video corpo

ETG 6500 Bio WM

ETG 6500 Bio WM

ETG 6500 Bio WM

Product catalog summary
Overview: The ETG 6500 BIO WM is a multicomponent gas analyzer designed for biogas applications, offering accurate and reliable measurement of O2, CO2, CH4, and H2S. It features a user-friendly interface with a touch screen monitor and supports data download via USB, Ethernet, and Wi-Fi connectivity.
Key Features:
  • Non-dispersive infrared (NDIR) technology for CO2 and CH4 measurement.
  • Electrochemical sensors for O2 and H2S.
  • Proprietary algorithm for N2 and calorific value calculation.
  • Low maintenance with no need for electrochemical cell replacement.
  • Temperature and pressure compensation to reduce span drift.
  • Modular design with high performance and low cost of ownership.
Specifications:
  • Response time: <1 liter/minute sample flow rate.
  • Data refresh rate: 1 second.
  • Warm-up time: 30 seconds ready, 3 minutes usable, 30 minutes full performance.
  • Operating temperature: 0° to 50° C; Humidity: up to 95% RH (non-condensing).
  • Power consumption: <300 W without condensate removal, <450 W with.
  • Touch screen: 5.7” resistive type.
  • External electrical supply: 100 to 240 Vac, 47-63 Hz.
  • Pneumatic connection: Rapid fittings 6.0 OD 4.0 ID.
Technical Data:
  • Methane (CH4): 0.01% resolution, 0-100% range, +/-1% F.S. accuracy.
  • Carbon Dioxide (CO2): 0.01% resolution, 0-100% range, +/-1% F.S. accuracy.
  • Oxygen (O2): 0.1% resolution, 0-25% range, +/-2% F.S. accuracy.
  • Hydrogen Sulphide (H2S): 0.1 ppm resolution, multiple ranges up to 10000 ppm, +/-2% F.S. accuracy.
Applications: Suitable for biogas plants, research centers, water treatment plants, biogas system integrators, and biomethane upgrading.
Mechanical and Environmental:
  • Dimensions: 60 x 80 x 35 cm; Weight: 25 Kg.
  • Case protection: IP55.
  • Operating altitude: -300 to 3000 m (-1000 to 10000 ft).
See more

Catalog excerpts

ETG 6500 Bio WM-1

ETG 6500 BIO WM multicomponent gas analzyer for Biogas O2, CO2, CH4, H2S MONITORING Suitable for applications: • Biogas Plants • Research Center • Water Treatment Plant • Biogas System Integrator • Biomethane Upgrading The ETG 6500 BIO WM series of gas analyzers by ETG are the ideal solution for Biogas/Biomethane updgrading measurement because of their accuracy, stability, reliability, broad measurement range, and the variety of available form factors. Unlike other analyzers, ETG 6500 non-dispersive infrared (NDIR) gas analyzers measure multiple gases in an instrument with a single optical path platform. ETG analyzers have the ability to measure CO2, H2S and O2 in addition to methane, and therefore provide the optimal combination of gases measurement for Biogas/Biomethane plants. We can measure H2S by NDUV technology that is maintenance free and doesn’t need Electrochemical cells replacement. The enhanced optics and electronics of our NDIR analyzers have virtually eliminated zero drift after the initial warm up period. The temperature and pressure compensation eliminates the major causes of span drift in the instruments. USER INTERFACE OVERVIEW Trend and digital value • Data downloadable on USB Pen Drive • Arm processor • Touch Screen monitor • Ethernet, Wi-Fi and USB Remoting • Low cost of ownership • Modbus, Profibus, Ethernet (optional) • Plug & Play • Customizable ETG Risorse e Tecnologia S.r.l. — Via Baione 2/K 10034 Chivasso (IT) — Tel: +39 011 192.708.90 — email: info@etgrisorse

 Open the catalog to page 1
ETG 6500 Bio WM-2

SPECIFICATIONS • NDIR (for CO2 and CH4) & ECD (for H2S, O2) technology • N2 and calorific value calculated by proprietary algorithm (standard) • Dust filter front panel • Condensate Removal System • Multipoint Sampling system • Highly modular & High performance • Totally developed in Italy Response Time Response time are specified at a sample flow rate of 1 liter per minute through the sample cell Data Refresh Rate Warm-up Time 30 seconds ready, 3 minutes useable, 30 minutes full performance Operating Temperature Operating Humidity Operating Altitude Zero & Span user selectable Power Consumption...

 Open the catalog to page 2

All ETG Risorse e Tecnologia catalogs and technical brochures

  1. ETG 9500 FTIR

    5  Pages

  2. ETG 6500 Syn

    2  Pages

  3. ETG PSS 100

    1  Page

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.