Defibrillator / Transcutaneous Pacer / Transvenous Pacer Analyzer The vPad-DF is the world’s first and only all-in-one Analyzer capable of testing Defibrillators, AED’s, Transcutaneous Pacemakers, and Transvenous Pacemakers. With its high-power test load, it can continuously withstand tests at maximum energy with no delay between discharges. Built on the proven Vision-Pad (vPad) platform, vPad-DF joins a family of vPad analyzers, simulators, and other solutions to carry out accurate and efficient testing through its intuitive user interface. Its ability to test automatically with user-programmable sequences and integrate with other analyzers and CMMS systems is unmatched. When paired with the optional vPad-VLM (Variable Load Module), vPad-DF will enable users to test defibrillators in accordance with IEC 60601-2-4 requirements, in addition to testing the performance of both Transcutaneous and Transvenous Pacemakers. Advanced analyzer with easy-to-use interface, efficiency High-resolution tablet display (1920 x 1200) Provides both Manual and Automated test modes Built-in and user customizable auto-sequences Optional vPad-VLM (Variable Load Module) to test in accordance with AAMI DF-80 and IEC 60601-2-4 requirements Reversible Paddle Plate Adapters to test with either defib paddles or hands-free cables Test report capabilities (built-in option) Integration with vPad Electrical Safety Analyzers 1 Integration with CMMS Systems 2 Screenshare Ready (user controlled) for Training and Support Wired or wireless control Battery or power supply operation Backward-compatible with vPad-Check automation for Phase 3 Defib/Pacer Analyzer short learning curv
Open the catalog to page 1Scope Output Accuracy Matters
Open the catalog to page 2From ECG R-wave peak to start of defib discharge -3000 ms prior to R-wave to +3000 ms after R-wave -300 ms prior to R-wave to +300 ms after R-wave -166 ms prior to R-wave to +166 ms after R-wave 0 to +700 ms after R-wave 1 ms ±1 ms 10 to 360 BPM in 1 BPM steps Coarse, Fine 104 BPM 150 BPM 100 to 300 BPM in 5 BPM steps 10 to 360 BPM in 1 BPM steps Coarse, Fine 104 BPM Coarse, Fine 150 BPM 100 to 300 BPM in 5 BPM steps 5 types Flat line, random baseline Accuracy Matters
Open the catalog to page 40.1 to 9.9 Hz in 0.1 Hz steps 0.1 to 9.9 Hz in 0.1 Hz steps 0.03 to 9.99 Hz in 0.01 Hz steps 10.0 to 300.0 Hz in 0.1 Hz steps 10 to 10,000 ms in 1 ms steps Square, Haver-triangle, Haversine, Sine2, or ISO 5841-1 0.05 to 5.00 mV in 0.01 mV steps 10 to 360 BPM in 1 BPM steps 1 to 300 ms in 1 ms steps ±1 % Sine 50 or 60 Hz (± 0.015 Hz) 0 to 10 mV p-p in 0.2 mV steps (transcutaneous) 0 to 100 mV p-p in 2 mV steps (transvenous) ±2 % 0.1 to 2.5 ms in 0.1 ms steps ± 1 μs, all pulse widths 0 (off) and ±2 to ±700 mV in 1 mV steps ±2 % Accuracy Matters
Open the catalog to page 6Accuracy Matters
Open the catalog to page 8Normal Sinus Rhythm Underdrive (85%) or Overdrive (115%) Aux Rate (10 to 300 BPM) VVI, VOO (30 to 200 PPM) NSR 12-lead ECG, Ventricular Paced Rhythm 10 to 300 BPM 10 to 250 mA VVI, VOO (30 to 200 PPM)
Open the catalog to page 9Square, triangle, haversine, Sine2, ISO 5841-1 0.05 to 5 mV ±5 % 1 to 300 ms ±1% 0.0 to 10.00 mV ±2 % Sine 50/60 Hz ±0.015 Hz 12-lead ECG, ventricular load Accuracy Matters
Open the catalog to page 10Triangular, on both Atrial and Ventricular channel with P-R interval adjustable 50 – 250 ms Underdrive (85%) or Overdrive (115%) Aux Rate (10 to 300 BPM) Atrial/ventricular load (500 Ω), ECG signal VVI, AAI, DDD, VOO, AOO, DOO (30 to 200 PPM) Triangular, on both Atrial and Ventricular channel with P-R interval adjustable 50 – 250 ms 10 to 300 BPM 2 to 50 mA VVI, AAI, DDD, VOO, AOO, DOO (30 to 200 PPM) Square, triangle, haversine, sine2, ISO 5841-1 0.05 to 50 mV ±5 % 1 to 300 ms ±1 % 0.0 to 100.0 mV ±2 % Sine 50/60 Hz ±0.015 Hz Atrial/ventricular load, ECG Signal
Open the catalog to page 112 % of reading + 0.1 J with 25, 75 Ω though 200 Ω loads 1 % of reading + 0.1 J with 50 Ω load 1 % of reading + 2 V with 25 Ω and 50 Ω loads 2 % of reading + 2 V with 75 Ω through 200 Ω loads 2 % of reading + 0.1 A with 25 Ω load 1 % of reading + 0.1 A with 50 Ω through 200 Ω loads
Open the catalog to page 12Ordering Information Part Number / Descriptions 8000-160 - vPad-DF, Defib Tester, Universal 8000-180 - vPad-VLM, Variable Load Module Optional Upgrade 6960-015 - vPad-DF Transvenous Pacer Testing Option Optional accessories 7200-660 - Cable, vPad-DF Hands-Free, Physio 7200-661 - Cable, vPad-DF Hands-Free, Philips 7200-662 - Cable, vPad-DF Hands-Free, Zoll 7200-663 - Cable, vPad-DF Hands-Free, Anderson 7200-664 - Cable, vPad-DF Hands-Free, HP Barrel 7200-659 - Cable, vPad-DF Unterminated Hands-Free 7200-667 - Cable, vPad-DF Red Pacer 2 Cable 7200-668 - Cable, vPad-DF Black Pacer 2 Cable 7500-170...
Open the catalog to page 13About Datrend Systems Inc. Datrend Systems is a Canadian owned company focused on developing and manufacturing leading edge biomedical test instruments and solutions for hospitals, clinics, third party service organizations, and medical device manufacturers. Utilizing the latest technology, we develop comprehensive Preventive Maintenance Solutions to help our customers save time and money whilst meeting stringent regulatory requirements. Datrend Systems subscribes to the regulatory standards of CSA, EU, UL, and UKCA. Products are designed and manufactured following the quality system principles...