Catalog excerpts

• Highly stable optics • Variable spectral bandwidth
Open the catalog to page 1
The SP-800-UV is a UV-Visible double-beam spectrophotometer with a variable spectral bandwidth. The spectrophotometer has an integrated user interface for local control and ease of navigation. The highly stable optics and two detectors measure the sample and reference simultaneously, optimizing measurement accuracy. The SP-800-UV has measurement modes for photometrics, concentration, multiwavelength, spectrum scanning, kinetics, quantitation, DNA, and protein analysis. Up to 10 wavelengths can be measured in the multi-wavelength measurement mode and quantitation curves can be created by...
Open the catalog to page 2
SP-800 Series Spectrophotometers Improved Optics SP-800-UV covers the UV-Visible wavelength range from 190 to 1100 nm with a variable spectral bandwidth ranging from 5 nm down to 0.5 nm using tungsten halogen and deuterium light sources. The range of bandwidths available allows a balance between resolution, accuracy and data precision to be maximized depending on the application, therefore providing a flexible platform to conform to multiple regulatory agencies for a variety of applications, all with one spectrophotometer. Three scan speeds are available enabling scan speeds of 100 to 2000...
Open the catalog to page 3
SP-800 Series Spectrophotometers Prism PC Software (continued) Prism PC software is supplied as standard and offers additional functionality, extensive post-measurement tools, unlimited results saving, and easy export of data. The Prism PC software can be used to fully control the functionality of the spectrophotometer. The measurement modes available in Prism mirror those of the instrument with measurement modes for photometrics, concentration, multi-wavelength, spectrum scanning, kinetics, quantitation, DNA, and protein analysis. Prism has pre-loaded methods for DNA analysis including...
Open the catalog to page 4
SP-800 Series Spectrophotometers Technical Specifications Model Wavelength range Wavelength resolution Wavelength accuracy Wavelength reproducibility Spectral bandwidth Photometric range Photometric accuracy Photometric reproducibility Stray light Stability Noise Multi-wavelength ±0.001 A/h at 500 nm after 15 minute warm-up 0.0005 A Up to 10 wavelengths, up to 20 wavelengths with PC software Ratio, difference, formulae with factors Spectrum range Scan speed Scan interval Auto peaks and valleys, zoom, addition, subtraction, peak ratios, smoothing, area under curve, wavelength table,...
Open the catalog to page 5
Technical Specifications (continued) Light source Lamp changeover Outputs Operating system Electrical supply Size (W x H x D) Weight DNA ratio, concentration, A320 correction Tungsten halogen and deuterium lamps 325 to 370 nm selectable USB and parallel Windows 2000, XP, Vista, Windows 7 115 VAC, 60 Hz or 230 VAC, 50 Hz 60.0 x 45.0 x 20.0 cm (23.5 x 7.9 x 17.7") 22 kg (48.5 lb) Ordering Information Cole-Pa rmer ‘Enquiries from Germany and France are now handled in our St Neots office by native speaking experts.
Open the catalog to page 6All Cole-Parmer Instrument Company catalogs and technical brochures
-
SP-500-600 Series
6 Pages
-
SP-300 GenovaNano Series
5 Pages
-
SP-400 GenovaBio Series
5 Pages
-
SP-400 Series
9 Pages
-
SP-300 Series
5 Pages
-
SP-200 Series
5 Pages
Archived catalogs
-
pH Buffers
1 Pages
-
Bench pH Meter Accessories
5 Pages
-
pH Electrodes
5 Pages
-
3520
1 Pages
-
3510
1 Pages
-
Fluorimetry Cuvettes
1 Pages
-
62 Series Fluorimeters
8 Pages
-
Colorimeter Accessories
2 Pages
-
6051 Bench Colorimeter
2 Pages
-
Conductivity Standards
2 Pages
-
9500 Bench DO2 Meter
1 Pages
-
6800 Spectrophotometer
2 Pages
-
67 Series Spectrophotometers
8 Pages
-
9500
1 Pages
-
970
1 Pages
-
3505
1 Pages
-
3345 Bench Ion Mete
1 Pages
-
3205 Portable Ion/pH Meter
1 Pages
-
4520
1 Pages
-
4510
1 Pages
-
470 Conductivity/TDS Meter
1 Pages
-
430 Conductivity/pH Meter
1 Pages
-
6051
2 Pages
-
6 2 Series Fluorimeters
8 Pages