ELECTRIC SAFETY SOLUTIONS Insulation monitor Insulation monitor in IT systems GENERAL INFORMATION The main purpose of the ES 1000A Insulation Monitor is to provide clinical and/or maintenance personnel with an insulation monitoring system in AC electrical installations with equipment with connected DC components, for medical locations with an IT system. The Insulation Monitor also oversees the load current, and the temperature of the transformer for medical use, of a distribution board for IT system in medical locations. It has a real-time clock through which the date and time of the equipment is set, allowing to store in the alarm memory, the date and time at which a particular alarm has occurred. ALARM MONITORING Ground wire connection failure Insulation leak TECHNICAL DATA Supply voltage Nominal frequency Consumption INSULATION MEASUREMENT Isolation transformer overcurrent Response Value (10A default) Relative error percentage Relative error percentage Response time Internal resistance Response Value Continuous voltage allowed (Ufg) Relative error percentage Network capacity allowed Response time Network communication failure CURRENT MONITORING Insulation failure detector Temperature in the isolation transformer COMMUNICATION Communication with multi-repeater TDS 149-2 Failure locator per output line TEMPERATURE MONITORING Communication with multi-repeater MRP 249-2A General Alarm 1 RS485 output 1 RS485 output 1 switched relay output