Model 10A ECG Simulator Model 20A ECG / Arrythmia Simulator Operating Manual
Open the catalog to page 1Model 10A / 20A Patient Simulators Operating Manual © 2006-2014 Datrend Systems Inc. Unit 130 - 4020 Viking Way Richmond, BC • CANADA • V6V 2L4 Tel: 800.667.6557 (North America Only) or 604.291.7747 • Fax 604.294.2355 e-mail customerservice@datrend
Open the catalog to page 3UNPACKING INSTRUCTIONS: Package Contents: Patient Simulator Battery: 9V type 1604 Operating Manual The Model 10A and 20A Patient Simulators are powered by a 9V alkaline battery (included). To install the battery, see section 2.1 of this Manual.
Open the catalog to page 4To order this manual, use Part Number 6100-450 Revision Revision History Description - 5 Year Warranty - Update Corporate Address - Update Corporate Address Copyright Datrend Systems Inc. (“DSI”) agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to DSI. Unpacking and Inspection Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage...
Open the catalog to page 5Warranty Warranty and Product Support DSI warrants this instrument to be free from defects in materials and workmanship under normal use and service for two (5) years from the date of original purchase. During the warranty period DSI will, at our option, either repair or replace a product that proves to be defective at no charge; provided you return the product (shipping, duty, brokerage and taxes prepaid) to DSI. Any and all transportation charges incurred are the responsibility of the purchaser and are not included within this warranty. This warranty extends only to the original purchaser and...
Open the catalog to page 6WARNING Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. DSI will not be responsible for any injuries sustained due to unauthorized equipment modifications. DSI DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION. THIS PRODUCT CONTAINS NO USER-SERVICEABLE COMPONENTS. UNAUTHORIZED REMOVAL OF THE INSTRUMENT COVER SHALL VOID THIS AND ALL OTHER EXPRESSED OR IMPLIED WARRANTIES.
Open the catalog to page 7MODEL 10A / 20A OPERATING MANUAL Table of Contents 1. OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 Models 10A and 20A . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1.1 Lead Interface: . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1.2 User Interface: . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2. OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1 Battery Installation . . . . . . . . . . . . . . . . . . ....
Open the catalog to page 9MODEL 10A / 20A OPERATING MANUAL DSI is a trademark of Datrend Systems Inc. All other trademarks referenced in this manual are the property of the respective trademark holders. Table of Contents # Page x
Open the catalog to page 10MODEL 10A / 20A OPERATING MANUAL Compact and lightweight, the Datrend Systems Model 10A and 20A patient simulators are designed to test patient monitors and telemetry systems, anywhere a quick ECG or arrythmia simulation is needed. The versatile 6-lead configuration provides an additional pre-cordial lead and allows you to check reference leads without the need for a full 12 lead layout. This unique V6 lead allows some monitors to simulate any of the other precordial leads (V6 - V9) without losing the signal from the standard five lead setup. 1.1.1 Lead Interface: RA, RL , LA , LL, V1, V6, Battery...
Open the catalog to page 11MODEL 10A / 20A OPERATING MANUAL 1.2 Specifications Model 10A Model 20A Battery Single LEDTest: (OK @ 7.5 VDC) 0.5, 1 and 2 mV PVC1, PVC2, Pair PVC, PVC 1 & 2 ongoing, Multifocal PVCs, Ventricular Fibrillation and Asystole
Open the catalog to page 12MODEL 10A / 20A OPERATING MANUAL 2. OPERATION2.1 Battery Installation The Model 10A and 20A are both powered by a standard 9V alkaline battery (9 V Type 1604). The 9V battery is included with the unit, but is not installed as shipped from the factory. Follow these instructions to install or replace the battery: 2.1.1 Ensure the power is switched OFF. 2.1.2 Locate the battery compartment on the rear of the instrument. Open the battery door by pressing on the door and simultaneously pushing the door outwards. 2.1.3 Observe polarity of the battery snap connections. Install or exchange the battery...
Open the catalog to page 13MODEL 10A / 20A OPERATING MANUAL Battery Test This test is conducted identically on the 10A and 20A simulators. With the power switch turned on (the default rate and amplitude settings will be flashing on the LEDs of the unit), touch the appropriate terminals of a standard 9V battery to the battery test terminals. If the battery has a charge of 7.5 VDC or more, the Battery / Lead Test LED will illuminate. Ensure that the polarity is correct - a battery placed in reverse polarity will not illuminate the LED. Lead Test This test is conducted identically on the 10A and 20A simulators. With the power...
Open the catalog to page 14MODEL 10A / 20A OPERATING MANUAL 2.4.1 ECG Testing: Attach patient snap leads to the appropriate terminals of the Simulator and power the Simulator ON. The default waveform should be displayed on the Patient Monitoring Device (PMD). Pressing the RATE button a single time will cycle the rate upward to the next setting. Pressing the RATE button repeatedly will cycle through the rates until released at the desired setting. Pressing the mV button will cycle the amplitude upward to the next setting. Selecting the 2 Hz waveform will display a 2 Hz square wave. The ECG monitor may display a rate of...
Open the catalog to page 15MODEL 10A / 20A OPERATING MANUAL 2.4.3 Sample Waveforms: Lead II (RA, LA, LL [ref.]): Normal Sinus Rhythm Ventricular Fibrillation Depending upon the monitoring equipment under test, the appearance of the waveform may differ slightly from the figures, above.