The Jenway 3510 is a versatile bench pH/mV meter designed for routine analysis. It offers up to three decimal place resolution and supports up to three calibration points, providing flexibility for enhanced performance needs. The device supports automatic calibration with pH buffers conforming to DIN, JIS, and NIST standards, as well as manual buffer entry.
Technical Specifications
  • pH Measurement: Range from -2.000 to +19.999, resolution options of 0.001/0.01/0.1, and accuracy of ±0.003. Calibration can be user-selected for 1, 2, or 3 points with automatic buffer recognition.
  • mV Measurement: Range of ±1999.9mV, resolution of 0.1/1mV, and accuracy of ±0.2mV.
  • Temperature Measurement: Range from -10 to 105°C, resolution of 0.1°C, and accuracy of ±0.5°C. Temperature compensation is available from 0 to 100°C.
  • Outputs: Analogue and RS232.
  • Connector: BNC.
  • Power: 9V AC ±10% at 50/60Hz.
  • Dimensions: 210 x 250 x 55 mm.
  • Weight: 850 g.
Key Features
  • Simultaneous readout of pH and temperature.
  • pH resolution up to 3 decimal places.
  • 1, 2, or 3 point calibration options.
  • Automatic or manual buffer selection.
  • Storage capacity for up to 32 results.
  • RS232 connection for printer or PC via DataWay.
Ordering Information
Part Code 351 001 includes the 3510 pH/mV meter, glass combination pH electrode, electrode stand and holder, ATC probe, BNC shorting plug, pH 4, 7, and 10 buffers, and UK power supply. Voltage variants are available as noted on page 94.
See more

Catalog excerpts

3510-1

Bench pH/mV Meter Simultaneous readout of pH and temperature pH resolution to 3 decimal places The 3510 is a versatile, simple to use pH, mV and temperature Automatic or manual buffer selection meter that is ideal for routine analysis. With up to three decimal place resolution and a choice of up to three calibration points the 3510 provides the user with added flexibility where future demands for enhanced performance may be required. A choice of pH calibration buffers to DIN, JIS and NIST standards can be used for automatic calibration, as well as manually entered buffer values. Technical Specification pH Range Accuracy Calibration User selectable 1, 2 or 3 point Automatic buffer recognition Jenway (2.00, 4.00, 7.00, 9.20 Temperature Range ATC and manual temperature 0 to 100°C compensation Outputs Ordering Information Part Code 3510 pH/mV meter supplied with glass combination pH electrode (924 005), electrode stand and holder (903 300), ATC probe (027 500), BNC shorting plug, pH 4, 7 and 10 buffers and * Voltage variants available see page 94

 Open the catalog to page 1

All Cole-Parmer Instrument Company catalogs and technical brochures

  1. SP-800 Series

    6  Pages

  2. SP-400 Series

    9  Pages

  3. SP-300 Series

    5  Pages

  4. SP-200 Series

    5  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.