vPad-BP
1 /1Page

vPad-BP

vPad-BP
1 /1Page

Catalog excerpts

vPad-BP-1

a vPad-BPTM - Performance Specifications Manufacturer Envelopes: GE Dinamap and Dinamap Pro Critikon Dinamap Plus Datascope Passport Welch-Allyn Spot LXi and 52000 Fukuda Dynascope Colin Press-Mate ...and more custom simulations Pressure Units: mmHg, mbar, kPa, in^O, cmH2O BP Presets: Systolic/Diastolic (mmHg) BP Simulation: Simulation type: oscillometric Rate Range: 20 - 240 BPM Accuracy: ±0.25 BPM in sync mode ±1 BPM otherwise Amplitude: 0 - 2 mL 1.25 mmHg into 500ml cuff Amplitude Range: 0 - 150% Amplitude Accuracy: better than 0.5% AutoSettings: unlimited, user definable Envelope Shift: ± 50 mmHg max Minimum Diastolic: 15 mmHg Maximum Systolic: 275 mmHg Manometer: Pressure Range: 0.0 to 400.0 mmHg Accuracy: ± 0.5 mmHg Resolution: 0.1 mmHg Regulated Pressure Source: Pressure Range: 10.0 to 400.0 mmHg Accuracy: ± 0.5 mmHg Resolution: 0.1 mmHg Leak Test: Automatic/manual Inflation Automatic Timer Leak Test Time: 30 - 600 seconds (user defined) Target Pressure: 20 to 400 mmHg Range: 0 to 200 mmHg/min User Definable Presets: 12 User Definable AutoSettings: unlimited OverPressure Test: Automatic/manual Inflation Range: 20 - 400 mmHg Release Time: 1-999 sec User Definable Presets: 12 User Definable AutoSettings: unlimited Standard Features / Accessories: • Autosequences • Unlimited User Defined Settings • vPad-A1 Power Base / Display • Universal Hose Adapter Kit Communication / User Interface: via vPad-A1 Base Unit Android 5" tablet: Touchscreen User Interface Wired (USB) or Bluetooth mode WiFi 16 GB memory Dual XBUS for Datrend test automation Power Supply: External AC adapter Internal rechargeable Li-Ion batteries (for 200+ simulations with full charge) Dimensions: 98mm x 275mm x 97mm (3.87" x 10.82" x 3.80") BP Unit (incl. A1 Base) Weight: 1080g (2.38lb) BP Unit (incl. A1 Base) 200g (0.44lb) wireless tablet interface Environment: 15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II Please contact the factory for the availability of other calibration tables, or visit our web site for updates at www.datrend.com All specifications subject to change without notice. DATRENDSystems Inc. 130 - 4020 Viking Way, Richmond BC, Canada V6V 2L4 Phone: 604.291.7747 • Fax: 604.294.2355 Toll-free (North America only): 800.667.6557 Email: [email protected] ©datrend systems inc. September 2016 Accuracy Matters • www.datrend.com

 Open the catalog to page 1

All Datrend Systems catalogs and technical brochures

  1. vPad-O2

    106  Pages

  2. venTest

    2  Pages

  3. vPad-Mini

    2  Pages

  4. vPad-ES2

    2  Pages

  5. vPad-Rugged2

    2  Pages

  6. Infutest Solo

    2  Pages

  7. vPad-RF

    2  Pages

  8. vPad-A1

    2  Pages

  9. vPad-IN

    2  Pages

  10. Model 20A

    1  Page

  11. FMS-3

    2  Pages

  12. AMPS-1

    2  Pages

  13. Oxitest Plus7

    2  Pages

  14. Phase 3 VLM

    2  Pages

  15. Phase 3

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.