modula
1 / 8Pages

Catalog excerpts

modula - 1

[ modula ] Central Reverse Osmosis System [ modula TP ] Double-stage Central Reverse Osmosis System

Open the catalog to page 1
modula - 2

The modular concept for permeate production [ modula ] Effective and reliable The modula reverse osmosis was developped to incorporate the proven and reliable DWA standard of dialysis water treatment into a flexible design concept. Based on an economic, cost effective and adaptable system designed to comply with different international dialysis requirements and regulations. The modula uses high quality materials and superior manufacturing to fulfill the objective to achieving maximum hygiene, optimum safety, long life and reliability of permeate production – with guaranteed economy through...

Open the catalog to page 2
modula - 3

Key features • Compact design / modular construction • Hygienic stainless steel tubing and frame • Water saving technology /  Permeate recirculation • Microprocessor controlled & monitored • Autoflush system • Secure inlet valve control • 3-way valve for permeate to ring or drain • Automatic on / off (two times per day) • sampling and disinfection points 2 (raw water and permeate) • 4-button control panel • Back-lit display • Monitoring parameters (permeate and concentrate flow, membrane pressure, raw water / permeate pressure, permeate temperature, permeate and raw water conductivity) •...

Open the catalog to page 3
modula - 4

Central double-stage Reverse Osmosis System [ modula TP ] By the double-stage setup the modula TP provides highest permeate quality even at a low feed water quality. As with the modula system the modula TP is designed to comply with different international dialysis requirements and regulations. Through two independent stages the modula TP provides for total operation safety. With one stage out of service the other can take over the complete duty to provide continuation of treatment. For heat disinfection of ring mains, media supply systems and dialysis machines modula TP can be easily...

Open the catalog to page 4
modula - 5

• Increased chemical and microbiological permeate quality through double-stage operation • Water Saving Technology through permeate recirculation • 5,7" multicolour touch screen • Compact design • Semi-automatic disinfection programmes • Energy-saving polyamide membranes • Permeate pressure sustaining valve for system pressure settings • Internal surface of passivated stainless steel piping with clamp connectors • Continuous monitoring of conductivity, pressure, temperature and permeate flow • Programmable shut on / off • Integrated self-test at power up • Emergency mode in case of...

Open the catalog to page 5
modula - 6

SpecificationPerformance Data Water Connections Electrical Data Dimensions (W x H x D) Weight (max.) Maximum dialysis water production capacity at 10 °C and with a back pressure of 2.5 bar Dialysis water pressure Retention rate System recovery Soft water feed Dialysis water feed Dialysis water return Mains plug Supply voltage 2 - 5 bar > 99 % NaCl up to 75 % Hose nipple for hose DN 20 Hose nipple for hose DN 20 Power consumption (per phase) 7 A (400 V-Versions) Electrical Safety Feed Water Ambient Conditions On-site Requirements Energy consumption Fuse Safety class Overvoltage category...

Open the catalog to page 6
modula - 7

SpecificationPerformance Data Water Connections Electrical Data Dimensions (W x H x D) Weight (max.) Maximum dialysis water production capacity (at 10 °C and 2,5 bar back pressure) Dialysis water pressure Retention rate System recovery Soft water feed Dialysis water feed Dialysis water return Mains plug Supply voltage 2 - 5 bar > 99 % NaCl Up to 85 % Hose nipple for hose DN 25 Hose nipple for hose DN 20 or DIN 11851 - DN 15 IG Electrical Safety Feed Water Power consumption (per phase) Energy consumption Fuse Safety class Overvoltage category Degree of pollution Protection class Quality...

Open the catalog to page 7
modula - 8

DWA GmbH & Co. KG Großer Sand 8 76698 Ubstadt-Weiher Phone: +49 7251 6900-0 Fax: +49 7251 6900-115 info@dwa-online.com www.dwa-online.com The Dialysis Water Speci

Open the catalog to page 8