QualityMon O2

QualityMon O2

Product catalog summary
Overview: The QualityMon O2 is an oxygen monitor designed to measure and monitor the oxygen content in breathing gas mixtures, primarily for medical applications. It is not suitable for personal protection or for monitoring during the production of gas mixtures.
Specifications:
  • Gas Type: Oxygen
  • Measuring Range: 18% to 100% of Oxygen in gas mixtures
  • Response Time: Less than 12 seconds at 90% of the final value
  • Display Accuracy: Less than 1% vol. O2 when calibrated to 100% vol. O2
  • Linearity Error: Less than 3% relative deviation to the characteristic curve
  • Drift: Less than 1% vol. O2 over 8 hours
  • Cross-Sensitivity: Less than 0.1% vol. O2 in reaction to various residual gases
  • Alarm: Visual and audible with manual override and mute function
  • Operating Temperature: +5°C to +50°C (41°F to 122°F)
  • Dimensions: 61 x 110 x 92 mm
  • Weight: 500g with 3 AAA batteries
  • Compatible Sensor: OOM111 by EnviteC Wismar GmbH by Honeywell
  • Sensor Life Expectancy: Minimum over 1,000,000 O2% hours
Standards and Certification: The device meets international standards EN ISO 80601-2-55, DIN EN 60601-1, EN ISO 14971, and holds a CE-Certification 13485: 2016.
Applications: The QualityMon O2 is suitable for functional monitoring of breathing gas mixtures in anesthesia, respiratory protective devices, and baby incubators.
See more

Catalog excerpts

QualityMon O2-1

Oxygen monitor QualityMon O2 The QualityMon O2 oxygen meter is used to determine and monitor the oxygen concentration in breathing gas mixtures in medical applications. The Quality MonO2 can be used for the functional monitoring of the breathing gas mixtures of anesthesia and respiratory protective devices, as well as baby incubators. The oxygen content of a breathing gas shall be monitored. Not suitable for personal protection. Not for monitoring during the production of gas mixtures. Safety index for patients and users: Device: Fulfilled international standards: CE-Certification: Technical Data Gas Type Measuring range Response time Display accuracy Accuracy Linearity error Drift Cross –sensitivity Alarm Alarm limits Mute alarm function Autosetfunction Offset voltage Operating temperature DimensionsW*D*H Weight Compatibleoxygen sensor Expected use of sensor life 0482 Internal battery-operated device EN ISO 80601-2-55, DIN EN 60601-1, EN ISO 14971 13485: 2016 certification Specification Oxygen 18%up to100% of Oxygen in Gas mixtures < 12 sec. at 90% of final value <1% vol. O2, when calibrated to 100 % vol.O2 <1% vol. O2, when calibrated to 100 % vol. O2 < 3% relative deviation to the characteristic curve < 1% vol. O2 over 8 hours < 0.1% vol. O2 in reaction to: 10 % CO2 residual N2 80% N2O residual N2 7.5% Halothane residual N2 7.5% Isoflurane residual N2 7.5% Enflurane residual N2 9% Sevoflurane residual N2 20% Desplurane residual N2 Visual and audible alarm 18 –101% (manuell alarm override 16–103%) 30 sec. ±2% of the current measured value < 200 μV in 100 % nitrogen over 5 min +5C° -+50C° / 41°F –122°F 61 x 110 x 92 mm 500g with3* Typ AAAbatteries OOM111by EnviteC Wismar GmbH by Honeywell Minimum over 1,000,00

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.