Sapphire® plus - 1

Sapphire'plus QS Instructions for Use Dental Curing Light with Optional Tooth Whitening Conforms to AAMI Std ES60601-1, IEC Stds60601-1-6, 80601-2-60 & 62366 Certified to CSA Std C22.2 NO. 60601-1-6

Open the catalog to page 1
Sapphire® plus - 3

(Read completely through and understand before proceeding) Each statement in this section is a WARNING Before connecting and operating the Sapphire® Plus, read all safety instructions. When used properly and with normal precautions, the Sapphire Plus is a safe and effective multi-purpose light system. However, the Sapphire Plus is a source of high electrical voltage, intense light, and heat. This product must only be used as indicated in this manual. This device must be used by dental professionals such as a Board Certified Dentist or Orthodontist. The operator must have specific training...

Open the catalog to page 3
Sapphire® plus - 4

High-intensity dental lights should not be used by operators or on patients who may be photo-sensitive. Use extreme caution with patients who have any conditions which may make any portions of their mouths heat sensitive. Do not expose soft oral tissues directly to the tip of the light or to the light coming out of the tip at close proximity. Do not direct the light on to unprotected gingiva or skin. Even a portion of the beam directed on unprotected gingiva or skin can cause a rapid rise in temperature. While the reflective surface of the teeth is less prone to this effect, the condition...

Open the catalog to page 4
Sapphire® plus - 5

Electromagnetic Compatibility For Sapphire® Plus NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. Precautions for electromagnetic compatibility (EMC): Equipment in medical environments, including this product, can produce...

Open the catalog to page 5
Sapphire® plus - 6

Sapphire® Plus Specifications: Input Voltage: 100-240 Vac, 50/60 Hz, 6/3A Wavelength Range: 400-500 nm Light Intensity (curing mode): 1975 - 2320 mW/cm2 (at time of manufacturing) Dimensions:131.5 (W) x 234.9 (D) x 263.3 (H) (mm) Weight: 5.8 kg / 12.8 lbs Basic System Contents Sapphire Light Source Light Guide and Tip Curing Light Protective Glasses (2 pairs) Instruction for Use Manual AC Power Cord Composite Test Rings (200 each) Important Start-up/Set-up Instructions (Follow these instructions prior to using the system) 1. Attach the AC power cord to the power inlet on the Sapphire Plus...

Open the catalog to page 6
Sapphire® plus - 7

4. not saturate the controller keypad area with disinfectant or other liquids. Wipe the keypad surface with a Do damp towel and dry completely. 5. the lamp access plate on the bottom of the system is removed to check the lamp, make sure that it is If replaced correctly. The corners of the plate must fit into the outline of the cutout and not rest on top of the edges of the surrounding metal. When installed properly, all surface areas are even with each other. If the access plate is misaligned, the lamp will not activate. If the access plate is misaligned the system will show “Locd” meaning...

Open the catalog to page 7
Sapphire® plus - 8

Manufactured By: Den-Mat Holdings, LLC 1017 West Central Avenue, Lompoc, CA 93436 USA Toll Free 800-433-6628 • Tel. 805-346-3700 denmat.com Controls and Functions 1. Sapphire® Plus Light Source 2. Sapphire Plus Control Panel: Includes control wheel and select button, mode indicator, display windows, and start/stop button. 3. Sapphire Plus Display Windows: Digital display of the pre-set time, program mode, and light meter. 4. iber Optic Light Guide: Flexible optical device for delivery of light. Make sure the light guide is pushed all F the way in until it stops. 5. Sapphire Plus Curing...

Open the catalog to page 8
Sapphire® plus - 9

Control Panel UP/DOWN arrows on CONTROL WHEEL increase and decrease time and output. LEFT/RIGHT arrows on CONTROL WHEEL toggle between modes. START/STOP starts/stops lamp in CURE, BLEACH and SCOPE modes. SELECT accepts mode and setting. Eyewear WARNING: The protective eyewear provided with the Sapphire Plus must be worn by the operator, assistants, and patient when the system is in use. Additional eyewear is available through DenMat. Light Guide and Tips The light guide provided with the Sapphire Plus provides an optical path for transmitting light energy from the source to the tips. Both...

Open the catalog to page 9
Sapphire® plus - 10

The curing handpieces for the new Sapphire Plus light have been designed to work with existing Sapphire lights (reverse compatible). At the same time, curing handpieces for existing Sapphire lights can be used with the new Sapphire Plus light (forward compatible). Using Existing Sapphire Curing Handpieces When using existing curing devices in the Sapphire Plus, it is necessary to manually select the correct mode of operation. When the Sapphire Plus light is powered up without a device installed in the light guide receptacle on the front of the light source, the error message “No HP” (no...

Open the catalog to page 10
Sapphire® plus - 11

When the curing tip is placed on the light meter window and the lamp is activated, the system automatically detects the light and changes to the RADIOMETER mode. The output reading of power density is shown on the display windows. To get an accurate reading, the curing tip must be centered directly in the light meter window. Set the timer to 5 seconds and remove the tip from the light meter window as soon as you have a measurement. LIGHT METER WINDOW Proper Light Meter Readings • uring: If light is below 300 mW/cm2, do not use for curing. High light intensity (over C 1500 mW/cm2) allows...

Open the catalog to page 11
Sapphire® plus - 12

To change from the C1 to the C2 timer, press the LEFT and RIGHT arrow keys on the CONTROL WHEEL. The display windows show “C2” in the MODE window and “05” in the SETTING window. The ACTIVATOR button on the curing handpiece will activate and deactivate the lamp. The TIMER SELECT button on the curing handpiece will alternate between the C1 and C2 timers. ATTENTION: DO NOT HOLD DOWN THE START/STOP OR ACTIVATOR BUTTON. PRESS AND RELEASE. Fast Start (Note: See the Warnings and Precautions in this section and at the beginning of the Instructions for use Manual before curing) When the curing...

Open the catalog to page 12