Density

Density

Density

Product catalog summary
Company Overview: Eurotec Medical Systems s.r.l. is a leading Italian company specializing in the design and manufacture of X-ray devices. All products are developed and produced in Italy, focusing on innovative technological and mechanical solutions. The company is ISO 13485 certified.
Product Overview: The Density Series is a range of DEXA bone densitometers designed for diagnosing osteoporosis. These devices utilize a patented suspended arm architecture and a solid aluminum frame, enhancing accessibility and maintenance.
Technical Specifications:
  • Principle of Operation: Utilizes samarium filtered energies (40keV, 75keV) with a filtration of 30.5mm Al eq. No external shielding is required.
  • Performance: Precision and accuracy are both less than 1%. Results include BMD, BMC, area, Z-score, and T-score.
  • X-ray Generator: Features a monobloc, HF, fixed anode with microprocessor-controlled stability. Tube current ranges from 0.125 to 1.75mA.
  • Environmental Requirements: Operates in temperatures between 15 to 27°C and humidity levels of 15% to 80%.
  • Software: Includes protocols for various body parts and is DICOM 3.0 compliant. Features automatic QC procedures and system diagnostics.
Models and Features:
  • Density PB and NFB: Whole body models using pencil and fan beam technology, respectively.
  • Density S and S-NFB: Compact models with similar technology distinctions.
  • Scan Time and Dose: Varies by model and body part, with specific times and doses outlined for spine, femur, forearm, and whole body scans.
Software Features: The dVision software is user-friendly, offering automatic segmentation, single-window operations, and daily quality control without external phantoms. It supports multiple languages and provides remote assistance capabilities.
Osteoporosis Assessment: The WHO defines osteoporosis based on BMD, with T-scores categorizing bone health. The Density Series provides accurate assessments using these standards.
Patient Accessibility: Designed for easy access by disabled and elderly patients, with a table height of 60cm.
See more

Catalog excerpts

Density-1

DENSITY SERIES TECHNICAL SPECIFICATIONS MODEL Density S-NFB Pencil beam Pencil beam STANDARD PROTOCOLS SCAN TIME * (Normal mode) Single detector Multiple detectors Single detector Spine, Femur, Forearm, Whole Body Spine, Femur, Forearm Spine, Femur, Forearm Spine: 119s Femur: 119s Forearm: 95s Whole body (2 meters): 9 min. Spine: 48s Femur: 48s Forearm: 34s Whole body (2 meters): 4 min. Spine: 119s Femur: 119s Forearm: 95s - Spine: 48s Femur: 48s Forearm: 34s - DOSE PATIENT * (Normal mode) Spine: 12.37uGy Femur: 12.37uGy Forearm: 0.82uGy Whole body: 4.71uGy Spine: 15.12uGy Femur: 15.12uGy Forearm: 1.00uGy Whole body: 5.75uGy Spine: 12.37uGy Femur: 12.37uGy Forearm: 0.82uGy - Spine: 15.12uGy Femur: 15.12uGy Forearm: 1.00uGy - SCAN AREA WEIGHT Multiple detectors Spine, Femur, Forearm, Whole body *Time and effective exposure depend on patient’s morphology Principle of operation: Separation: samarium filtered (K-edge) Energies: 40keV, 75keV Filtration: 30,5mm Al eq External shielding: Not required: x-ray safety requirements may be different in some countries Environmental Dose Rate: <0.5µSv/h for typical usage @1m from x-ray source Performance: Precision: <1% Accuracy: <1% Reporting results: BMD, BMC, area, Z-score, T-score, Z-Score %, T-Score % Reference curves: NANHES, local databases Calibration: multi-points BMD and area Continuous Electronics Autocalibration: adaptive peak threshold setting Linearity range: 0,2g/cm² ÷ 2g/cm² Patient weight limit: 160kg X-ray generator: Type: x-ray monobloc, HF, fixed anode, microprocessor controlled stability Tube current: 0.125 - 1.75mA Constant potential source at 86keV Cooling: oil convection Collimator: single hole /multiple holes, 2mm diam. Beam size: 2mm X-ray detector: Type: ultra high efficiency NaI(Tl) scintillator integral line PMT Detector HV : negative, -200 to -1200V, stabilized Environmental requirements: Temperature: 15 to 27°C Humidity: 15% to 80% (non condensing) Minimum room dimensions: 3x3m Power requirements: 230VAC, 500W A different DEXA bone densitometer E U R O T E C M E D I C A L S Y S T E M S s . r. l . is one of the most important companies in Italy designing and manufacturing X-ray devices. Eurotec Medical Systems units are designed, developed a n d m a n u f a c t u r e d i n I T A LY, a i m i n g t o t h e m o s t i n n o v a t i v e technological and mechanical solutions. EUROTEC MEDICAL SYSTEMS is ISO 13485 certified. Software: Protocols: spine, Femur (right/left), Forearm (right/left), Total Body/Body composition (PB and NFB models), Pediatric (optional) Dicom 3.0 compliant (optional) Built-in automatic QC procedures and system diagnostics Bone edge detection: automatic Selectable quality (High resolution, Normal, Fast) Pre-set exam parameters automatically adjusted according to patient’s morphology PC and OS minimum requirements: Operating system: windows 64 - bit (7/8/8.1/10) Memory: 4GB Hard disk: 500GB Display: LCD or LED, 22”, resolution 1920*1080’’ Printer: inkjet or laser color Connection: Ethernet or 802.1n Wi-Fi (optional) Accessories: Washable and anallergic mattress Washable and anallergic pillow Covering clothes EUROTEC MEDICAL SYSTEMS srl Via Aldo Moro 5/7 - 24020 Scanzorosciate (BG) Italy tel: +39 0355297470 - FAX: +39 0350662392 [email protected] - www.eurotec-ms.com A bone densitometer for all: Easy to access for disabled and elderly with mobility problem. Table height is just 60cm. Specification are subject to modifications without prior notice. 22/10/2019 Rev.

 Open the catalog to page 1
Density-2

DENSITY SERIES dVISION The smart Density software OSTEOPOROSIS Osteoporosis is a medical condition in which the bones become brittle and fragile from loss of tissue, typically as a result of hormonal changes, or deficiency of calcium or vitamin D. Osteoporosis is one of the major diseases affecting the elderly and more precisely is the seventh most common complaint. The incidence among people over age 50 is about 50% mostly among women. b a: Normal bones - b:Osteoporosis WHO (World Health Organization) defined osteoporosis based on bone mineral density (BMD). ESSENTIAL DESIGN Density series are...

 Open the catalog to page 2

All Eurotec Medical Systems catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.