Exergen Temporal Artery Thermometry

Exergen Temporal Artery Thermometry

Product catalog summary
Overview: The document provides instructions and guidelines for using the Exergen Temporal Artery Thermometer, emphasizing the correct procedure to ensure accurate temperature readings.
Procedure:
  • Scan Across Forehead: Place the probe on the center of the forehead, depress the button, and slide the probe in a straight line to the hairline.
  • Slide Behind Ear: Lift the probe, touch behind the ear halfway down the mastoid process, and slide down to the soft depression behind the earlobe.
  • Read Temperature: Release the button to read the temperature.
Accuracy Assurance:
  • Ensure the skin is exposed by brushing hair aside.
  • Measure straight across the forehead, not down the side of the face.
  • For patients in a lateral position, measure only the up-side to avoid falsely high readings.
Rationale for Measurement Locations:
  • Forehead vs. Side of Face: The temporal artery is closer to the skin surface on the forehead.
  • Behind the Ear: This area helps eliminate false low readings due to evaporative cooling from sweating.
  • Behind the Earlobe: Accessible and reliable due to high blood flow, especially in cases of diaphoresis.
Special Considerations:
  • Covered or Traumatized TA Area: Use the area behind the ear lobe for accurate readings.
  • Infants: High perfusion allows measurement at either the TA or behind the ear lobe.
  • Obese Patients: The temporal artery remains accessible.
Color Indicators: Red, Orange, Yellow indicate hot; Green, Black, Blue, Violet indicate cold.
Contact Information: Exergen Corporation, Watertown, MA 02472, Phone: 617.923.9900, Website: www.exergen.com
See more

Catalog excerpts

Exergen Temporal Artery Thermometry-1

Exergen Temporal Artery Thermometry Green, Black, Blue, Violet = Cold Measure exposed skin. Brush hair and bangs aside if covering the area to be measured. Measure straight across the forehead and not down the side of face. Think of a sweatband. Measure only the up-side on a patient in a lateral position. The down-side will be insulated preventing the heat from dissipating, resulting in falsely high readings. Why scan across the forehead and not down side of the face? The TA is less than 2mm below the skin surface at the forehead, but tends to run deep below the skin at the side of the face. Why measure behind the ear as well as the TA? Measuring both areas eliminates the possibility of a false low temperature caused by the evaporative cooling from diaphoresis, which many times is not obvious. Why not use only the area behind the ear lobe? If your patient exhibits any grade of diaphoresis, vasodilation will be present. While perspiration results in a low temperature in the TA area, the area directly behind the ear lobe will exhibit the high blood flow necessary for the arterial measurement, and is the last area of the body to sweat. And, it is very accessible. The area behind the ear only works as a sole site if there is 100% vasodilation. In the absence of diaphoresis (or head trauma), vasodilation in the neck area is uncertain, making this area too variable to be reliable as a sole site. What if the TA area is completely covered with dressings or has been traumatized by burns or lacerations,? This is a good example of when the temperature could be obtained just behind the ear lobe, as the area will be 100% vasodilated in response to rushing more blood to the injury to promote the healing process. What about an infant Vasodilation is a normal state for infants, and the high rate of perfusion allows the measurement to be made either at the TA or behind the ear lobe (as long as the area is exposed) when the TA area is not readily accessible. What about a very obese patient The temporal artery is trapped between the skull and the skin making the TA easily accessible even on a morbidly obese patient. 1. Scan Across Forehead Place probe flush on center of forehead and depress button Lift probe from forehead, touch behind ear halfway down the mastoid process (the bone in back of the ear) and slide down to the soft depression behind the earlobe Keeping button depressed... Slide probe in a straight line across forehead to the hairline With button still depressed... Release button and read

 Open the catalog to page 1

All Exergen Corporation catalogs and technical brochures

  1. TAT-5000S-MUSB

    14  Pages

  2. TAT-5000S

    20  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.