Product catalog summary
Overview: The Dual Wire Myograph System - 420A is designed for the simultaneous testing of two vessels with diameters ranging from 30 μm to 3 mm. It allows for independent testing or comparative studies by mounting vessels in a divided or single chamber.
Specifications:
  • Chamber: Made of acid-resistant stainless steel with a minimum volume of 2.7 ml. It can be divided for separate testing or combined for comparative studies.
  • Temperature Control: The system operates within a range of 15.0 to 50.0 °C with a resolution of 0.1 °C and stability of ±0.2 °C.
  • Transducer: Measures force/tension with an output reading in mN, a range of ±200 mN, and a resolution of 0.01 mN.
  • Output: Data communication via USB 2.0 with four analogue output channels and a range of ±2.5 V.
Features:
  • Manual micropositioners for precise tension control.
  • Glass windows in the chamber base for morphological or fluorescence measurements.
  • Digital output directly integrated with LabChart Pro for data analysis.
  • Optional electronic valve for easy chamber control.
Procedures: Vessels are mounted as ring preparations over parallel wires. The system allows for manual control of vessel circumference and stretch, with force transducers measuring tension development. The chamber is heated and can be filled with a physiological salt solution at 37°C, maintaining vessel viability for up to 12 hours.
Applications: The system is suitable for examining vessel contractility and reactivity under isometric conditions. It supports comparative studies of treated vs. untreated or diseased vs. healthy samples by exposing vessels to identical drug concentrations.
Contact Information: For further inquiries, contact Danish Myo Technology A/S or DMT-USA, Inc. via the provided email addresses and phone numbers.
See more

Catalog excerpts

420A-1

DUAL WIRE MYOGRAPH SYSTEM - 420A Manually operated micropositioners for accurate tension control Chamber can be easily divided to keep 2 mounted samples separate Glass windows in chamber base facilitates morphological or fluorescence measurements Digital output. Data directly piped into Labchart Pro The Dual Wire Myograph System - 420A is designed for simultaneous testing of two vessels with diameters of 30 μm - 3 mm, independently. The vessels are mounted as ring preparations by threading them over two parallel wires and securing the wires to two supports or “jaws”. One support is attached to a precision micrometer, allowing manual control of vessel circumference and stretch. The other support is attached to a force transducer for measurement of force/tension development. The base of the chamber contains glass windows allowing morphological observations or fluorescence measurements on an inverted microscope. Typically, the preparation is kept in the heated vessel chamber in a physiological salt solution at 37°C, bubbled with oxygen where the vessels remain viable for up to 12 hours. The preparation is mounted in a heated 10 ml acid-resistant stainless steel chamber, which can be covered with a lid with ports for rapid suction/draining, refilling and bubbling of oxygen supply. Following mounting and equilibration, the passive length-tension relationship of the vessel is determined. During the actual experiment, the circumference of the vessel is kept constant. Compounds can be added directly to the chamber, and the vessel’s contractility and reactivity are measured under isometric conditions. In the Dual Wire Myograph System - 420A, the chamber can be divided for independent testing of each vessel. The two vessels also can be mounted in a single chamber by removing the divider. This allows direct comparative studies of vessels from treated/untreated conditions or diseased vs. healthy patient samples, for example, by exposing the vessels to identical concentrations of drug or compound. It is thus possible to examine whether a given pathological state is associated with altered morphology or reactivity. As an option, an electronic valve can be added to the system for easy control and emptying of the chamber. The Wire Interface with touch screen makes it easy to set up and use. Furthermore, the Wire Interface is compatible with the DMT Device Enabler allowing automatic recognition of supported devices by LabChart, use of multiple devices simultaneously, correct units and ranges in LabChart channels and simultaneous recording of data into LabChart alongside a PowerLab. The DMT Device Enabler allows the Dual Wire Myograph System - 420A to stream data directly i

 Open the catalog to page 1
420A-2

DUAL WIRE MYOGRAPH SYSTEM - 420A CHAMBER: Chamber volume (min) Chamber material Acid resistant stainless steel Vessel size Vessel normalization Micrometer resolution Mounting type TEMPERATURE: Range TRANSDUCER: Output reading Force calibration OUTPUT: Data communication Analogue output channels Analogue output range

 Open the catalog to page 2

All Danish Myo Technology A.S (DMT) catalogs and technical brochures

  1. 820MO

    2  Pages

  2. 630ma

    2  Pages

  3. 360CW

    2  Pages

  4. 320a

    2  Pages

  5. 840MD

    2  Pages

  6. 820MS

    2  Pages

  7. 120CP

    2  Pages

  8. 114P

    2  Pages

  9. 112PP

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.