X - ray Machine Specifications
1 / 4Pages

Catalog excerpts

X - ray Machine Specifications - 1

Dansereau Dental Equipment X - ra y M a c h i n e S p e c i f i c a t i o n s Wa l l M o u n t e d And Pa s s T h r u M o u n t e d Dansereau Health Products, Inc. 1581 Commerce Street Corona, CA 92880 (800) 423 - 5657 Website WWW.DHPDENTAL.COM

Open the catalog to page 1
X - ray Machine Specifications - 2

X R AY M A C H I N E S Pass Thru X ray Framing and Support Specifications 49” Mounting Plate Location requires door hinges to be within 3 “ from bottom of cabinet NOTE: Critical Element is Location of ELT Wire from the Main Controller Sub Controller must be mounted in a manner that it will withstand a 10lb s

Open the catalog to page 2
X - ray Machine Specifications - 3

Pass Thru X ray Mounting Specifications Sub Controller Wall Mount Template Electrical Requirements: Power Supply 120VAC=/-10% with a 3 Wire Grounded Dedicated Circuit. 20 Amperes - Wire Size 12AWG - Line Voltage must be within 2 - 3% at 10.8 Amps. Interconnecting wiring between Sub Controller and Main Controller 4 Conductor, 20AWG, 300V with a Maximum wire run of 33 feet Critical Element All Wire to be thru this Hole !!! NOTE: Contact Dansereau for actual Full Size Templates

Open the catalog to page 3
X - ray Machine Specifications - 4

Wall Mt. X ray Framing /Support Specifications If the X-RAY is mounted in a manner other than specified by Belmont Mfg the support of the wall must be checked and verified to be adequate. Sub Controller must be mounted in a manner that it will withstand a 10lb shear load. Critical Element All Wire to be thru this Hole !!! Electrical Requirements: Power Supply 120VAC=/-10% with a 3 Wire Grounded Dedicated Circuit. 20 Amperes - Wire Size 12AWG - Line Voltage must be within 2 - 3% at 10.8 Amps. Interconnecting wiring between Main Controller: Sub Controller and Main Controller 4 The Main...

Open the catalog to page 4