Technical Data Description: The A616S InCare
Syringe Pumps are available in two product variants: InCare and TCI. They support dosing units such as g, mg, µg, ng, IU, and mmol, with infusion rates ranging from 0.1 to 1,500 ml/h. The infusion volume can be set between 0.1 to 9,999 ml, and infusion time ranges from 1 minute to 99 hours and 59 minutes. The device allows for manual or automatic bolus administration and purging of the infusion line.
Performance Data: The pumps offer an accuracy of ±2% with CODAN qualified syringes and ±5% with customer qualified syringes. They feature a Keep Vein Open (KVO) rate configurable from off to 5.0 ml/h and an alarm system for unused pumps.
History and Interface: The device logs up to 1,000 events with timestamps and can connect to central systems via a nurse call interface. It includes an RS-232 interface for configuration and can connect to docking stations for Ethernet communication.
Operation: The pump features a monochrome LCD display with adjustable brightness and a night mode. It supports multiple infusion modes, including standard operation, dose calculation, TIVA, and TCI, with a medication database required for some modes.
Safety: Safety features include automatic pressure release, visual and audible alarms, DEFI proofing, and a key lock. The pressure limit for occlusion is adjustable between 200 to 1,200 mbar.
Power Supply: The device operates on a 1,800 mAh NiMH battery, providing up to 9 hours of operation at 5 ml/h. It supports mains power supply from 100 to 240 V.
Product Specification: The pump measures 320 x 100 x 180 mm and weighs 2.0 kg with the battery. It operates in temperatures from 5°C to 40°C and is compliant with various safety and usability standards.
Compliance: The device meets standards such as IP22, IEC 60601-1, and EN 60601-2-24, among others.
Accessories: Available accessories include CODAN docking stations and the ARGUSservice software for maintenance and configuration.