Product catalog summary
Specifications Overview
The document outlines the specifications for Dansereau Dental Equipment's X-ray machines, specifically focusing on wall-mounted and pass-thru mounted configurations. Key elements include mounting requirements, electrical specifications, and critical installation notes.
Mounting Specifications
  • Wall Mounted: The main controller must be mounted on two 2” x 4” wood studs with 16” centers. The mounting hardware should withstand a 100lb shear load and a 450 drawal force.
  • Pass Thru Mounted: The sub-controller must withstand a 10lb shear load. The mounting plate location requires door hinges to be within 3 inches from the bottom of the cabinet.
Electrical Requirements
  • Power Supply: 120VAC +/-10% with a 3-wire grounded dedicated circuit.
  • Current: 20 Amperes, Wire Size: 12AWG.
  • Line Voltage: Must be within 2-3% at 10.8 Amps.
  • Interconnecting Wiring: Between sub-controller and main controller, 4 conductor, 20AWG, 300V with a maximum wire run of 33 feet.
Critical Installation Notes
  • Ensure all wiring passes through specified holes as indicated in the templates.
  • Contact Dansereau for full-size templates for accurate installation.
  • Verify wall support adequacy if installation deviates from specified methods.
See more

Catalog excerpts

X - ray Machine Specifications-1

Dansereau Dental Equipment X - ra y M a c h i n e S p e c i f i c a t i o n s Wa l l M o u n t e d And Pa s s T h r u M o u n t e d Dansereau Health Products, Inc. 1581 Commerce Street Corona, CA 92880 (800) 423 - 5657 Website WWW.DHPDENTAL.COM

 Open the catalog to page 1
X - ray Machine Specifications-2

X R AY M A C H I N E S Pass Thru X ray Framing and Support Specifications 49” Mounting Plate Location requires door hinges to be within 3 “ from bottom of cabinet NOTE: Critical Element is Location of ELT Wire from the Main Controller Sub Controller must be mounted in a manner that it will withstand a 10lb s

 Open the catalog to page 2
X - ray Machine Specifications-3

Pass Thru X ray Mounting Specifications Sub Controller Wall Mount Template Electrical Requirements: Power Supply 120VAC=/-10% with a 3 Wire Grounded Dedicated Circuit. 20 Amperes - Wire Size 12AWG - Line Voltage must be within 2 - 3% at 10.8 Amps. Interconnecting wiring between Sub Controller and Main Controller 4 Conductor, 20AWG, 300V with a Maximum wire run of 33 feet Critical Element All Wire to be thru this Hole !!! NOTE: Contact Dansereau for actual Full Size Templates

 Open the catalog to page 3
X - ray Machine Specifications-4

Wall Mt. X ray Framing /Support Specifications If the X-RAY is mounted in a manner other than specified by Belmont Mfg the support of the wall must be checked and verified to be adequate. Sub Controller must be mounted in a manner that it will withstand a 10lb shear load. Critical Element All Wire to be thru this Hole !!! Electrical Requirements: Power Supply 120VAC=/-10% with a 3 Wire Grounded Dedicated Circuit. 20 Amperes - Wire Size 12AWG - Line Voltage must be within 2 - 3% at 10.8 Amps. Interconnecting wiring between Main Controller: Sub Controller and Main Controller 4 The Main Controller...