
  • Independent control of near UV and cool white VIS lamps to prevent overexposure and allow flexibility in testing.
  • Automatic and manual modes for test duration control.
  • Dimmable lamps for specific light intensity testing (230V models).
  • Programmable alarms for lamp replacement notifications.
  • Specular aluminum reflective surfaces for uniform light distribution.
  • Stainless steel interior for easy maintenance.
  • Two light detectors with integral radiometer for precise light intensity measurements, calibrated to NIST standards.
  • Precise temperature control via PID controller.
  • Humidity control in specific models to isolate hydration effects during testing.
Main Screen Display: The interface includes an icon-based, multi-language touch-screen for easy programming of temperature, humidity, and light meters. It also features pull-out light banks for easy bulb replacement and a modular refrigeration system for easy servicing.
Serviceability: The chambers are designed with off-the-shelf components for easy maintenance and offer factory services such as validation, calibration, and preventative maintenance.
Specifications:
  • UVA and VIS lamp types are fluorescent, with typical UVA light intensity ranging from 2-30 W/m2 and VIS light intensity from 2-35 klux.
  • Temperature range is 10°C to 35°C with lights on, and humidity range is 40-70% RH for specific models.
  • Interior construction includes reflective specular aluminum and stainless steel, with a total internal volume of 10.5 cu. ft.
  • Electrical requirements vary by model, with options for 115V, 208/230V, and 230V configurations.
Optional Accessories: Various replacement kits for light banks and chart recorders are available, along with a universal condensate recirculating system and extended temperature options.
See more

Catalog excerpts

Photostability chambers-1

Photostability chambers 6540 Series, Designed per ich q1b

 Open the catalog to page 1
Photostability chambers-2

PHOTOSTABILITY CHAMBERS The 6540 Series Photostability Chambers were specifically designed to perform near UV and visual light testing with fluorescent lamps per ICH Q1B Guidelines, Option 2. This second generation bench-top unit controls light and temperature conditions through an easy-to-use color touch-screen interface. Lights automatically shut off after a specified exposure level or time duration. Uniform light distribution and high intensity levels allow quick response for forced degradation testing and confirmatory studies. Precise humidity control (Models 6545-1, 6545-2 and 6545-3) is...

 Open the catalog to page 2
Photostability chambers-3

STANDARD FEATURES CONTINUED • Shelves slide-out for quick and easy loading and unloading of samples. • Strategic lamp spacing provides enhanced light uniformity. Overhead lighting results in a large illumination area for maximized direct lighting. • Door safety switch turns lights off when the door is opened. The lockable door prevents accidental disturbance of in-process tests. • Two access ports (one per shelf) are standard. MAIN SCREEN DISPLAY • Performance is managed through an eye level, icon-based, color, touch-screen interface for user convenience. • The easy-to-use controls Active display...

 Open the catalog to page 3
Photostability chambers-4

specifications & optional accessories Model Lamp Type, UVA Lamp Type, VIS Light Detectors UVA Light Intensity, Typical VIS Light Intensity, Typical Lamps Dimming Typical Time Required to Reach ICH Recommendation of 200 W-hr/m2 Typical Time Required to Reach ICH Recommendation of 1.2 million lux-hr Light Uniformity Temperature Range Temperature Control Temperature Uniformity Humidity Range Humidity Control # of Shelves, Size Usable Shelf Area, Each Usable Shelf Area, Total Maximum Weight per Shelf Interior Surfaces Interior Construction Internal Volume Exterior Dimensions Interior Dimensions #...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.