Product catalog summary
Overview: The VAMP Closed Blood Sampling System is designed for venous and arterial blood management protection. This guide provides a quick reference for priming, drawing clearing volume, and obtaining blood samples using the system.
Priming Procedure:
  • Squeeze and hold the flexures to slightly raise the plunger, facilitating the flow of the priming solution.
  • Ensure the shut-off valve is open (parallel to the tubing) and hold the sample site above the VAMP reservoir at a 45° angle.
  • Initiate flow by pulling the Snap-Tab of the TruWave disposable pressure transducer.
  • Slowly deliver the priming solution to remove air, then close the plunger and connect to the patient’s catheter.
Drawing the Clearing Volume:
  • Squeeze the flexures firmly and slowly pull up the reservoir plunger over 3-5 seconds until it reaches its 5 mL capacity.
  • Close the shut-off valve by turning the handle perpendicular to the tubing.
Blood Sampling from Z-site Needleless Sample Site:
  • Swab the sample site and insert a cannula with a syringe or direct-draw unit.
  • If using a direct-draw unit, insert a vacuum tube and draw the required blood volume.
  • After the last sample, remove the vacuum tube first, then pull out the VAMP direct-draw unit.
  • Open the shut-off valve by turning the handle parallel to the tubing.
  • Push down on the plunger smoothly over 3-5 seconds until the flexures lock in place and all fluid is reinfused.
  • Flush the system clear by pulling the Snap-Tab on the TruWave transducer and swab the sample site to remove excess blood.
Caution: This device is restricted to sale by or on the order of a physician. Refer to the full instructions for use for detailed prescribing information, including indications, contraindications, warnings, precautions, and adverse events.
Trademarks: Edwards, Edwards Lifesciences, the stylized E logo, TruWave, and VAMP are trademarks of Edwards Lifesciences Corporation.
See more

Catalog excerpts

VAMP Closed Blood Sampling System-1

VAMP Closed Blood Sampling System Venous Arterial blood Management Protection system Quick Reference Guide Priming 1. queeze and hold flexures to S slightly raise plunger to facilitate flow of priming solution. 2. With the shut-off valve in the open position (parallel to the tubing), hold sample site above the VAMP reservoir at a 45° angle. 3. Provide flow by pulling the Snap-Tab of the TruWave disposable pressure transducer. 4. Slowly deliver priming solution to remove air. Close plunger and connect to patient’s catheter.

 Open the catalog to page 1
VAMP Closed Blood Sampling System-2

Drawing the clearing volume 5. irmly squeeze the flexures and F slowly (over 3-5 seconds) pull up the reservoir plunger until the reservoir has reached its 5 mL volume capacity. 6. lose shut-off valve by turning C the handle perpendicular to tubing. Drawing blood samples from Z-site needleless sample site 7. Swab the sample site. Push cannula with syringe or direct-draw unit into sample site. 8. f using direct-draw unit, put vacuum I tube into the unit. Draw the required volume of blood for a sample. Repeat to complete blood study requirements.

 Open the catalog to page 2
VAMP Closed Blood Sampling System-3

9. After the last sample has been drawn, remove the vacuum tube first, then grasp the VAMP direct-draw unit by the cannula and pull straight out. 10. pen shut-off valve by turning the O handle parallel to tubing. S 11. moothly and evenly over 3-5 seconds, push down on the plunger until the flexures lock in place in the fully closed position and all fluid has been reinfused into the line. F 12. lush the VAMP system clear by pulling the Snap-Tab on the TruWave transducer and swab the sample site ensuring removal of any excess blood.

 Open the catalog to page 3
VAMP Closed Blood Sampling System-4

CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards, Edwards Lifesciences, the stylized E logo, TruWave, and VAMP are trademarks of Edwards Lifesciences Corporation. All other trademarks are property of their respective owners. © 2017 Edwards Lifesciences Corporation. All rights reserved. PP--US-2012v1.0 Edwards Lifesciences • edwards.com One Edwards Way, Irvine CA 92614 USA

 Open the catalog to page 4

All Edwards Lifesciences catalogs and technical brochures

  1. EVOQUE

    7  Pages

  2. PASCAL

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.