ADT 672 series

ADT 672 series

ADT 672 series

Product catalog summary
Overview
The ADT672 series precision pressure calibrators utilize advanced microprocessor technology and silicon pressure sensors to provide accurate pressure calibration for gauges, transmitters, and switches. These calibrators are portable and can measure pressure, read current or mV from transducers, and supply excitation voltage for sensor calibration. They achieve 0.025% full-scale accuracy up to 10,000 psi and 0.1% full-scale accuracy up to 36,000 psi.
Features and Applications
  • Gauge calibration and precision pressure measurement
  • Calibration of pressure transmitters and testing of pressure switches
  • Testing of safety relief valves and pressure regulators
  • Hydraulic leak testing
  • Pressure measurement uncertainty: 0.025%FS or 0.05%FS
  • Pressure ranges up to 36,000 psi
  • Measurement of mA or V with high accuracy
  • 24V loop supply for powering transmitters during tests
  • Hart Communication capability and advanced temperature compensation
  • Data logging and dual readout with Min/Max/Hold functions
  • Large, backlit display with 6-digit resolution
  • Rechargeable battery or AC adapter power options
  • NIM traceable calibration certificate available
Specifications
  • Accuracy: 0.025% or 0.05% of full scale
  • Gauge types: Gauge, Compound, Absolute, Differential pressure
  • Pressure ranges: Various, up to 36,000 psi (2,500 bar)
  • Media compatibility: Gas and liquid, depending on pressure range
  • Burst pressure: Varies by model, typically 3X rated pressure
  • Electrical connections: 0.156 inch sockets
  • Data logger: 30 files, 40 records per file, manual and automatic modes
  • Power: Rechargeable Li-ion battery, 40-hour working time, 4-hour recharge
  • Enclosure: Aluminum alloy case, 316L SS wetted parts
  • Communication: RS232 interface
Environmental Conditions
  • Compensated and operating temperature: 14°F to 122°F (-10°C to 50°C)
  • Storage temperature: -4°F to 158°F (-20°C to 70°C)
  • Humidity: Less than 95%
Ordering Information
  • Model numbers and optional accessories available
  • Pressure port types: NPT, Autoclave, or custom
See more

Catalog excerpts

ADT 672 series-1

APPLICATIONS ·Gauge calibration ·Precision pressure measurement ·Pressure transmitters calibration ·Pressure switch testing ·Safety relief valve testing ·Pressure regulator testing ·Hydraulic leak testing SPECIFICATIONS Accuracy 02: 0.025% of full scale 05: 0.05% of full scale OVERVIEW With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the ADT672 series precision Gauge types Gauge pressure Compound Pressure pressure calibrators provide a pressure calibration solution for Absolute pressure gauges, transmitters, and switches in a wide pressure range. Differential pressure In a highly portable package, it can not only measure pressure precisely with a built-in pressure sensor, but can also read the current or mV produced by the transducer, and can even supply an excitation voltage to power sensors or transmitters during 36,000 psi (2,500 bar), every silicon pressure sensor in gauges has been specially aged, tested and screened before assembly. ADT672 series precision pressure calibrators are unmatched in performance and reliability. FEATURES ·Pressure measurement uncertainty: -02: 0.025%FS -05: 0.05%FS ·Pressure ranges to 36,000 psi (2500 bar) ·Measure mA or V with 0.02% RD + 0.003%FS accuracy calibration. To reach 0.025% F. S. accuracy over ranges to 10,000 psi (700 bar) and 0.1% F. S. accuracy over range to Pressure Pressure ·Power transmitters during test using 24V loop supply ·Pressure switch test ·Hart Communication capability ·Advanced temperature compensation ·Dual readout ·Min/Max/Hold to capture changing measurements ·Data logging ·Large, easy to read display with 6-digit resolution ·Back lighted display ·Rechargeable battery or AC adapter ·NIM traceable calibration certificate (optional) Note: [1]. Sealed gauge pressure for above 1000 psi [2]. G=Gas, L=Liquid [3]. 0.025% FS

 Open the catalog to page 1
ADT 672 series-2

Absolute Pressure Pressure Pressure Media Range Differential Pressure Pressure Pressure Media Range Range Compound Pressure Pressure Pressure Media Range Range Electrical Measurement and Source Accuracy Range 24V±0.5V max:50mA Protect at: 120mA Status OPEN/CLOSED Pressure Port ,, 1 4 NPT(≤15,000psi) Autoclave F-250-C, 9/16" - 18 UNF female or Autoclave M-250-C, 9/16"-18 UNF male(>15,000psi) 0.156 inch (4mm) test hose (for differential pressure) Other connections available per request Electrical Connection 0.156 inch (4mm) sockets Over Pressure Warning: 120% Data Logger Storage capacity: 30 files,...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.