Introduction
The document discusses the use of the rapid MICRO N cube for nitrogen analysis in various polymers. This instrument measures low-level nitrogen concentrations with a detection limit of 10 mg/kg using the Dumas method, which is matrix-independent and does not require matrix-dependent calibration.

Instrument and Methodology
The rapid MICRO N cube is used for measuring nitrogen concentrations. Solid polymer samples, each weighing approximately 20 mg, are prepared by weighing into tin boats and pressing to remove air. The Dumas method involves reducing nitrogen oxides to N2, which is then measured using a thermal conductivity detector.

Sample Analysis
Four types of polymer samples were analyzed:
  • Polypropylene granulate: 122 mg/kg nitrogen, SD 13 mg/kg, RSD 11%
  • Polyethylene granulate: 1985 mg/kg nitrogen, SD 25 mg/kg, RSD 1%
  • Polyethylene foil: 56 mg/kg nitrogen, SD 4 mg/kg, RSD 7%
  • HDPE granulate: 342 mg/kg nitrogen, SD 13 mg/kg, RSD 4%
The results demonstrate the instrument's suitability for analyzing nitrogen concentrations in polymers with high precision.

Conclusion
The rapid MICRO N cube is effective for nitrogen analysis in polymer samples, offering high sensitivity and matrix-independent results, making it ideal for the polymer industry.
See more

Catalog excerpts

MICRO N cube-1

Application note Nitrogen analysis in various polymers with the rapid MICRO N cube The rapid MICRO N cube measures low level nitrogen concentrations with a detection limit of 10mg/kg using the Dumas method. After reduction of the oxides of nitrogen the resulting N2 is measured using a thermal conductivity detector. Until now such low level concentrations were measured via oxidation to NO followed by chemiluminescence detection. This method is, however, matrix dependent and needs a matrix dependent calibration. Solid polymer samples have been weighed into tin boats. After closing the sample they have been pressed with a manual pressing tool in order to remove remaining air from the sample. The sample weight of the polymer samples was about 20 mg. instrument: rapid MICRO N cube sample: 20 mg polypropylene granulate polyethylene granulate HDPE granulate polyethylene foil The results show that the rapid MICRO N cube is very well suitable for the analysis of nitrogen concentrations in polymer samples. Nitrogen concentrations down to 56 mg/kg have been analyzed with a high precision. The rapid MICRO N cube offers the matrix independent Dumas method in combination with high sensitivity, ideal for applications in the polymer industry. Elementar Analysensysteme GmbH Donaustraße 7 63452 Hanau, Germany phone: +49 6181 9100-0 email: [email protected] web: www.elementar.de

 Open the catalog to page 1

All Elementar catalogs and technical brochures

  1. BiovisION

    4  Pages

  2. AromavisION

    4  Pages

  3. iso TOC cube

    4  Pages

  4. EnvirovisION

    4  Pages

  5. soli TOC

    4  Pages

  6. enviro TOC

    4  Pages

  7. UNICUBE

    4  Pages

  8. vario-EL-cube

    4  Pages

  9. iso

    4  Pages

  10. Liquiface

    2  Pages

  11. GC-5

    4  Pages

  12. 18 O-analysi

    8  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.