DuraShock

DuraShock

DuraShock

Product catalog summary
Instructions:
Before using the DuraShock, review the indications, contraindications, and FAQs. Ensure proper fitting and integration into the patient's prosthetic system. The stiffness of the DuraShock is determined by the patient's weight and activity level. It is not designed for loads over 100 lbs (45 kg) in tension.

Warranty:
The product comes with a 1-year warranty from the date of fitting. The warranty is void if the weight rating is exceeded or if the pylon is modified improperly.

Product Specifications:
The DuraShock is available in various models based on weight and activity levels, with specific build heights and torque settings. The manual provides detailed specifications for Long Pylon, Short Pylon, and MiniShock models.

Installation:
Follow installation instructions carefully to avoid voiding the warranty. The pylon can be installed with the pyramid receiver in either position. Alignment screws should be tightened to specified torque settings. A damping ring is included to adjust performance.

Maintenance and Care:
The DuraShock is maintenance-free and filled with a permanent lubricant. Do not attempt to lubricate or remove the cap screw. Alignment screws should be checked annually.

Troubleshooting:
Common issues include excessive rotation or vertical displacement, which can often be resolved by securing the damping ring or checking weight limits.

Frequently Asked Questions:
The manual addresses questions about adjusting the device, orientation of attachment, and care instructions, including cleaning and handling exposure to water.

Contact Information:
Fillauer LLC and Fillauer Europe contact details are provided for further assistance.
See more

Catalog excerpts

DuraShock-1

DuraShock Product Manual

 Open the catalog to page 1
DuraShock-2

Instructions • Please review the indications, contraindications, and FAQ sections of the manual before use. These instructions should be read prior to fitting and followed to ensure the proper integration of the DuraShock into the patient’s prosthetic system. • The DuraShock stiffness is based on weight and activity level. Please provide accurate patient information so that the appropriate product may be selected. • The DuraShock is not intended for loads over 100 lbs. (45 kg) in tension. Activites such as snowboarding, water skiing, and attempting to remove the prosthesis by pushing on the prosthetic...

 Open the catalog to page 2
DuraShock-3

Product Specifications DuraShock, Long Pylon Moderate Activity High Activity Build Height Torque Setting Product Number DuraShock, Short Pylon Moderate Activity High Activity Build Height Torque Setting Product Number DuraShock, Short Moderate Activity High Activity Build Height Torque Setting Product Number

 Open the catalog to page 3
DuraShock-4

Installation Deviating from the installation instructions or modifying the DuraShock in any way will void any product warranty and could lead to product failure and injury to the patient. Orientation The pylon may be installed with the pyramid receiver in the proximal or distal position (up or down) without effect on the performance. If the DuraShock is frequently exposed to wet environments, it is best to place the pyramid receiver toward the distal end. Adjustments to alignment should be done below the DuraShock to assure that the DuraShock remains in line with the patient’s weight at heel...

 Open the catalog to page 4
DuraShock-5

Ensure the ring always has some tension on it to keep it from sliding off the shock. The dampener ring is released by sliding the two grooved sections apart. Cutting the 30 mm Pylon The minimum lengths for the long pylons are 5.75 in. (14.6 cm) for the 30 mm version and 6.75 in. (17.1 cm) for the 34 mm version. The minimum lengths for the short pylons are 4.75 in. (11.4 cm) for the 30 mm version and 5 in. (12.7 cm) for the 34 mm version. The pylon may be cut using a standard tube cutter, hack saw, or band saw. Care should be taken to ensure this valuable pylon is not cut too short. Pylons are...

 Open the catalog to page 5
DuraShock-6

Maintenance and Care The DuraShock pylon is designed to be maintenance free. Lubrication The DuraShock pylon is filled with a special permanent lubricant that will last the life of the unit. No lubricant should be used on or in the DuraShock as it may cause failure and will void the warranty. Cap Screw The cap screw in the top of the DuraShock pylon is not to be removed. It is the factory lubrication port. Turning this screw does not adjust the function in any way. Tampering with or removing this screw will void the warranty. Adjustment The alignment screws should be checked and torqued to the...

 Open the catalog to page 6
DuraShock-7

Frequently Asked Questions Can the rotation and shock function be adjusted or fine-tuned? Yes, using the damping ring, you may fine-tune the performance of the device. See instructions under “Damping Ring” on page 4 of this manual. Does it matter which end of the DuraShock attaches to the socket? No. It may be attached in either orientation—pyramid proximal or distal. If the DuraShock is frequently exposed to wet environments, it is best to place the pyramid receiver toward the distal end. Daily Care for the Patient Changes in Performance Any changes in performance or noises that might develop...

 Open the catalog to page 7
DuraShock-8

2710 Amnicola Highway Chattanooga, TN 37406 423.624.0946 Fillauer Europe www.fillauer.com

 Open the catalog to page 8

All Fillauer catalogs and technical brochures

  1. Orthotics

    74  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.