1. Catalogs
  2. Convergent Technologies
  3. Convergys ESR-Analyzer-Range

Convergys ESR-Analyzer-Range

Convergys ESR-Analyzer-Range

Convergys ESR-Analyzer-Range

Product catalog summary
Overview: The document provides a comparative analysis of the Convergys® ESR series, which includes models ESR 10s, ESR 10, ESR 20, and ESR 100, along with ESR Vacuum Tubes. These devices are used for Erythrocyte Sedimentation Rate (ESR) testing in medical laboratories.
Specifications:
  • Area of Application: All models are designed for ESR testing.
  • Bar Code Reader: Optional for ESR 10s and ESR 10; integrated for ESR 20 and ESR 100.
  • Printer: Optional for ESR 10s and ESR 10; integrated for ESR 20 and ESR 100.
  • Reading Channels: ESR 10s and ESR 10 have 10 channels, ESR 20 has 20 channels, and ESR 100 has 100 channels.
  • Analytical Capacity: Varies from 20 tests/hour to 400 tests/hour depending on the model and mode.
  • Loading Capacity: Ranges from 10 to 100 samples at a time.
  • Loading Pattern: Random or batch (with mixing) for ESR 20 and ESR 100; random for others.
  • Mixing: Internal for ESR 10 and ESR 20; external for ESR 10s and ESR 100.
  • Interface: RS 232 for printer and barcode scanner; additional host interface for ESR 20 and ESR 100.
  • Stored Patient Results: Up to 500 results for ESR 10, ESR 20, and ESR 100.
  • Display: LCD with backlight for ESR 10s and ESR 10; graphical LCD for ESR 20 and ESR 100.
  • Instrument Size and Weight: Varies by model, with weights ranging from 0.9 kg to 5 kg.
  • Power Requirements: 12 Vdc with varying amperage; external power supply input of 100 - 240 Vac.
  • Operating Conditions: Temperature range of 15 – 32 °C and humidity of 45% - 85% non-condensed.
  • Reading Time: Selectable between 15, 30, or 60 minutes.
  • Measuring Method: IR beam with precision of +/- 0.2 mm.
  • Results Precision: +/- 1 mm.
  • Temperature Correction: Automatic compensation to 18° (Manley).
  • Results Format: Selectable in Westergren mm/30’, mm/1h, mm/2h.
  • Blood Draw Level Acceptance: 0.90 to 1.20 ml.
Ordering Information: Each model has a specific reference number and description, detailing the number of channels and integrated features like mixers, barcode scanners, and printers.
Additional Notes: Convergent Technologies reserves the right to change specifications without prior notice. Full specifications are available upon request.
See more

Catalog excerpts

Convergys ESR-Analyzer-Range-1

® Convergys ESR series Convergys® ESR 10s C Convergys® ESR 10 Conv vergys Convergys® ESR 20 R Convergys® ESR 100 Convergys® ESR Vaccum Tubes

 Open the catalog to page 1
Convergys ESR-Analyzer-Range-2

Convergent Technologies GmbH & Co. KG Software Center 1, D-35037 Marburg, Germany Tel: +49 6421 8869948 | Fax: +49 6421 8869958 Homepage: www.convergent-technologies.de e-mail: [email protected] 500 results Yes, after 10 minutes LCD 2x40 character with backlight 240 x 120 x 200 mm ~ 2 kg 12 VDC 0.5 A 100 - 240 Vac 50 - 60 Hz 15 – 32 °C 45% - 85% not condensed 15, 30 or 60 minutes IR beam +/- 0,2 mm +/- 1 mm automatic compensation to 18°(Manley) in Westergren mm/30’, mm/1h, mm/2h. 0.90 to 1.20 ml Yes, after 10 mintues LCD 2x40 character with backlight 180 x 100 x 140 mm ~ 0.9 kg...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.