1. Catalogs
  2. EKF Diagnostics
  3. Hemolysis QC: Datasheet (EN EU Rev 1.0 05-2025)

Hemolysis QC: Datasheet (EN EU Rev 1.0 05-2025)

Hemolysis QC: Datasheet (EN EU Rev 1.0 05-2025)

Hemolysis QC: Datasheet (EN EU Rev 1.0 05-2025)

Product catalog summary
Overview: The document provides information on the Hemolysis QC system designed for precise measurement of free hemoglobin in the supernatant of stored red cell concentrates (RCC). It highlights the system's efficiency, accuracy, and practical application in estimating hemolysis levels.
Specifications:
  • Measurement Range: 0.3 - 30.0 g/L
  • Precision: CV < 2% for Hb ≥ 5 g/L; SD < 0.2 g/L for Hb < 5 g/L
  • Calibration: Factory calibrated against HiCN reference method
  • Sample Volume: 20 μL
  • Operating Temperature: 15 °C to 35 °C
  • Cuvette Storage Temperature: 2 °C to 50 °C
  • Dimensions: 150 x 90 x 40 mm
  • Weight: 135 g
Features:
  • Fast results within one to two seconds
  • Reagent-free microcuvettes with a long shelf life of 30 months
  • Automatic self-check and maintenance-free operation
  • Photometric measurement with scattering compensation
  • Connectivity: USB connection to PC
Components:
  • Hemolysis QC Analyzer
  • Hemolysis QC Test Cuvette
  • Hemolysis QC Control Pack
Usage Instructions: Collect the sample, insert the filled cuvette, press down gently, and remove the cuvette quickly after the "UP" signal appears. The result will display within seconds.
Additional Information: The document is intended for general information, and the product should be used according to the provided instructions/manuals. The manufacturer is EKF-diagnostic GmbH, located in Barleben, Germany.
See more

Catalog excerpts

Hemolysis QC: Datasheet (EN EU Rev 1.0 05-2025)-1

Hemolysis QC Precise free hemoglobin measurement of supernatant of RCC Test supernatant of stored red cell concentrate Simplifies the estimation of the degree of hemolysis Reagent-free microcuvettes Scan here to view product webpage

 Open the catalog to page 1
Hemolysis QC: Datasheet (EN EU Rev 1.0 05-2025)-2

Results in one-two seconds Linear measuring range: 0.3 - 30.0 g/L Precision: CV: < 2 % for Hb > 5 g/L; SD: < 0.2 g/L for Hb < 5 g/L Factory calibrated against HiCN reference method Photometric, broad spectrum with compensation of scattering • Automatic self-check between every measurement and maintenance free • Two levels of liquid control are available • Long microcuvette shelf life: 30 months from date of production even after opening the bag • Operating ambient temperature: 15 °C to 35 °C • Cuvette storage temperature: 2 °C to 50 °C • Operation by power supply or computer • Hemolysis QC Test...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.