The document provides technical specifications and applications for the Cubic PM5000 Optical Particle Counter, designed by Cubic Sensor and Instrument Co., Ltd, for use in clean rooms.
Specifications
  • Operating Principle: Laser scattering.
  • Measured Particle Range: 0.3μm to 10μm.
  • Measurement Range: 0 to 10,000,000 pcs/L.
  • Resolution: 1 pcs/L.
  • Working Conditions: -30°C to 70°C, 0-95% RH (non-condensing).
  • Storage Conditions: -40°C to 85°C, 0-95% RH (non-condensing).
  • Counter Efficiency: 70% at 0.3μm, 99% at ≥0.5μm.
  • Measuring Error: ±30 pcs/L for ≤100 pcs/L, ±30% of reading for >100 pcs/L.
  • Response Time: 1 second.
  • Power Supply: DC 5V±0.1V, Ripple wave <50mV.
  • Dimensions: W85*H74*D24.9 mm.
  • Digital Output: UART_TTL/IIC (3.3V/5V).
  • Life Span: Over 5 years of continuous working.
Performance and Output
The PM5000 offers real-time and time-averaged outputs, with longer averaging times providing higher accuracy and stability. A Quick Response Mode is available for rapid environmental changes, reducing the average time from 60s to 15s while maintaining accuracy.
Comparison and Testing
Comparison tests with the TSI9306 standard equipment show the PM5000's performance across various particle sizes and concentrations.
Applications
The PM5000 is suitable for portable particle counters and facility monitoring systems in clean workshops or rooms. It is applicable in pharmaceutical, food, electronic workshops, dust-free laboratories, and clean operating rooms.
Contact Information
For further inquiries, contact Cubic Sensor and Instrument Co., Ltd at [email protected] or visit their website at www.gassensor.com.cn.
See more

Catalog excerpts

Optical Particle Count Technology for Clean Room Cubic PM5000-1

Optical Particle Count Technology for Clean Room Cubic PM5000 Cubic Sensor and Instrument Co.,Ltd 1

 Open the catalog to page 1
Optical Particle Count Technology for Clean Room Cubic PM5000-2

Optical Particle Counter-Cubic PM5000 Laser Particle Sensor Specification Operating principle Laser scattering Measured particle range Measurement range Working condition Storage condition Counter efficiency Measuring error ≤100pcs/L:±30pcs/L >100pcs/L:±30% of reading Condition: 0°C ~ 40°C, 50±10%RH Reference instrument: TSI9306 Response time Time to first reading Power supply Working current Standby current Digital output Life span

 Open the catalog to page 2
Optical Particle Count Technology for Clean Room Cubic PM5000-3

Optical Particle Counter-Cubic PM5000 Real time and time average output Particle Concentration /pcs/L Real time Output 10S Average Output 30S Average Output 60S Average Output  Concentration is 1000pcs/L for the example.  Longer time average, higher accuracy and less deviation. Time/s Particle Concentration /pcs/L  When the particle concentration is changing, measurement stability is better and vibration is less when the average time is longer.  The accuracy is better with longer time delay to reflect the change.

 Open the catalog to page 3
Optical Particle Count Technology for Clean Room Cubic PM5000-4

Optical Particle Counter-Cubic PM5000 Average and Quick Response Output Particle Concentration Time To keep stable output is the first target. If some application scenarios need PM5000 to give fast response to environment change, then a output can be designed in the software of PM5000. Quick Response Mode When the sensor detect a rapid and continuous change of particle concentration in 6 seconds, the sensor could switch to quick response mode and output a shorten average time (from 60s to 15s). In quick response mode, the response time is increased significantly and benefit from certain average,...

 Open the catalog to page 4
Optical Particle Count Technology for Clean Room Cubic PM5000-5

Optical Particle Counter-Cubic PM5000 Comparison test of PM5000 and TSI9306(standard equipment) is implemented with various concentration and channels. Particle concentration pc/L Sampling Sequence No. Particle concentration pc/L Sampling Sequence No. Sampling Sequence No.

 Open the catalog to page 5
Optical Particle Count Technology for Clean Room Cubic PM5000-6

Application of Cubic PM5000 PM5000 can be used in portable particle counters, or installed in facility monitoring system of clean workshop or clean room. Portable Particle Counter Facility Monitoring System of Industrial Clean Workshop

 Open the catalog to page 6
Optical Particle Count Technology for Clean Room Cubic PM5000-7

Application of Cubic PM5000 Clean environment is required in many fields, airborne particle concentration should be controlled and monitored any time. PM5000 can be used in all these fields. Pharmaceutical Workshop Electronic Workshop Food Workshop Sophisticated Equipment Workshop Dust-Free Laboratory Clean Operating Room

 Open the catalog to page 7
Optical Particle Count Technology for Clean Room Cubic PM5000-8

Cubic Sensor and Instrument Co.,Ltd Tel: +86-027-8162 8827 Fax: +86-027-87401159 Email: [email protected] Web: www.gassensor.com.cn Add: Fenghuang No.3 Road, Fenghuang Industrial Park, Eastlake Hi-tech Development Zone, 430205, Wuhan, China

 Open the catalog to page 8

All Cubic Sensor and Instrument Co.,Ltd. catalogs and technical brochures

  1. About Cubic

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.