Dansereau Dental Equipment D r y Va c u u m I n s t a l l S p e c s Dansereau Health Products, Inc. 1581 Commerce Street Corona, CA 92880 (800) 423 - 5657 Website WWW.DHPDENTAL.COM UL - 60601 Issued 2003/04/25 Ed. Rev 2006/004/26 CSA C22.2#601.1 Issued 1990/01/01 / (R2005) USFDA Medical Device Lic.# 2027216

Open the catalog to page 1
Super Flow 10 - 2

DHP Dry Vacuum Pump Install Dimensions 24” 24” Required Air Flow Perimeter 12 Inches NOTE: Air Flow is critical element in Dry Vacuum System. A minimum 12 inches of clearance is required to ensure proper air flow. An air vent fan for circulation in the Dry Vacuum Area is also require. Optimally an Air Conditioning Vent is recommended by not requir

Open the catalog to page 2
Super Flow 10 - 3

DHP Dry Dual Vacuum Pump Install Dimensions 55” NOTE: Air Flow is critical element in Dry Vacuum System. A minimum 12 inches of clearance is required to ensure proper air flow. An air vent fan for circulation in the Dry Vacuum Area is also require. Optimally an Air Conditioning Vent is recommended by not required. Required Air Flow Perimeter 12 Inc

Open the catalog to page 3
Super Flow 10 - 4

DHP Vacuum Pump - Pre-installation Guide Location The DHP Dry Vacuum System should be installed in a well ventilated area. The DHP Dry Vacuum System is air cooled and without proper air circulation the lifespan of the vacuum pump could be shortened significantly. The sound level of the vacuum pump, when in operation, is very quiet and can be placed within a dental facility. The ambient temperature in the dental office equipment room should never exceed 40 degrees Fahrenheit minimum and 100 degrees Fahrenheit maximum. The dental office equipment room will require a minimum 5 air changes per...

Open the catalog to page 4
Super Flow 10 - 5

Specifications: Voltage 240 VAC / Hz 50 - 60 - If 208V Present "Buck Booster Required " • Output H.P. - 2 / Current Amps 11 • Free Delivery (CFM) 70 / Vacuum In. Hg. 16" max 8” continuous • Weight (lbs.) 97 • Width (inches) 16 • Length (inches) 20 • Height (inches) 16 • Tank (gallons) 16 • Includes: Low voltage control, separator tank with automatic drain and check valve • Copper NO LONGER RECOMMENDED FOR VACUUM LINES OR VENTING

Open the catalog to page 5
Super Flow 10 - 6

DHP Dry Vacuum Specifications Vacuum Line: See Chart Page 4. Dansereau Health Products, Inc. strongly recommends Schedule 40PVC for use as the Vacuum Line. However, be aware some Building Codes require a metal type vacuum line and require Copper as the Vacuum Line. Copper will eventually deteriorate over years of service, Schedule 40 PVC will not. NOTE: IF YOUR VACUUM LINE INSTALLATION IS OUTSIDE THE SCOPE LISTED BELOW CONTACT DANSEREAU FOR SPECIAL INSTRUCTIONS. Operatory Termination - See Vacuum Sizing Chart for Specs. Site Requirements: Environment Conditions Operating Conditions - Indoor...

Open the catalog to page 6
Super Flow 10 - 7

OVERHEAD PLUMBING If you are required to run Overhead Plumbing the specifications below will offer the best installation conditions for optimal performance. Incorrect installation of overhead vacuum piping will ensure low quality vacuum pressure. Plumbing Rules for Overhead Piping: • Keep the vertical lift height as short as possible. • The vertical pipe must tee into the top of the horizontal line. • Run the vertical line in 1/2” pipe. • Run the horizontal overhead line in 1-1/2” or preferably 2” size pipe. For more than 5 Operatories consult with Dansereau. • Nitrous oxide scavengers must...

Open the catalog to page 7
Super Flow 10 - 8

BRANCH LINE TERMINATION TO TWO OPERATORIES THIS BRANCH LINE TERMINATION IS SHOWN CORRECTLY. NEVER USE THE INCORRECT EXAMPLE SHOWN AT LEFT FOR VACUUM SYSTEMS. MOST COMMON VACUUM PLUMBING ERRORS 00 NOT ALLOW ANY RIPE TO BRANCH OFF ANOTHER RIPE BELOW THE CENTERLINE OF THE MAIN OR BRANCH LINE PIPE. SEE FIG. S FOR CORRECT BRANCH LINE TAKE OFF.

Open the catalog to page 8
Super Flow 10 - 9

DHP Dry Vacuum Specifications Electrical Requirements LINE VOLTAGE - Single phase 240v/50/60hz electricity is required for proper operation of the D H P Dry Vacuum System. All electrical sources to the DHP Vacuum Pump MUST BE PROPERLY GROUNDED! All DHP Dry Vacuum Systems are operated by a LOW VOLTAGE SWITCH or the AUTO WASH DOWN SYSTEM WILL NOT OPERATE. LOW VOLTAGE - A 18/3 low voltage rated wire should be run from one central location to allow the DHP Dry Vacuum System to be turned on and off. Plumbing Requirements WATER LINE - A 1/2" cold water source is required for proper operation of...

Open the catalog to page 9