Cube 80

Cube 80

Cube 80

Product catalog summary
Technical Specifications:
The system operates on 230Vac@50Hz in Europe and 110-120Vac@60Hz in the USA/Canada, with an absorbed electric power of 265VA. It includes a fuse block with 2 x 5.0 AT delayed fuses. The dimensions are 650 x 570 x 690 mm, and it weighs 60 kg. The operating room temperature ranges from +15 to +35°C, with storage up to 45°C, and allowable relative humidity from 20% to 80% without condensation.

Central Unit and Display:
The central unit features an Intel XScale PXA 255 microprocessor with 32 MB FLASH and 64MB SDRAM. It has a TFT 800x600 color touch screen display.

Control and Analytic Section:
The control unit includes a microprocessor card on an owner bus, and the internal analytic section has an 89-position chain for test tubes. It uses two pairs of optic-electronic elements (LED & analogical sensor).

Printer and Interface:
The system includes an alphanumeric printer with thermal paper, printing 36 characters per line at a speed of 20mm/sec. Interfaces include 2 x RS232C, 2 USB Host, 1 USB Client, and 1 Slot Compact Flash.

Safety and Compliance:
The device is classified under Protection Category CLASS I, complies with EN61010-1 for device security, and EN61326-1 for EMC. It falls under Installation Category II.

Consumables and Accessories:
The system uses a dedicated tube with a transponder RF for charging with the exact number of tests, available in 10,000, 5,000, and 1,000 test options. ESR Control Cubes are available in different configurations.

System Features:
The analyzer is fully automated for determining erythrocyte sedimentation rate (ESR) in primary EDTA tubes. It supports various models and brands of EDTA tubes, allowing simultaneous processing without restrictions. The system ensures high productivity with up to 95 results per hour and continuous loading of sample holder racks. It provides automated mixing of samples and delivers the first result in 22 minutes.

User Interface and Safety:
The touch screen interface is user-friendly and intuitive, offering complete traceability of samples. The system minimizes operator risk by eliminating contact with blood samples and waste fluid production. It automatically sorts ESR and non-ESR samples through a bi-directional connection to the laboratory information system.

Conclusion:
The system is designed as a walk-away solution, fitting the needs of modern clinical laboratories by reducing operator time, improving data handling safety, and minimizing bio-contamination risks.
See more

Catalog excerpts

Cube 80-1

walk-away system mcccE UIEOOE TECHNICAL CHARACTERISTICS < CL < CO o o CURRENT ABSORBED ELECTRIC POWER FUSE BLOCK DIMENSIONS WEIGHT ROOM TEMPERATURE ALLOWABLE RELATIVE HUMIDITY CENTRAL UNIT DISPLAY CONTROL UNIT PERIPHERALS INTERNAL ANALYTIC SECTION OPTIC GROUPS PRINTER INTERFACE PROTECTION CATEGORY SECURITY OF THE DEVICE EMC INSTALLATION CATEGORY CATALOG NUMBER Europe 230Vac@50Hz, Usa/Canada 110-120Vac@60Hz 265VA 2 x 5.0 AT (Delayed) (5 x 20 mm) UL 650 x 570 x 690 mm ( l x h x d ) 60 Kg In exercise from +15 to +35°C, Warehouse up to 45°C From 20% to 80% without condensation Microprocessor Intel XScale PXA 255 32 MB; FLASH 64MB SDRAM TFT 800x 600 colour with Touch Screen Microprocessor card on owner bus 89 position chain for the appropriate test tube Two couples of optic-electronic elements (Led & analogical sensor) Alphanumeric with thermal paper wide 58mm, 36 characters per line, speed 20mm/sec. 2 x RS232C, 2 USB Host, 1 USB Client, 1 Slot Compact Flash CLASS I EN61010-1 EN61326-1 II 10380 FULLY AUTOMATED BENCH-TOP ANALYZER FOR THE DIRECT DETERMINATION OF ERYTHROCYTE SEDIMENTATION RATE (ESR) IN PRIMARY EDTA TUBES CONSUMABLES & CHECK DEVICE TRANSPONDER RF (Dedicated tube with trasponder RF to be used to charge the VES-Matic Cube System with the exact number of tests) 10.000 tests 5.000 tests 1.000 tests ESR CONTROL CUBE 4 x 9 mL 2 x 9 mL Normal - 2 x 9 mL Abnormal ESR CONTROL CUBE 2 x 9 mL 1 x 9 mL Normal - 1 x 9 mL Abnormal niECCE DIESSE Diagnostica Senese SpA Via del Pozzo, 5 Loc. San Martino 53035 Monteriggioni (SI] Italy Tel: +39 0577 319560/61 Fax: +39 0577 318763 www.diesse.it [email protected] UICOOC DIAGNOSTIC SYSTEMS

 Open the catalog to page 1
Cube 80-2

and versatility different models and brands of EDTA tubes can be processed simultaneously by the instrument, without restrictions suitable for laboratories receiving samples from different collection points greater sample stability: the test can be performed within up to 24 hours from blood collection continuous loading of the sample holder racks for a better work-flow in the laboratory automated mixing of samples ensuring the complete disaggregation of erythrocytes first result obtained in 22 minutes only HIGH PRODUCTIVITY up to 95 nesults/houn FEATURES THE SYSTEM IS REALLY WALK-AWAY AND PERFECTLY...

 Open the catalog to page 2

All Diesse Diagnostica Senese catalogs and technical brochures

  1. RackSmart

    2  Pages

  2. Cube 200

    2  Pages

  3. ING CubeTrack

    2  Pages

  4. MINI-CUBE

    2  Pages

  5. ChorusTrio

    2  Pages

  6. Chorus Line

    2  Pages

  7. AUTODAT

    2  Pages

  8. ESR Line

    2  Pages

  9. CUBE30touch

    2  Pages

  10. VACU-CODE

    2  Pages

  11. Catalog 2017

    54  Pages

  12. VES-MATIC 30

    2  Pages

  13. VES-MATIC 20

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.