Conformance Statement Overview
The document outlines the Clarius Mobile Health Corp.'s "Scanner" device, which uses ultrasound technology to produce medical images. It supports DICOM standards for communication and management of medical imaging data, facilitating image transfer to DICOM servers and automatic patient information filling via DICOM Worklists.
Introduction
This document is intended for healthcare professionals, software engineers, and designers familiar with DICOM. It details the Clarius DICOM implementation, supporting various data compression methods like Bitmap (uncompressed), Lossless JPEG, and JPEG-LS Lossless to optimize image storage and transfer.
Networking
  • Implementation Model: Describes data flow for storage, worklist, and verification processes. The Storage Application Entity sends images to remote entities, while the Modality Worklist Application Entity downloads worklists from DICOM servers.
  • Function Definition of AE: Details operations of Storage, Modality Worklist, and Verification Application Entities, including connectivity verification and image transfer protocols.
  • AE Specifications: Lists SOP Classes for various application entities, including Ultrasound Image Storage and Modality Worklist Information Model.
  • Association Policies: Clarius initiates one association at a time, does not support asynchronous communication, and provides implementation identifying information.
Media Interchange
Focuses on character set support and security measures.
Security
Outlines security profiles, association level security, and application level security to ensure data protection during image transfer.
Appendix A: IOD Details
Provides detailed information on supported IODs, including Ultrasound Image IOD Modules and Attributes, with extensive tables mapping attributes and modules.
Technical Specifications
The document outlines technical specifications and conformance requirements for Clarius DICOM implementation, focusing on storage, query, and verification activities.
Storage SOP Classes
All SOP Classes supported by the Storage Application Entity exhibit consistent behavior. The Application Entity Title used in Association Negotiation Request is either the Client AE Title or the Scanner’s Serial Number, truncated to 16 characters. The document details handling of C-STORE response statuses, including success, failure, and warning scenarios.
Query Worklist Information (C-FIND)
Specifies proposed presentation contexts for querying worklist servers and outlines SOP-specific conformance for Modality Worklist SOP Class, including optional matching key attributes and response status handling behavior for C-FIND operations.
Verify Connectivity (C-ECHO)
Describes proposed presentation contexts for verifying connectivity and SOP-specific conformance for the Verification SOP Class, detailing handling of C-ECHO response statuses.
Media Interchange
The Clarius DICOM implementation does not support media interchange.
Support of Character Sets
Supports the default repertoire designated by ISO-IR 6 and character set ISO_IR 192.
Security
No specific security measures or profiles are supported. The application does not accept Association Open Requests, and no application-level security measures are implemented.
Appendix A: IOD Details
Details on supported IODs for Ultrasound Image Storage and Ultrasound Multi-frame Storage SOP Classes, listing modules and attributes, indicating which are implemented and which are not.
Handling of Attributes
Specifies handling of various attributes, including patient information, study details, and series information. Attributes not implemented are noted, and handling procedures for implemented attributes are described.
General Overview
Provides detailed specifications and handling instructions for various modules related to ultrasound imaging and data management, including attributes, tags, and types for different modules such as General Equipment, Image, Cine, and US Region Calibration.
Specifications
  • General Equipment Module: Attributes include Manufacturer, Institution Name, and Software Versions. The Manufacturer is specified as 'Clarius Mobile Health Corporation'.
  • General Image Module: Attributes like Instance Number and Image Comments are detailed, with specific handling instructions for each.
  • Cine Module: Focuses on Frame Time Vector for cine captures, providing precision in milliseconds.
  • Image Pixel Module: Describes pixel data handling, including RGB triplets and compression methods.
  • US Region Calibration Module: Details on ultrasound region attributes, including spatial format and region location boundaries.
Procedures
  • Attributes are categorized by their tag and type, with specific handling instructions provided for each.
  • Handling instructions include default values, conditions for presence, and data formats.
Standards and Norms
  • Follows DICOM standards for attribute naming and handling.
  • Includes references to specific DICOM Nema Standards for certain attributes.
Recommendations
  • Ensure all attributes are correctly populated according to the specified handling instructions.
  • Follow the UCUM standard for measurement units, with noted exceptions.
Key Data from Tables
  • Table A-14: Maps Image Type Value 3 to Clarius Workflow, categorizing different anatomical areas and procedures.
  • Table A-15: Maps Transducer Types to Clarius Models, specifying different probe models.
  • Table A-22 and A-23: Provide mappings for View Code and View Modifier Code Sequences, detailing how system labels are translated into code values.
Critical Information
  • Attributes are marked with types indicating their necessity and conditions for implementation.
  • Specific character sets and timezone offsets are defined for data consistency.
Overview
Provides a detailed description of various DICOM code macros and their attributes, focusing on handling and requirements of each attribute within the DICOM standard.
Code Macros
  • Code Code Macro: Value type is "CODE". Includes Concept Name Code Sequence and Content Sequence with relationship types.
  • Date Code Macro: Value type is "DATE". Includes Concept Name Code Sequence and conditional date value.
  • Time Code Macro: Value type is "TIME". Includes Concept Name Code Sequence and conditional time value.
  • Image Code Macro: Value type is "IMAGE". Includes Referenced SOP Sequence and Content Sequence with relationship type.
  • SCoord Code Macro: Value type is "SCOORD". Includes Concept Name Code Sequence, Graphic Data, Graphic Type, and Content Sequence with relationship type.
Attribute Mapping
  • Details mapping of attributes across different DICOM IODs such as Modality Worklist, Image IOD, and Multi-frame IOD.
  • Attributes include Scheduled Procedure Step Start Date/Time, Modality, Performing Physician’s Name, Study Instance UID, Accession Number, Referring Physician’s Name, Institution Name/Address, Patient's Name/ID/Birth Date/Sex.
Type Codes
  • Type 1: Required with a valid value.
  • Type 2: Required but may be unknown or zero-length.
  • Type 3: Optional.
  • Type 1C: Conditional Type 1.
  • Type 2C: Conditional Type 2.
See more

Catalog excerpts

Clarius DICOM Conformance Statement-1

UNCONTROLLED DOCUMENT WHEN PRINTED 2025-03-27T01:41:09-07:00-83e24aa-47ef2f812e-12.0.4 51-10-00001-DHF-Rev 27

 Open the catalog to page 1
Clarius DICOM Conformance Statement-2

1. Conformance Statement Overview 6 2.3. Definitions, Terms, and Abbreviations 7 3.1.1. Application Data Flow 9 3.1.2.1. Storage Application Entity and Structured Report Application Entity 10 3.1.2.2. Modality Worklist Application Entity 10 3.1.2.3. Verification Application Entity 10 3.2.1. Storage Application Entity Specification 10 3.2.2. Modality Worklist Application Entity Specification 11 3.2.3. Verification Application Entity Specification 11 3.2.4. Structured Report Application Entity Specification 11 3.2.5.4. Implementation Identifying Information 12 UNCONTROLLED DOCUMENT WHEN PRINTED...

 Open the catalog to page 2
Clarius DICOM Conformance Statement-3

3.2.6. Association Initiation Policy 12 3.2.6.1. Activity - Send Images (C-STORE) 12 3.2.6.1.1. Proposed Presentation Contexts 12 3.2.6.1.2. SOP-Specific Conformance for Storage SOP Classes 13 3.2.6.2. Activity - Query Worklist Information (C-FIND) 14 3.2.6.2.1. Proposed Presentation Contexts 14 3.2.6.2.2. SOP-Specific Conformance for Modality Worklist SOP Class 14 3.2.6.3. Activity - Verify Connectivity (C-ECHO) 16 3.2.6.3.1. Proposed Presentation Contexts 16 3.2.6.3.2. SOP-Specific Conformance for Verification SOP Class 17 5. Support of Character Sets 19 6.2. Association Level Security 20 6.3....

 Open the catalog to page 3
Clarius DICOM Conformance Statement-4

Figure 3-1. Application Data Flow Diagram (Storage) 9 Figure 3-2. Application Data Flow Diagram (Worklist) 9 Figure 3-3. Application Data Flow Diagram (Verification) 9 Figure 3-4. Application Data Flow Diagram (Structured Report) 10 Table 1-1. Network Services 6 Table 2-1. Definitions, Terms, and Abbreviations 7 Table 3-1. SOP Classes for AE Storage 10 Table 3-2. SOP Classes for AE Worklist 11 Table 3-3. SOP Classes for AE Verification 11 Table 3-4. SOP Classes for AE Structured Report 11 Table 3-5. DICOM Application Context for AE 11 Table 3-6. DICOM Implementation Identifying Information 12...

 Open the catalog to page 4
Clarius DICOM Conformance Statement-5

Table A-9. Cine Module 27 Table A-10. Image Pixel Module 27 Table A-11. US Region Calibration Module 27 Table A-12. Multi-frame Module 28 Table A-13. US Image Module 28 Table A-14. Image Type Value 3 mapping to Clarius Workflow 29 Table A-15. Transducer Type Mapping to Clarius Models 30 Table A-16. SOP Common Module 30 Table A-17. Image Pixel Description Macro Attributes 31 Table A-18. Optional View and Slice Progression Direction Macro Attributes 31 Table A-19. Code Sequence Macro Attributes 31 Table A-20. View Code Sequence Mapping 32 Table A-21. View Modifier Code Sequence Mapping 33 Table...

 Open the catalog to page 5
Clarius DICOM Conformance Statement-6

The "Scanner" device developed by Clarius Mobile Health Corp. uses ultrasound technology to scan patients and produce medical images of various formats. DICOM is the standard for the communication and management of medical imaging information and related data [ 1]. The Clarius implementation provides functionality to transfer saved images or live image data to DICOM servers in several selectable image formats. In addition, the app also allows the user to download DICOM Worklists from their DICOM Worklist Servers to automatically fill in patient information. The following is the table of Supported...

 Open the catalog to page 6
Clarius DICOM Conformance Statement-7

This document is intended for health-related workers, software engineers, and designers. It is assumed that the reader has a working knowledge of DICOM. The Clarius DICOM implementation uses a subset of the functionality and features present in the DICOM Standard. The output allows for multiple choices of data compression: Bitmap (uncompressed) - Implicit VR Little Endian, Explicit VR Little Endian • The Implicit VR Little Endian and Explicit VR Little Endian Transfer Syntax are used for uncompressed images and the pixel data is straight RGB triplets per pixel for width x height pixels. Lossless...

 Open the catalog to page 7
Clarius DICOM Conformance Statement-8

Term Definition DIMSE DICOM Message Service Element IOD Information Object Definition ISO International Organization for Standardization PDU Protocol Data Unit PDV Protocol Data Value SCP Service Class Provider (DICOM Server) SCU Service Class User (DICOM Client) • NEMA PS3 Digital Imaging and Communications in Medicine (DICOM) Standard - http://medical.nema.org/ • DCMTK - DICOM Toolkit - https://dcmtk.org/ UNCONTROLLED DOCUMENT WHEN PRINTED 2025-03-27T01:41:09-07:00-83e24aa-47ef2f812e-12.0.4 51-10-00001-DHF-Rev 27

 Open the catalog to page 8
Clarius DICOM Conformance Statement-9

3. Networking 3.1. Implementation Model 3.1.1. Application Data Flow Figure 3-1. Application Data Flow Diagram (Storage) The Storage Application Entity sends images to a Remote Application Entity. Sending images is done for each study completed, and for all images in the study. There is the option of selecting 1 or more Remote Application Entities upon completion of the study, or later in a separate list of completed studies. Figure 3-2. Application Data Flow Diagram (Worklist) The Modality Worklist Application Entity downloads worklists from the user's DICOM worklist server listed in the DICOM...

 Open the catalog to page 9
Clarius DICOM Conformance Statement-10

Figure 3-4. Application Data Flow Diagram (Structured Report) The Storage Application Entity sends images to a Remote Application Entity. Sending images is done for each study completed, and for all images in the study. There is the option of selecting 1 or more Remote Application Entities upon completion of the study, or later in a separate list of completed studies. 3.1.2. Function Definition of AE 3.1.2.1. Storage Application Entity and Structured Report Application Entity The selection of Remote Application Entities upon completion of the study, or the selection of a Remote Application Entity...

 Open the catalog to page 10
Clarius DICOM Conformance Statement-11

The DICOM standard application context name for DICOM 3.0 is proposed: Table 3-5. DICOM Application Context for AE Application Context Name 1.2.840.10008.3.1.1.1 Clarius initiates one Association at a time for each destination and each request is processed one at a time. Clarius does not support asynchronous communication (multiple outstanding transactions over a single Association). UNCONTROLLED DOCUMENT WHEN PRINTED 2025-03-27T01:41:09-07:00-83e24aa-47ef2f812e-12.0.4 51-10-00001-DHF-Rev 27

 Open the catalog to page 11
Clarius DICOM Conformance Statement-12

The implementation information for these Application Entities are: Table 3-6. DICOM Implementation Identifying Information OFFIS DCMTK 3.6.5 1.2.276.0.7230010.3.0.3.6.5 3.2.6. Association Initiation Policy 3.2.6.1.1. Proposed Presentation Contexts Table 3-7. Proposed Presentation Contexts for Storing Single / Multi-frame Images Presentation Context Table Abstract Syntax Transfer Syntax Name UID Name List UID List Ultrasound Image 1.2.840.10008.5.1.4.1.1.6.1 Storage Implicit VR Little Endian Explicit VR Little Endian Hierarchical, First- Order Prediction Ultrasound Multi- 1.2.840.10008.5.1.4.1.1.3.1...

 Open the catalog to page 12