Specifications Overview
The Flexor EL150™ is a benchtop chemistry system designed for high throughput and reliability. It can perform up to 133 photometric tests and 266 ISE tests per hour. The system features a combined rotor for samples and reagents, with 30 refrigerated reagent positions and 25 barcode-readable sample positions. The pipetting system includes programmable syringes for precise reagent and sample handling.

Key Features
- Cuvette rotor with 48 semi-disposable cuvettes, supporting over 10,000 tests per rotor.
- Quartz-iodine lamp for light source, with a wavelength range of 340-800 nm.
- Photometric range of -0.1 to 3.0 absorbance with a resolution of 0.001 abs.
- Analytical modes include kinetic measurement, mono- and bichromatic end point measurement, and non-linear calibration curves.
- Quality control features include Westgard rules and Levey-Jennings plots.

Standards and Environmental Conditions
The system complies with CE-IVD, CB, and UL standards. It operates within a temperature range of 15-32°C and humidity of 15-85% RH. No dedicated plumbing is required.

Interface and Software
The Flexor EL150 features a state-of-the-art host-query interface via RS232 or Ethernet. It includes a touch screen navigation system running on Windows™ Embedded, designed for ease of use and productivity.

Innovations and Maintenance
- Integrated barcode reader for positive sample identification.
- Dry ISE module with patented solid-state electrode technology.
- Proactive maintenance kit for cost-effective annual maintenance.

Company Overview
ELITech Clinical Systems is a leader in benchtop chemistry systems, focusing on proximity laboratories. The Flexor series is known for its reliability and ease of use, supported by a global distribution network.

Conclusion
The Flexor EL150 System is a compact, fully-featured chemistry analyzer offering enhanced performance and minimal maintenance. It is ideal for primary, STAT, or backup testing needs, providing reliable results with reduced operational costs.
See more

Catalog excerpts

Flexor EL150-1

Specifications REAGENT AND SAMPLE SYSTEM • One rotor combining both sample and reagent positions; • Inner rotor ring: - refrigerated reagent positions 30 for 10 ml, 25 ml and 50 ml reagent bottles; - Cooled to 10°C +/-4°C at normal laboratory conditions; - All reagent positions can be assigned as R1, R2 and R3; • Outer rotor ring: - 25 barcode readable positions; - All positions fit 13x75 mm primary and secondary tubes and pediatric cups; - All positions can be used for Calibrators, Controls, Normal, Pediatric and STAT samples. Pipetting system • 1000 μl Reagent syringe: - R1 volume 110 - 400 μl; - R2 volume 0 - 180 μl; - R3 volume 0 - 180 μl; - Programmable in 1 μl steps; • 100 μl Sample syringe; - Sample volume 2-30 μl; - Programmable in 0.1 μl steps. • Programmable dilution ratios 1:5 up to 1:200 in one step increments with 3 possible diluents; • Pre-heated probe with level detection, collision protection and integrated mixer. CUVETTE ROTOR • Cost effective, semi-disposable cuvette rotor with 48 cuvettes, path length 7mm; • >10.000 tests per rotor; • Typical reagent consumption 250 μl per test; • Measuring temperature 37˚C, controlled by Peltier elements. LIGHT SOURCE • Quartz-Iodine lamp 12V-20W. WAVELENGTH RANGE • 340 - 800 nm; • Optical unit with 8 position filter wheel; • Automatic wavelength selection; • 340, 405, 505, 546, 578, 620, 660, 700 nm standard installed; • Other wavelenghts available on request. Interface • tate of the art Host-Query interface S available; • Host: RS232 or Ethernet (TCP/IP) through LIS-2A protocol. PHOTOMETRIC RANGE • -0.1 to 3.0 Absorbance; • Resolution 0.001 Abs. ANALYTICAL MODES (SINGLE, DUAL AND TRIPLE REAGENT SYSTEM) • Kinetic measurement with linearity check; • Mono- and bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction; • Two point measurement; with or without slope blank; • Graphic plot of all measuring points; • Predilution and automatic reflex dilution as needed; • Non-linear calibration curves; • Prozone check for immunology tests; • Cut-off declaration; • Calculated tests. QUALITY CONTROL • Up to 15 different controls can be defined, 3 per test; • Westgard rules; • Levey-Jennings plots; • Quality control statistics. Water Consumption • ~950 mL per hour max, continuous operation. STANDARDS and regulations • CE-IVD; • CB; • UL. DIMENSIONS (BenchTop , self Contained Unit) • 0cm (36in) x 75cm (30in) x 60cm (24in) 9 (W x H x D); • Weight: 75 kg (165 lbs.). About ELITech Clinical Systems… ENVIRONMENTAL CONDITIONS • Temperature: 15 – 32 °C (59 – 90 °F); • Humidity: 15 – 85% RH; • Altitude: up to 2000 m; • Plumbing: No dedicated system water or drain required; • Electrical: - Voltage: 100 – 240 Vac; - Frequency: 50 – 60 Hz; - Power (max): 400VA. ELITech Clinical Systems is a global leader in BenchTop Chemistry Systems, with over 15,000 Chemistry Systems delivered. Our mission is to enable better medical decisions by bringing high-value diagnostic solutions and services to laboratories that are closer to the patient (proximity laboratories). ELITech Clinical Systems leadership originates from our ability to continuously provide differentiated products and services focused to meet the needs of proximity laboratories. ELITech Clinical Systems markets the Flexor Series, designed to deliver trouble free results laboratories can trust. Integrated PC • Touch screen navigation; • Operating System: Windows™ Embedded. Robust, Always gets the job done Trouble Free Operations with True workhorsereliability proven system Simplifies lab operations & reduces error Best in class operator interface, easy to train, BAR CODE READER • Hand held CCD bar code reader used for test requisition, reagent identification and automated programing of assays, controls, and calibrators. Right Results, On Time with mistake Designedand error proofing reducing capabilities Dry ISE Module • Patented Solid State Electrode Technology; • Indirect Measurement; • Dilution 1:14; • Measures Sodium, Potassium, Chloride and CO2. Best vendor services, application & lab operations support Management of difficult samples Cost effective, minimizes labor, maintenance & operate & maintain Proven methodologies, referenced and traceable The Flexor Series builds on the established robustness and reliability of the current Flexor systems, resulting from decades of proven experience in the design and manufacture of award winning BenchTop laboratory products. Together with a dedicated range of ready-to-use reagents from ELITech Clinical Systems, the Flexor Series provides an ideal chemistry solution for laboratories that are closer to the patient. Over the past decades, the ELITech Group Companies have established a solid distribution network that operates world-wide. Thousands of customers have experienced the quality, convenience and reliability of ELITech products. Local support is provided by dedicated and well trained sales and service organizations. Positive sample identification • Positive Sample Identification (PSID) via integrated barcode reader. Reads all popular formats including Codes 39, 128, 93, CODABAR, and Interleaved 2/5. PRINTER • Printer supported by MS Windows™. PROACTIVE MAINTENANCE KIT • Complete parts kit for annual preventive maintenance. CHEMISTRY system Distributor contact information: Content of this document is subject to change without prior notification Flexor EL150™ is a trademark of Vital Scientific BV Windows™ is a trademark of Microsoft Delivering Trouble Free Results Laboratories can Trust THROUGHPUT • Up to 133 photometric tests / hour; • Up to 266 ISE tests / hour. Delivering trouble free results Belgium Brazil France Italy Middle East & Africa Serbia Switzerland The Netherlands T: + 44 1442 869320 UK United States T: + 1 855 354 8324 + 1 855 - ELITECH