ELECTRIC SAFETY SOLUTIONS IT power supply systems Insulation panel for IT system GENERAL INFORMATION The IT system insulation board provides a protected insulated power supply system, for surgical units and other critical care areas, designated as locations for wet procedures. Automatic switch Isolation transformer for medical use Transformer protection Insulation monitor Line insulation fault detector General ground busbar Circuit breakers TECHNICAL DATA ELECTRICAL CHARACTERISTICS 400V~ 3N | 230V~ 230V~ | 230V~ Three phase | Single phase 50-60Hz 3 to 10kVA Input voltage Output voltage Phases Frequency Power INSULATION MONITOR Power supply Frequency Safety class Fault locator Classification Monitored alarms Temperature in the isolation transformer Modbus connection Manual bypass element Isolation transformer overcurrent Communication failure ISOLATION TRANSFORMER Voltage Frequency Safety class Leak current Power Insulating class No load current Insulation resistance Ground wire connection failure Insulation leak AUTOMATIC SWITCH Blackout FORCED VENTILATION Max flow Power consumption TEMPERATURE SENSOR Precision electric safety for life etkho.c
Open the catalog to page 1ELECTRIC SAFETY SOLUTIONS IT power supply systems Insulation panel for IT system CONNECTION DIAGRAM
Open the catalog to page 22 Pages