VAMP Closed Blood Sampling System Venous Arterial blood Management Protection system Quick Reference Guide Priming 1. queeze and hold flexures to S slightly raise plunger to facilitate flow of priming solution. 2. With the shut-off valve in the open position (parallel to the tubing), hold sample site above the VAMP reservoir at a 45° angle. 3. Provide flow by pulling the Snap-Tab of the TruWave disposable pressure transducer. 4. Slowly deliver priming solution to remove air. Close plunger and connect to patient’s catheter.
Open the catalog to page 1Drawing the clearing volume 5. irmly squeeze the flexures and F slowly (over 3-5 seconds) pull up the reservoir plunger until the reservoir has reached its 5 mL volume capacity. 6. lose shut-off valve by turning C the handle perpendicular to tubing. Drawing blood samples from Z-site needleless sample site 7. Swab the sample site. Push cannula with syringe or direct-draw unit into sample site. 8. f using direct-draw unit, put vacuum I tube into the unit. Draw the required volume of blood for a sample. Repeat to complete blood study requirements.
Open the catalog to page 29. After the last sample has been drawn, remove the vacuum tube first, then grasp the VAMP direct-draw unit by the cannula and pull straight out. 10. pen shut-off valve by turning the O handle parallel to tubing. S 11. moothly and evenly over 3-5 seconds, push down on the plunger until the flexures lock in place in the fully closed position and all fluid has been reinfused into the line. F 12. lush the VAMP system clear by pulling the Snap-Tab on the TruWave transducer and swab the sample site ensuring removal of any excess blood.
Open the catalog to page 3CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards, Edwards Lifesciences, the stylized E logo, TruWave, and VAMP are trademarks of Edwards Lifesciences Corporation. All other trademarks are property of their respective owners. © 2017 Edwards Lifesciences Corporation. All rights reserved. PP--US-2012v1.0 Edwards Lifesciences • edwards.com One Edwards Way, Irvine CA 92614 USA
Open the catalog to page 4All Edwards Lifesciences catalogs and technical brochures
-
EVOQUE
7 Pages
-
PASCAL
4 Pages
-
VAMP adult system
5 Pages
-
PediaSat Oximetry Catheter
5 Pages
-
OptiSite
2 Pages
-
Edwards Fogarty® Catheters.
3 Pages
-
Vascular Therapies
28 Pages
-
ev 1000 brochure
5 Pages
-
ED MagnaEase Bro
3 Pages
-
MagNa PerForMaNce MiTraleaSe
4 Pages
-
VAMP Adult Brochure
4 Pages
-
Swan-Ganz Brochure
6 Pages
-
ccNexfin System
3 Pages
-
PediaSat
4 Pages
-
Sepsis_management
4 Pages
-
Global Myocardial Protection
4 Pages
-
Edwards Protection Cannulae
119 Pages
-
Fluid Optimization
3 Pages
-
Flo Trac
4 Pages
-
Retrogradebrochure
4 Pages
-
Venous Cannulae
4 Pages
-
Arterial Cannulae
4 Pages
-
EV1000 Clinical Platform
5 Pages
-
PediaSat Catheter
4 Pages
-
PreSep Catheter
6 Pages
-
VolumeView Set
4 Pages
-
FloTrac Sensor
4 Pages
-
Product catalog
28 Pages