CAP-130 - 1

Series 130 Horizontal Flow Wall Modules allow time for any particulates to purge or blow off from the face of the HEPA filters. It also serves as a burn in time for the equipment prior to production operation. While the units are fully tested prior to shipping at the factory, a problem can sometimes develop because of unseen shipping damage or damage during the installation process. HEPA Filters Sizes – The HEPA filters used in the units are of standard industrial sizes. The size will be: 24 x 24 x 3 24 x 30 x 3 24 x 36 x 3 23-5/8 x 47-5/8 x 3 30 x 30 x 8.75 30 x 36 x 8.75 30 x 48 x 8.75 Media Pack: 2.875 with aluminum separators. Gasket: 1/4 x 3/4 closed cell, gasket or upstream side of the filter Filter Frames: Extruded aluminum Filter Efficiency: • 99.99% on 0.3 Micron (standard) “HEPA” • 99.99% on 0.12 Micron “ULPA” • Other Shipments Clean Air Products takes every reasonable precaution to ensure that your horizontal air flow cabinet arrives without damage. However, damage can occur in any shipment, and it is important that you note visible damage immediately with a notation on the consignee’s copy of the freight bill. Terms are F.O.B. factory, unless otherwise stated. Your inspection of either visible or concealed damage is the basis of your filing claim (which you must do at once) against carrier. An inspection then must be made to verify the claim against the carrier. Performance Specification All equipment is thoroughly inspected at the Clean Air Products’ factory at the time of shipment. Quality control is maintained by General Description The Series 130 horizontal module comes in a variety of sizes and styles. Some of the included information will depend on the style of unit purchased. The Series 130 units are constructed of painted steel as standard. This finish is designed to be compatible with normal cleanroom environments. When cleaning is required, a mild soap and water solution is recommended with a distilled water rinse. The distilled water is recommended because it does not have any mineral hardness that may leave a “mineral film” that can be wiped off and cause particulate contamination. Care should be used when cleaning the protective grills so the HEPA filters are not damaged during the cleaning process. NEVER attempt to clean the HEPA filter except by purging the filter. Rubbing or blowing the HEPA with high pressure air will only cause leaks. When the units are installed, let them run for a few days to “purge” the filters and to give the units some noncritical run time. Having the units run (with the prefilters in place) with “none cleanroom” activity going on around the units will not hurt the HEPA filter or appreciably shorten the life of the filter. (Prolonged running of the unit may load the filters somewhat, but, by letting them run, clean air coming out of the unit will prevent any contamination from getting onto the HEPA filter face.) The alternative is to tape plastic over the HEPA filters to protect them from dirt during these periods of time. By letting the units run for a time before “critical operation” start it will Model CAP 132 - 4 x 6 With top mounted blower and prefilter housing. For more information or to download or fax this product from the web, simply go to: www.cleanairproducts.com/130

Open the catalog to page 1
CAP-130 - 2

Series 130 Horizontal Flow Wall Modules Technical Data continued constant surveillance over the product, beginning at receipt of purchased material and concluding with a final inspection which certifies performance to Class 100 conditions of Federal Standard 209d, as well as to the unique requirements of each project. In all instances where product quality cannot easily be assessed on the end item, the product is inspected during subassembly fabrication. All electrical components are UL approved; all mechanical components are fabricated or purchased and inspected to performance requirements...

Open the catalog to page 2
CAP-130 - 3

front of the hood. Air currents will enter the hood and contaminate the interior. Many organizations have a 6 inch line marked on the table top. All work is performed behind this line to minimize contamination due to drafts. 7. Spills should be cleaned up immediately. Sponge the liquid near the HEPA filter first to keep the liquid away from the filter. Should liquid splash onto the HEPA filter, DO NOT attempt to wipe the filter. Wiping the filter will cause filter leaks. The liquid will dry, discolor and plug a small area of the filter in front. The filter is nearly 4 inch deep, and a small...

Open the catalog to page 3
CAP-130 - 4

Series 130 Horizontal Flow Wall Modules Technical Data continued state speed control located on the side of the blower. By turning the potentiometer on the cover of the electric box you change the speed of the blower. Some units have two direct drive blowers feeding a single plenum. On this type of unit, both blowers should be set at about the same speed. NOTE: on dual direct drive blowers the speed controls of both blowers need to be set very close to each other. If one is set too slow, it will not start properly. After the units speed is set, shut the units off. Let both blower wheels...

Open the catalog to page 4
CAP-130 - 5

solution can be used and wiped on the vinyl. This will leave a very thin soap film that will help prevent a static charge from building on the surface of the curtains. There are also cleanroom static control solutions that are commercially available. Many people have used both and had no problems. Care should be taken that these liquid solutions do not pose a contamination problem for your operation. There is an optical conductive vinyl curtain material that is also available if your product is static sensitive. This material has a carbon grid embedded inside the clear vinyl. Please consult...

Open the catalog to page 5

All Clean Air Products catalogs and technical brochures

  1. CAP-301

    5 Pages

  2. CAP-109BL

    3 Pages

  3. CAP-112

    2 Pages

  4. CAP-116

    4 Pages

  5. CAP-901

    2 Pages

  6. CAP-1532

    5 Pages

  7. CAP701

    8 Pages

  8. CAP-591

    6 Pages

  9. CAP-558

    3 Pages

  10. CAP-560

    3 Pages

  11. CAP-582

    2 Pages

  12. cap-550

    4 Pages

  13. cap-556

    3 Pages

  14. CAP-575

    5 Pages

  15. CAP-577

    5 Pages

  16. CAP1414

    5 Pages

  17. CAP1421

    2 Pages

  18. CAP1412

    2 Pages

  19. CAP-18

    2 Pages

  20. CAP701KD-ST

    2 Pages