SPECIFICATION Product Name: Laser Particle Count Sensor Module Item No.: PM5000S Version: V0.1 Date: 27th Jan, 2021
Open the catalog to page 1
Content Revision
Open the catalog to page 2
Laser Particle Count Sensor Module PM5000S Applications ▪ Clean Room ▪ Pharmaceuticals Industry ▪ Precision Machinery Industry ▪ Microbiological Industry Description PM5000S laser particle sensor module is based on laser scattering technology, which can accurately detect and calculate the number of suspended particles with different particle sizes in the air. The sensor can simultaneously output quantity of particles in 6 channels including 0.3um, 0.5um, 1.0um, 2.5um, 5.0um and 10um (PCS/L). ▪ It can output 0.3um, 0.5um, 1.0um, 2.5um, 5.0um, 10um 6 channel particle number at the same time....
Open the catalog to page 3
Specifications Laser Particle Sensor Specification Operating principle Laser scattering Measured particle range Measurement range Working condition Storage condition Counter efficiency Measuring error >100pcs/L:±30% of reading Condition: 0°C ~ 40°C, 50±10%RH Reference instrument: TSI9306 Response time Time to first reading Power supply Working current Standby current Digital output Life span
Open the catalog to page 4
Dimensions and Connector 1. Dimensions (Unit mm, tolerance ±0.2 mm) UART sending (TTL level @3.3V~5V)/IIC clock UART receiving (TTL level @3.3V~5V)/IIC data Output mode exchange TTL level @3.3V high level or floating is UART communication mode, low level is I²C 3. Connector Description The interface connector of Sensor is A2501WV-8P. The pitch is 2.5mm. The connection cable with female connector at both ends can also be customized.
Open the catalog to page 5
Product Installation The surface of air inlet and air outlet closing to inner wall of user machine, this is the best installing method. If cannot do like this, there should be air isolation structure between inlet and outlet, thus to avoid the air flow return in the machine. The appropriate installation ways are recommended as below. Recommended installation Internal profile
Open the catalog to page 6
User Attention ▪ When install PM5000S sensor module in your system or equipment, please make sure of unobstructed air-inlet and air-outlet. And make sure that there is no huge airflow faced to air-inlet and air-outlet. The inner profile and recommended installation as instruction in previous page to avoid dust deposit on the sensitive apparatus surface for the influence of the sensor accuracy. ▪ The air inlet size in the user machine’s inner wall should not be smaller than the sensor’s air inlet size. ▪ It should not be installed directly in the air duct for purifying when applied in air...
Open the catalog to page 7
UART Communication Protocol 1.General Statement 1) The data in this protocol is all hexadecimal data. For example, “46” for decimal [70]. 2) [xx] is for single-byte data (unsigned, 0-255); for double data, high byte is in front of low byte. 3) Baud rate: 9600; Data Bits: 8; Stop Bits: 1; Parity: No 4) It is default by continuously mode after powering on. Working mode will not be saved after powering off. 2.Format of Serial Communication Protocol Sending format of software: Start Symbol Detail description on protocol format: Protocol Format Start symbol Sending by software is fixed as [11H],...
Open the catalog to page 8
>0.3um, particles number, unit: pcs/L >0.5um, particles number, unit: pcs/L >1.0um, particles number, unit: pcs/L >2.5um, particles number, unit: pcs/L >5.0um, particles number, unit: pcs/L >10um, particles number, unit: pcs/L >0.3um, particles number = DF25*256^3 + DF26*256^2 + DF27*256^1 + DF28 >0.5um, particles number = DF29*256^3 + DF30*256^2 + DF31*256^1 + DF32 >1.0um, particles number = DF33*256^3 + DF34*256^2 + DF35*256^1 + DF36 >2.5um, particles number = DF37*256^3 + DF38*256^2 + DF39*256^1 + DF40 >5.0um, particles number = DF41*256^3 + DF42*256^2 + DF43*256^1 + DF44 >10um,...
Open the catalog to page 9
//open particle measurement //module is under particle measurement open status //close particle measurement // module is under particle measurement closed status 4.3 Set up and Read Particle calibration coefficient Send: 11 02 07 DF1 CS // Set up particle calibrated coefficient Send: 11 01 07 E7 // Read particle calibrated coefficient Response: 16 02 07 DF1 CS Function: Read/set up particle calibration coefficient Note: Calibration coefficient =DF1/100 Calibration coefficient setting valid range:0.1~2.5. 4.4 Read Software Version Number Send: 11 01 1E D0 Response: 16 0E 1E DF1~DF13 [CS]...
Open the catalog to page 10
I²C Communication Protocol 1. Brief Introduction a. This is an I²C protocol for PM5000S. The sensor module is lower computer, which is not able to initiate communication automatically. Communication is initiated via main controlled board, which reads data and sends control commands. b. Communication clock frequency <=100Khz 2. Communication Common START: start signal, send by main controlled board; STOP: stop signal, send by main controlled board; ACK: acknowledge signal, send by the sensor module if in bold; otherwise, send by main controlled board; NACK: non-acknowledge signal, send by...
Open the catalog to page 11
3.2 Read Data Command Send by main controlled board: START+READ+ACK+P1+ACK+P2+ACK+……+P32+NACK+STOP Data Device address Byte content Sensor address and read/write command This byte is 0x51 when read data Frame header Frame length Number of byte, not including length of device address (from P1 to P32, 32 bytes in total) Sensor status Status “1” Means sensor is closing. Status “2” Means sensor is under measuring. Status “7” Means the faulty, temperature is too high or too low, or fan speed is too high or too low. Status “0x80” Means measuring data is stable (only for dynamic or timing...
Open the catalog to page 12
Product Code Description Structure Type 000:Structure Number ※ Product Application 5: Cleanroom Product Category PM:Dust Sensor Accessories can be matched with sensor Accessories PN To measure air flow Gasboard 7500H-OPC Ultrasonic Flow Meter rate, measurement range is 0~5L/min For more information, please contact with Cubic team.