SensoStar GLHsix

SensoStar GLHsix

SensoStar GLHsix

Product catalog summary
Overview: This document provides technical specifications and features of a maintenance-free biosensor designed for simultaneous measurement of glucose, lactate, and hemoglobin from whole blood samples. It emphasizes ease of use and efficiency, making it suitable for small laboratories and field testing.
Technical Specifications:
  • Measuring Range:
    • Glucose: 0.6 – 50 mmol/L (11 – 910 mg/dL)
    • Lactate: 0.5 – 30 mmol/L (4.5 – 270 mg/dL)
    • Hemoglobin: 1.86 – 15.5 mmol/L (3 – 25 g/dL)
  • Measuring Time:
    • Single/dual measurement for glucose and/or lactate: 40 seconds
    • Single measurement for hemoglobin: 90 seconds
    • Combined measurement for glucose/lactate and hemoglobin: 90 seconds
  • Sample Volume: 10 µL
  • Sample Types: Whole blood, serum, plasma
  • Precision (CV for 20 samples in series):
    • Glucose (12 mmol/L): < 1.5%
    • Lactate (10 mmol/L): < 2.0%
    • Hemoglobin (8 mmol/L): < 2.0%
  • Dimensions: 200 mm (W) x 150 mm (D) x 170 mm (H)
  • Weight: 2 kg
Features and Benefits:
  • Utilizes biosensor technology for glucose and lactate and photometric measurement for hemoglobin.
  • Offers high specificity, accuracy, and precision with a short time to result.
  • Features automatic calibration and extended sensor operation time.
  • Provides individual glucose plasma values from whole blood samples.
  • Facilitates easy sample collection with a flexible sample matrix.
  • Includes an optional sample barcode reader and a slider for series measurement of up to six samples.
  • Equipped with an EDP interface and printer, and is ideal for small laboratories and field testing with an optional battery.
See more

Catalog excerpts

SensoStar GLHsix-1

CdLH £>ix • Simultaneous Measurement • Maintenance Free Biosensor • Individual Glucose Plasma Value from Whole Blood CHOOSING QUALITY.

 Open the catalog to page 1
SensoStar GLHsix-2

Efficient operation combined with convenient use · Biosensor technology for glucose and lactate; photometric measurement of hemoglobin - High specificity - High accuracy - Excellent precision - Short time to result - Extended sensor operation time · Automatic calibration · Individual glucose plasma value from whole blood samples · Easy sample collection · Flexible sample matrix · Optional sample barcode reader · Slider for measurement in series up to six samples · Ideal for small laboratories and field testing (battery optionally available) · EDP interface and printer Technical specifications...

 Open the catalog to page 2

All DiaSys Diagnostic Systems GmbH catalogs and technical brochures

  1. respons®910

    4  Pages

  2. DiaSys

    104  Pages

  3. HbA1c net FS

    6  Pages

  4. InnovaStar®

    6  Pages

  5. respons®920

    4  Pages

  6. Cystatin C FS

    4  Pages

  7. Electrolytes

    4  Pages

  8. D-Dimer FS

    4  Pages

  9. oneHbA1c FS

    4  Pages

  10. CRP U-hs

    4  Pages

  11. Lactate FS

    2  Pages

  12. Prealbumin FS

    2  Pages

  13. LP-PLA2

    4  Pages

  14. Ethanol FS

    2  Pages

  15. CRP FS

    4  Pages

  16. CK-MB FS

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.