Air Pocket Issues:
  • Light spots in or near the center of the sheet may indicate leaks in the door gasket, probe throughput, ventilator, or pipe joints.
  • Insufficient vacuum may also be a cause.
  • Service may be needed to address these issues.
Over Sterilization:
  • Occurs when the temperature is too high relative to the pressure, or the exposure time is too long, or both.
Wet Steam or Non-Condensable Gases:
  • Presence of spots, figures, or circle-formed figures in the center may indicate low pressure or temperature, or both.
  • Insufficient drying can also be a factor.
  • Steam may contain non-condensable gases, most often carbon dioxide.
Test Approval:
  • Uniform black/dark brown color in a 134°C cycle indicates approval.
  • Uniform black/dark grey color in a 121°C cycle also indicates approval.
See more

Catalog excerpts

Test pack-1

Air pocket (light spot in or near the centre of the sheet) - Door gasket or throughput of the probe, or ventilator or pipe joint leaks. - Insufficient vacuum - Service needed Over sterilization - Temperature too high in relation to the pressure, or too long exposure time or both. Wet steam or non condensable gases (spots/figures/circle formed figure in the center) - Pressure or temperature or both too low. - Insufficient drying. - Steam contains non condensable gases, most often carbon dioxide. -—f* - Uniform black/dark brown in 134 oC cycle - Uniform black/dark grey in 121 oC cycle

 Open the catalog to page 1

All Clinichem catalogs and technical brochures

  1. EVITTTDF

    6  Pages

  2. EVIT TTDF

    2  Pages

  3. PT20

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.