Product catalog summary
Overview: The document provides technical specifications and operational guidelines for AirSep NewLife Elite and Intensity stationary oxygen concentrators. These devices are designed for medical use, providing oxygen at various flow rates and pressures.
Specifications:
  • Flow Rates: Elite: 1.0-5.0 LPM; Intensity: 2.0-8.0 LPM; Intensity 10: 2.0-10.0 LPM.
  • Weight: Elite and Intensity: 54 lb (24.5 kg); Intensity 10: 58 lb (26.3 kg).
  • Sound Level: Elite: ±48 dB(A); Intensity: ±52 dB(A); Intensity 10: ±55 dB(A).
  • Power Consumption: Elite: 350 Watts; Intensity: 410 Watts; Intensity 10: 590 Watts.
  • Outlet Pressure: Elite: 7 psi (48 kPa); Intensity/Intensity 10: 20 psi (138 kPa).
Operational Guidelines:
  • ON/OFF Operation: The unit is turned on by pressing the power switch upward and off by pressing it downward. An audible alarm sounds briefly at startup.
  • Nasal Cannula and Tubing: Elite supports up to 50 ft (15.2 m) of tubing; Intensity/Intensity 10 supports up to 200 ft (61 m).
  • Humidifier Bottle: Compatible with all models at any flow setting.
Maintenance:
  • Air Intake Filter: Clean weekly with soap and water.
  • Product Filter: Designed to last the life of the unit, no replacement schedule.
  • Preventative Maintenance: Replace the felt filter every 5,000 hours if present.
Performance:
  • Oxygen Concentration: Varies by model and flow rate, generally between 90% and 95%.
  • Hour Meter: Tracks total and accumulated hours, with reset functionality for accumulated hours.
Alarms: Various audible and visual indicators for conditions such as low/high pressure, low O2 concentration, high temperature, and power failure.
Recommendations: Regular testing of oxygen concentration is advised upon delivery and at intervals set by the provider.
See more

Catalog excerpts

AirSep® NewLife® Elite and Intensity-1

Quick Reference Guide AirSep® NewLife® Elite and Intensity Stationary Oxygen Concentrators Flow Rates Elite: Intensity: Intensity 10: Elite: Intensity: Intensity 10: Sound Level Elite: Intensity: Intensity 10: Power Consumption Elite: Intensity: Intensity 10: 350 Watts 410 Watts 590 Watts Outlet Pressure Because of its high outlet pressure, the NewLife Intensity and Intensity 10 can be used for special applications including (but not limited to): large volume jet nebulizers, venti-masks, medication nebulizers, and the Y-connection of two concentrators to achieve high flow rates. With the unit plugged in, press the power switch upward in the ON (I) position. The audible alarm will sound briefly and oxygen flow will begin. If the unit has O2 Monitoring, the yellow light may remain on during warm-up for up to 5 minutes after power up. To turn the unit off, press the power switch downward in the OFF (0) position. All lights will turn off and oxygen flow will cease. Nasal Cannula and Tubing Elite: Standard oxygen tubing and nasal cannulas may be used with the Elite. Maximum tubing length is 50 ft (15.2 m). Intensity/Intensity 10: Standard or high flow oxygen tubing and nasal cannulas may be used with the NewLife Intensity and Intensity 10. The maximum tubing length is 200 ft (61 m). Humidifier Bottle A humidifier bottle may be used with the NewLife Elite, Intensity, and Intensity 10 at all flow settings. It attaches directly to the oxygen outlet of the conce

 Open the catalog to page 1
AirSep® NewLife® Elite and Intensity-2

The Air Intake Filter (Gross Particle Filter) is located on the back of all NewLife units and should be cleaned weekly by the patient. It is removed from the concentrator and washed with a warm solution of soap and water. Make sure the filter is rinsed and dried thoroughly before replacing it in the NewLife. Product Filter Elite: The product filter is a disk shaped filter prior to the outlet flow. This filter is designed to last the life of the unit and does not have a replacement schedule. Intensity: The product filter, which is located within the unit’s product tank, is designed to last the...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.