Oxitest Plus 7 P u l s e O x i m e t e r Te s t e r The front panel keyboard design makes Puritan Bennett sensor port, developed in adjustment of the five preset oxygen saturation cooperation with Nellcor®. This exclusive levels, six preset heart rates and four preset feature allows correct selection of the amplitudes a breeze. In addition, features instrument’s simulation output to the specific such as Signal Artifact simulations, preset R-Cal value of the attached Nellcor® probe. patient parameters, Alarm Testing and Report printing make Oxitest Plus 7 truly unique. Accur acy M att er s • www.datr end.com Oxitest Plus 7 features the exclusive Nellc
Open the catalog to page 1Oxitest Plus 7 – Performance Specifications Saturation (SpO2): Range of adjustment is according to pulse oximeter manufacturer and/or model. Masimo, Mindray, Nellcor, and Nellcor OEM’s (e.g., Critikon; Protocol; Philips; etc.): Variable from 35% to 100% in steps of 1% SpO2 presets at 35, 70, 80, 90, 93, and 97% Alarm Tests: Five automated test sequences for determining oximeter alarm response time to Low SpO2 ; Low Heart Rate; High Heart Rate; Low Perfusion; and Motion Artifact conditions. Also determines time required for oximeter alarm to self-clear after test conditions are returned to normal....