Catalog excerpts

Clinical Study ONCOERA® Closed System Drug Transfer Devices Smoke Test VIAL ADAPTER VIAL STOPPER - VA104F.20 TEST LAB NANOLAB Laboratories Group Exposure to Hazardous Drugs As the risks associated with occupational exposure to hazardous drugs become increasingly apparent, there is increasing awareness of the methods, procedures and devices involved in the preparation and administration of such drugs. All standards aim to provide healthcare workers with maximum protection, by minimizing the contact of hazardous drugs with the employees and environment. Reputable organizations such as NIOSH (National Institute of Occupational Safety and Health) and ISOPP (The International Society of Oncology Pharmacy Practitioners) recommend the use of closed system drug transfer devices (CSTD), which prevent the escape of toxic drugs and their aerosols into the environment, in addition to the use of personal protective equipment.
Open the catalog to page 1
Aim The purpose of this study was to evaluate the vapor containment efficiency of ONCOERA® CSTDs, in order to evaluate the airtight sealing properties during the preparation of hazardous drugs. Method Draeger Air Flow Tester Kit was used as a drug substitute simulating normal preparation conditions. Each product was observed during its operation for any release of smoke into the environment. As the positive control group, smoke release was observed from the syringe without the syringe adapter attached. The syringe adapter was locked into the needle-free connector of the demonstration kit....
Open the catalog to page 2
Smoke release of Air Flow Tester Kit Smoke release from the syringe as a positive control SF 201 prevents smoke release from the syringe SF 201 and Vial Stopper prevent smoke release from the ports SF 401 prevents smoke release from the syringe SF 401 and Vial Stopper prevent smoke release from the ports
Open the catalog to page 3
Clinical Study ONCOERA® Closed System Drug Transfer Devices Smoke Test According to the study using smoke simulation, ONCOERA® Vial Stopper and ONCOERA® Syringe Adapters together form a closed system, preventing the release of toxic gases and aerosols to the environment. In this way, it helps to ensure drug safety, employee safety and environmental safety during the preparation of hazardous drugs. ERASER MEDICAL LLC Pınar Cd. No:78 Pınarbaşı / Bornova / İzmir - TÜRKİYE Phone: +90 232 479 80 60 Fax: +90 232 479 80 25 www.eraser.com.tr, info@eraser
Open the catalog to page 4All ERASER MEDIKAL TIB. CIH. SAG. UR. PAZ. SAN. VE TIC LTD STI catalogs and technical brochures
-