Catalog excerpts
For Stable Internal Fixation of the Sternum Titanium Sternal Fixation System Surgical Technique
Open the catalog to page 1Introduction Titanium Sternal Fixation System 2 Surgical Technique Surgical Technique 7 Alternative Technique with Self-drilling Screws 17 Product Information Implants 21 MR Information The Titanium Sternal Fixation System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration or image artifact in the MR environment. The safety of the Titanium Sternal Fixation System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury. Titanium Sternal Fixation System Surgical Technique DePuy...
Open the catalog to page 2Titanium Sternal Fixation System. For stable internal fixation of the sternum. The DePuy Synthes Titanium Sternal Fixation System provides stable internal fixation of the sternum following a sternotomy or fracture of the sternum. Broad variety of sternal plates Different titanium plates according to the anatomical structures and patient’s need are available: – Straight locking plates for a stable sternal rib-to-rib fixation – Sternal body plates for minimal dissection – Star-shaped and H-shaped locking plates for fixation of the manubrium – Titanium sternal locking straight plate without...
Open the catalog to page 3Titanium Sternal Fixation System Self-tapping and self-drilling locking screws Both self-tapping and self-drilling locking screws are available to accommodate surgeon preference. Secure and stable locking The plate functions like an external fixator, applied internally. The screwhead of the sternal locking screws locks securely into the threaded plate hole to provide stable fixation (Figure1). Emergency release pin on all plates‡ All plates consist of two parts joined by a U-shaped release pin in the cross section. The release pin allows a quick and easy sternal reentry in cardiac emergency...
Open the catalog to page 4Titanium Sternal Fixation System Construct strength comparison Constructs loaded in tension in lateral direction. All tests were performed on 10 mm thick composite polyurethane foam test blocks composed of an inner (cancellous) core of 5 lb/ft 3 polyurethane foam with a 1.25 mm thick (cortical) shell of 10 lb/ft 3 polyurethane foam laminated to the exterior.* Sample test construct (X-Plate setup shown) 0 Cerclage wire (8x, 0.8 mm) Stainless steel X-Plates, wide (2x, 460.040) Titanium * echanical test data on file at DePuy Synthes. M Mechanical test results may not necessarily be indicative...
Open the catalog to page 5In 1958, the AO formulated four basic principles, which have become the guidelines for internal fixation.1,2 In 1958, the AO formulated four basic principles, which have become the guidelines for internal xation1, 2. Anatomic reduction Anatomic reduction Fracture reduction and fixation to Fracture reduction and xation to restore anatomical relationships. restore anatomical relationships. 1 Early, active mobilization Early, active Early and safe mobilization and Early and safe mobilization rehabilitation of the injured part rehabilitation of the injured and the patient as a whole. and the...
Open the catalog to page 6Primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum, to stabilize the sternum and promote fusion. Warnings: – hese devices can break during use (when subjected to T excessive forces or outside the recommended surgical technique). While the surgeon must make the final decision on removal of the broken part based on associated risk, in doing so, we recommend that whenever possible and practical for the individual patient, the broken part should be removed. – edical devices containing stainless steel which may M elicit an allergic reaction in...
Open the catalog to page 7Surgical Technique 1 Position patient Position the patient with the arms tucked along the sides. Avoid placing arms at 90° on arm boards, as this makes chest closure more difficult. 2 Debride (for secondary closure of the sternum) Remove existing wires. Debride the involved sternal edges until they are free of devitalized tissue and down to bleeding tissue. Hemostasis should be obtained. A curette can be used to remove any nonviable cartilaginous rib. Precautions: A sternal bone specimen should be sent to pathology to assess for osteomyelitis. Antibiotic treatment should be based on the...
Open the catalog to page 8Surgical Technique 3 Expose ribs laterally, if necessary Beginning medially, elevate the pectoralis major muscles with overlying soft tissue attached to create flaps and permit later approximation in the midline. It is usually not necessary to perform a second incision at the shoulder to release the pectoral muscle insertion. Following debridement and muscle elevation, pulse lavage the entire surgical site with an adequate volume of saline with antibiotics. 4 Measure sternal edges Instrument 319.11 Depth Gauge Using the depth gauge, measure the sternal edges adjacent to each rib where a...
Open the catalog to page 9Reduce sternum 398.903 Sternal Reduction Forceps, angled, with teeth 398.985 Bone Reduction Forceps, large Reduce the sternum using reduction forceps on both the superior and inferior aspects of the sternum. When placing the forceps, care should be taken to avoid the intercostal and mammary vessels and nerves. Alternative instrument 398.902 Sternal Reduction Forceps Note: Sternum can also be reduced with stainless steel surgical wire, if desired. Precaution: Avoid direct contact of stainless steel wires with titanium implants to avoid galvanic corrosion. 6 Cut and contour template...
Open the catalog to page 10Surgical Technique 7 Select plate Instrument 319.990 Plate and Rod Cutter Select the appropriate length titanium sternal locking plate. Center the release pin on the sternum with sufficient plate length on each side to allow a minimum of four locking screws on each side. The plate can be cut to length, if necessary, using the plate and rod cutter. Alternative instrument 391.967* Shortcut Plate Cutter (2 required) As an alternative, shortcut plate cutters may be used. Note: All steps of preparation and implantation of the Sternal Locking Plate have to be done, whenever possible, with the...
Open the catalog to page 11All Depuy Synthes catalogs and technical brochures
-
2.0 mm LCP® Distal Ulna Plate
20 Pages
-
Building on Success
16 Pages
-
RADIUS OF CURVATURE
3 Pages
-
Introducing The Variable Angle
12 Pages
-
HEALIX Anchor™ 3.4 mm
2 Pages
-
Small Battery Drive II
4 Pages
-
HEALIX ADVANCE
4 Pages
-
3.5 mm LCP™ Medial
15 Pages
-
MatrixRIB®FixationSystem
86 Pages
Archived catalogs
-
2.4 mm VA LCP™
4 Pages
-
Mandible Trauma Solutions
2 Pages
-
Power line II
4 Pages
-
Concorde
28 Pages
-
LCP Intercarpal
31 Pages
-
LCS® COMPLETE™
2 Pages
-
Synthes TPLO.
8 Pages
-
SynFix-LR System
56 Pages
-
ATB Anterior Tension Band Plate
32 Pages
-
CONDUIT™
15 Pages
-
Brochure_FINAL
2 Pages
-
DePuy Synthes
81 Pages
-
Anspach
3 Pages
-
Orthopedic Foot Instruments
32 Pages
-
PINNACLE® Hip Solutions
12 Pages
-
Corail
24 Pages
-
S-ROM® NOILES™
68 Pages
-
TRI-LOCK® Product Rationale
12 Pages
-
Reclaim Surgical Technique
44 Pages
-
Speed
2 Pages
-
attune
80 Pages
-
HAMMERLOCK® 2
2 Pages
-
DePuy Glenoid Solutions
2 Pages
-
Trauma Solutions. Elbow
4 Pages
-
Polar
4 Pages
-
Alveolar Distractor.
4 Pages
-
Piezoelectric System
4 Pages
-
Air Power Line II
6 Pages
-
LCP Clavicle Hook Plate
4 Pages
-
TruMatch Pin Guides
16 Pages
-
P F N A
8 Pages
-
SKILL, DEDICATION,
16 Pages
-
Orthopaedics. Overview
20 Pages
-
DURALOC
16 Pages
-
Marathon Cemented Cup
20 Pages
-
REEF Surgical Technique
16 Pages
-
MatrixNEURO
8 Pages
-
Anspach XMax
4 Pages
-
Anspach eMax 2 Plus
4 Pages
-
Small Electric Drive
4 Pages
-
Air Pen Drive
4 Pages
-
Colibri II
4 Pages
-
Spine
25 Pages
-
Expert Hindfoot Arthrodesis Nail
48 Pages
-
LCP Distal Fibula Plates
32 Pages
-
TomoFix
60 Pages
-
Expert Tibial Nail PROtect
16 Pages
-
Expert Tibia Nail
84 Pages
-
Sacral Bars
16 Pages
-
Pelvic C-Clamp
20 Pages
-
Low Profile Pelvic System
16 Pages
-
Proximal Femoral (Hook) Plate
24 Pages
-
LCP
24 Pages
-
PFNA
112 Pages
-
HCS 1.5, 2.4, 3.0
36 Pages
-
LCP Wrist Fusion
32 Pages
-
LCP Compact Hand
28 Pages
-
VA-LCP Elbow
48 Pages
-
Distal Radius
44 Pages
-
Olecranon
30 Pages
-
LCP Hook Plate
28 Pages
-
DHP & Olecranon
4 Pages
-
LCP S-A
4 Pages
-
Epoca
4 Pages
-
Philos
32 Pages
-
MultiLoc
68 Pages