SOPRANOICE LASERSYSTEM Diode Laser System Model No. :3-Waves Soprano Ice Laser USER MANUAL
Open the catalog to page 1Brief Introduction 1. Treatment principle 755/808/10464nm diode laser hair removal technology is based on the selective absorption of light and heat. The laser goes through the epidermis layer to reach the root of hair follicles, laser can be absorbed and converted into heat to destroy the hair follicle and the surrounding oxygen tissue; Meanwhile, the particular super cooling system can creat certain protect the epidermis layer from burn by laser energy, which ensure At the same time, 755/808/10464nm diode laser heat hair follicles to 48°C under lower energy density mode, hair follicle cells...
Open the catalog to page 2SOPRANOICE LASER SYSTEM The operation and control parts are in the mainframe, the built-in integrated micro controller can monitor the operation of equipment continually. The control panel include: touch LCD screen (show the system working mode and parameters), key switch,emergency switch. The treatment handpiece include: laser emitter and transmission system, handheld operation by user. 755/808/1064nm laser treatment is a safe and reliable way to remove unwanted hair. As long as the trained and qualified medical personnel operate and maintain properly, it can make sure the safety operation....
Open the catalog to page 3SOPRANOICE LASER SYSTEM Using sealed white gem crystal to transmit l laser to the patient's skin, the light pulse can only be emit from the front surface of the gem crystal. • Shut off power by emergency switch when accident happens. • Keys witch is top reventabuse of the machine. • The buzzer beeps when the device is ready toemit light pulse. Emergency Button 12. 1inch Color Touch Screen Handpiece(15*20mm) Light Indicator Key Switch Handpiece Connector
Open the catalog to page 4SOPRANOICE LASER SYSTEM Machine Label Air Switch Power Code Foot Switch Air outlet Water input&output
Open the catalog to page 5SOPRANOICE LASER SYSTEM Perfect Custormsizsed Holder Self-locking Connector Largest Copper Cooler Medical Standard Wheels Brushless Pump
Open the catalog to page 6SOPRANOICE LASER SYSTEM SOPRANOICE LASER SYSTEM Soprano Ice PLATINUM Price is Much More Acceptable Power is Stronger USA COHERENT LASER BARS Can Treat More Than 10000+ Thin screen More Stylish designed Looks Modern Quality Finished Well-defined Structure Easy Maintenance
Open the catalog to page 7SOPRANOICE LASER SYSTEM The lightest weight handle 900g Only Easy Operated by Lady Pretty Cold Cooling Effect REAL PAIN FREE
Open the catalog to page 8SOPRANOICE LASER SYSTEM Screen Interface: 6 、 Cautions 6.1Safety prevention To ensure the safe operation, pls pay attention to safety prevention and read the warning below carefully: ① Except the company's authorized technical staff, other person is forbidden to examine and repair the laser machine, especially the internal machine, including power supply, cooling systems,optical componentsand diode laser handpiece ② Make sure rated voltage conform to the voltag ein your country(200/208Vor 220/230V) ③ Maintenance only after the machine is turned down and power cord is unplugged, to avoid Danger...
Open the catalog to page 9SOPRANOICE LASER SYSTEM ④ The gemcrystal should always bekept clean.It should be cleaned after each treatment ⑤ It is banned to start machine if laser hand piece leak water; if already start, it must be shut Down immediately. Diode laser may hurt the eyes or cause flame and burns. Necessary protective measures Should be taken in operation. ② Excessively emit laser light to treatment area may damage the skin, cause proliferation / atrophy/ abnormal pigmentation. ③ Do not look directly at the lasere mittedf rom the handpiece. Do not air-launch laser light, When not in use, the handpiece should...
Open the catalog to page 10SOPRANOICE LASER SYSTEM ① Temperature rise after the object absorb light energy,it is necessary to take ② Preventive measures to reduce danger of igniting the near in flammable. ③ Implementation of the preoperative skin treatment, please do not use alcohol or acetone and Other combustibles.If necessary, use soap and water. ④ If use soap and water to clean or disinfect the machine, only after it dry completely can the machine restarts. 6.5 System safety device The machine is equipped with some system safety devices, all operating or related personnel Must be familiar with their location and usage:...
Open the catalog to page 11SOPRANOICE LASER SYSTEM 8 、 System Instruction This chapter introduces the main component parts,control system and technology parameters. Notices: Only authorized service engineers can open the side panel, and maintain internal each module. User banned to disassemble the machine. Control panel: User perate the machine though the contro lpanel,it includes the following components: A. Key switch Used to power on and off, the machine can be open on by key provided by our ompany B. Emergency Stop It is used to in emergency. When depress it, the handle can not be lightened up and no light can Be simmered...
Open the catalog to page 12SOPRANOICE LASER SYSTEM Only authorized technician can onduct installation^ include: • Unpacking,place the machine in preselected location. • Check the integrity of equipment and itscomponents. • Infuse deionized water. • Installing the treatment head piece • Test the Output accuracy, calibration parameters,and check the function of all the parts
Open the catalog to page 13SOPRANOICE LASER SYSTEM ① Before unpacking, it is necessary to check whether the installation location meets the requirements. ② The yellow / green ground wire inside the sheath ground though wall line to make sure the Equipment ground.Good grounding is essential for safety. ③ No transient input power, voltage and current spikes, collapse or surge. Therefore, the machine should not share transmission line with lift, air-conditioning systems, high-power Motors and other large loads. We suggest separating the machine's power cord and separate air-break switch. ④ The main fuse is 16A circuit breaker...
Open the catalog to page 14SOPRANOICE LASER SYSTEM Turn on the key switch to start operation system, the interface displays as follows few second later. ① laser power,percentage adjustable; ② Pulse width( 1-400ms),adjustable; ③ Repetitionrate(1-10Hz); ④ Fluence(energy density) automatic display by different parameters; ⑤ laser head life monitor: total10000*1000shots;now:x*1000shots; ⑥ Treatment assistant: Set totals hots energy for treatment area; Target energy,adjustable; Current energy;reset; ⑦ Skin cooling level,adjustable; ⑧ Status(working,or error) ⑨ Ready:turn on the handle(ready to simmer) ⑩ Standby:turn off the...