Basic Rules

Basic Rules
1 / 11 PagesView full catalog

Basic Rules

Product catalog summary
Generalities
  • Conservative basic rules and patient medical history should be observed.
  • Consider periodontology and overall situation before implant-supported restoration.
  • Recommendations for implants: 6-8 in the upper jaw and 4-6 in the lower jaw depending on denture type.
  • Ensure tension-free fit of the superstructure for implant-supported dentures.
  • For single molar restoration, a minimum of one ø 4.0 mm implant is recommended.
Specifics
  • Champions® implants can be inserted minimally invasively or with a full-flap method.
  • Primary stability of 20 to 40 Ncm should be achieved in D1 and D2 bone.
  • Torque wrench settings should start at 20 Ncm and can be adjusted up to 40 Ncm.
Immediate Restoration and Immediate Loading
  • Immediate loading is possible with a passive-fit denture supported by at least 4 implants.
  • Primary stability of 20 to 40 Ncm is required.
  • Use only designated original Champions® instruments for implant placement.
Condensers / Implant Diameter Choice in D3/D4-bone
  • Condensers can transform soft bone into hard bone, aiding in immediate implantation.
  • Implant diameter is determined by the Condenser that achieves primary stability.
  • Follow a sequence of increasing Condenser diameters to achieve desired stability.
Drilling Sequences
  • Start pilot drilling 0.5 to 1 mm palatal or lingual to the alveolar ridge center.
  • Drilling speeds: max 250 rpm in cortical bone, 70 rpm in cancellous bone.
  • Ensure primary stability by providing crestal bone relief for titanium implants.
  • Perform Bone Cavity Check (BCC) in all five dimensions after drilling.
Condenser Usage
  • Condensers are used for bone condensing and determining optimal implant diameter.
  • Use only in soft to medium-hard bone (D3 + D4).
  • Manual and mechanical operations have specific torque limits to prevent damage.
See more

Catalog excerpts

Basic Rules-1

PRODUCT INFORMATION Basic Rules & Drilling and Condenser Protocol

 Open the catalog to page 1
Basic Rules-2

3. Immediate Restoration and Immediate Loading 4. Condensers/Implant diameter choice in the D3/D4-bone 1) Please observe conservative basic rules and individual medical history of the patient. 2) Before an implant-supported and prosthetic restoration, among others periodontology and the overall situation should be considered and assessed. 3) Please follow the recommendations of the "Konsenskonferenz Implantologie" (consensus conference Implantology): In the upper jaw, a minimum of 6 implants /teeth is recommended for a removable denture and 8 implants /teeth for a fixed denture. In the lower...

 Open the catalog to page 2
Basic Rules-3

1) All Champions® implants can be inserted both using the minimally invasive method (without raising a mucoperiosteal flap, i.e. in a “flapless” way) and in a “classic” manner (with “full-flap”). They are particularly well suited for insertion following the secured MIMI® technique. An incision made with a scalpel or diamond turbine without raising a flap is also considered minimally invasive, as is a punch or direct penetration of the gingiva. 2) With Champions (R)Evolution® implants, the Shuttle (or the Gingiva-Clix to be clipped over it) should ideally terminate in a subgingival or equigingival...

 Open the catalog to page 3
Basic Rules-4

3.Immediate Restoration and Immediate Loading Immediate loading is usually possible when ... ... a passive-fit denture is supported by a minimum of 4 implants /teeth. ... primary stability of 20 to 40 Ncm is achieved. ... w ith fewer than 4 implants /teeth (including single-tooth gaps), transitioning to the Secondary Osseointegration Stability phase (SOS) is recommended. The single crown should be integrated 8 weeks post-surgery! Please note: Champions® implants must only be placed and treated with the designated original Champions® instruments such as Drills, Condensers, Insertion Aids, and...

 Open the catalog to page 4
Basic Rules-5

4.Condensers / Implant diameter choice in the D3/D4-bone If the bone at the surgical site is soft (D3/D4), it is possible to transform soft bone into hard bone using Condensers in a minimally invasive way, which means to increase the D3/D4 bone densification, thus preparing it, for example, for immediate implantation. This process is known as “Osseous Metamorphosis.” Additionally, Condensers assist implantologists in selecting the appropriate implant diameter because it’s not the anatomy that determines the implant diameter, but rather the diameter of the corresponding Condenser that achieves...

 Open the catalog to page 5
Basic Rules-6

1) Whenever possible – both in the maxilla and mandible – start the pilot drilling 0.5 to 1 mm in a palatal or lingual position to the center of the alveolar ridge. 2) After anesthesia in the MIMI technique, measure the mucosal thickness with the yellow Drill (laser markings start at 2 mm, then 4 and 6 mm, and continue at 10, 12, 14 mm, etc.) until the instrument makes contact with bone. 3) rilling should be done at a maximum of 250 rpm in the cortical bone and at D 70 rpm in the cancellous bone. 4) To ensure primary stability in the cancellous bone of the D1 and D2 bone, it is generally recommended...

 Open the catalog to page 6
Basic Rules-7

Condensers Generalities: Condensers are used for condensing bone and as test instruments for determining the optimal implant diameter. Use them only in soft to medium-hard bone (D3 + D4). Champions® Condensers should not be used in hard bone (D1/D2). It is strongly recommended that you should use the Condensers according to the Condenser- Protocol. Failure to observe the maximum torque (40 Ncm) can cause damage to the handpiece. View of Condensers with Torque Wrench holder View of Condensers with Handpiece (HP)-Connection Before surgery: Care should be taken to ensure that the components, instruments,...

 Open the catalog to page 7
Basic Rules-8

D1 + D2, ø 3.5 mm Champions (R)Evolution® & Square “NEW ART” • Pilot drilling: yellow – full working length according to implant length. • Preparation: white, blue • Crestal relief: brown, orange, green • Relief: brown, orange, green – not deeper than 2 mm subcrestally • Implant insertion D1 + D2, ø 4.0 mm Champions (R)Evolution® & Square “NEW ART” • Pilot drilling: yellow – full working length according to implant length. • Preparation: white, blue, brown • Crestal relief: orange, green, red • Relief: orange, green, red – not deeper than 2 mm subcrestally • Implant insertion D3, ø 4.5 mm Champions...

 Open the catalog to page 8
Basic Rules-9

D3, ø 5.0 mm Champions (R)Evolution® & Square “NEW ART” • Pilot drilling: yellow – full working length according to implant length. • Condensers: silver, light blue, green, yellow, blue, pink • Implant insertion D4, ø 5.5 mm Champions (R)Evolution® & Square “NEW ART” • Pilot drilling: yellow – full working length according to implant length. • Condensers: silver, light blue, green, yellow, purple, blue, pink, green • Implant insertion D1 + D2, ø 2.5 mm Ball-Head • Pilot drilling: yellow – full working length according to implant length. • Condenser: silver • Implant insertion Product Information...

 Open the catalog to page 9
Basic Rules-10

• Pilot drilling: yellow - full working length according to implant length. • Preparation: white • Implant insertion • Pilot drilling: yellow - full working length according to implant length. • Condensers: silver, light blue, green, yellow, purple • Implant insertion • Pilot drilling: yellow - full working length according to implant length. • Preparation: white • Implant insertion Product Information Basic Rules Rev. 05/2025-10 Page 10/11

 Open the catalog to page 10
Basic Rules-11

Manufacturer in the EU: Champions-Implants GmbH Managing Director: Dr. med. dent. Armin Nedjat Im Baumfeld 30 | Champions Platz 1 D-55237 Flonheim Germany Tel. +49 (0) 6734 91 40 80 | Fax +49 (0) 6734 10 53 [email protected] champions-implants.com CE 0297 Champions® is a registered trademark of Champions-Implants GmbH Product Information Basic Rules

 Open the catalog to page 11

All Champions Implants GmbH catalogs and technical brochures

  1. CERTIFICATE

    1  Page

  2. Mucosal Punch

    3  Pages

  3. Rescue Tool

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.