The Mainstream NDIR ETCO2 Sensor by Wuhan Cubic Optoelectronics Co., Ltd is designed to measure CO2 concentration at the end of an exhaled breath using infrared absorption spectroscopy. It is primarily used for monitoring respiratory end tidal CO2 concentration and respiration rate, and can be easily integrated into monitors or anesthesia machines.
Specifications
  • Measurement Range: 0 to 10% vol; 0 to 20% vol
  • Warm-up Time: 10 seconds
  • Accuracy: ±2mmHg @<5% CO2 (at BTPS); ±10% of reading @ >5% CO2 (at BTPS)
  • Respiration Rate Range: CM2200: 2~15 bpm; CM2200B: 2~10 bpm
  • Response Time: CM2200: Detector: 28ms; System: 200ms; CM2200B: Detector: 125ms; System: 200ms
  • Data Interface: TTL or RS232
  • Power Supply: DC 5.0V±5%
  • Working Temperature: 5ºC~50ºC
Features
  • Fast response, long lifespan, low maintenance costs
  • High accuracy with temperature and pressure compensation technology
  • EMI system to avoid electromagnetic interference
  • Compatible with Respironics CapnoStat5 or CPT Protocols
Applications
The sensor is suitable for use in clinical anesthetic machines, CACR, PACU, ICU, pre-hospital rescue, and respirators.
See more

Catalog excerpts

Mainstream NDIR ETCO2 Sensor-1

Mainstream NDIR ETCO2 Sensor For Healthy Respiratory Wuhan Cubic Optoelectronics Co., Ltd

 Open the catalog to page 1
Mainstream NDIR ETCO2 Sensor-2

Mainstream NDIR ETCO2 Sensor Introduction It is designed by the principle of infrared absorption spectroscopy for CO2 concentration at the end of an exhaled breath, which is widely used to monitor the respiratory end tidal CO2 concentration and respiration rate. It is easy to be integrated to the monitors or anesthesia machines with an 8-pin socket and the preparation of the corresponding display/control software. Features ■ Fast response, long lifespan, low maintenance costs ■ Compensation technology used for high accuracy of temperature and pressure ■With EMI system, avoiding the interference of...

 Open the catalog to page 2
Mainstream NDIR ETCO2 Sensor-3

Applications ■ Clinical anesthetic machine ■ CACR, PACU, ICU, pre-hospital rescue ■ Respirator Tel: 86-27-81628827 Website: www.gassensor.com.cn Email: [email protected]

 Open the catalog to page 3

All Cubic Sensor and Instrument Co.,Ltd. catalogs and technical brochures

  1. About Cubic

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.